Unistat T350w HT Temperature Control System Datasheet

Contenido

  1. Overview p. 1
    Product name and general introduction to the Unistat T350w HT unit
  2. Spec pp. 1-2
    Technical data per DIN 12876 including performance and dimensions
  3. p. 2 table, 3 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
min. ambient temperature5 °C
from Serial-No.:S619282
1.0/26

Technical data according to DIN 12876

ParameterValue
Operating temperature range65...300 °C
Minimum temperature with water cooling15 °C
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power96 kW
Cooling power at 300°C60 kW
Cooling power at 200°C60 kW
Cooling power at 100°C30 kW
Circulation pump: MK pump
max. delivery200 l/min
max. delivery pressure5,5 bar
Pump connectionM38x1,5 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG1 1/4 male
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity33 l
Filling capacity expansion tank56 l
Overall dimensions WxDxH **800x1007x1606 mm
Net weight460 kg
Power supply (3 Phase)460V 3~ 60Hz
max. current (3 Phase)130 A
Fuse (3 phase)3x160A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
max. ambient temperature40 °C

** Overall dimensions WxDxH

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a 460 V 3-phase 60 Hz supply with a 130 A max current and 3x160 A fusing, so the electrical infrastructure must be sized accordingly.

Determines feeder, breaker and fuse sizing before the unit can be installed.

Source
“Power supply (3 Phase) 460V 3~ 60Hz”
Installation

Three-phase devices drawing more than 63 A are delivered without cable and without plug, so the installer must provide the power connection.

Affects installation scope and cost; cabling and plug must be procured separately.

Source
“Three-phase devices with current consumption greater than 63A --> without cable, without plug”
Installation

Cooling water must be supplied at 1 bar minimum differential pressure and no more than 6 bar, with G1 1/4 male connections.

Defines the cooling water supply requirements and connection size for the site.

Source
“min. cooling water differential pressure 1 bar”
Operation

The unit is designed for cooling water intake between 5 °C and 20 °C, and cooling power drops as the cooling water temperature rises.

Cooling water temperature directly affects achievable cooling capacity and must be verified at the site.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

If cooling water temperature falls below the condensation point, the cooling water pipes in the unit must be insulated.

Prevents condensation and corrosion damage inside the unit during cold water operation.

Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Selection

The Unistat T models have no active refrigeration system, so reaching 65 °C requires adequate heat loss from the application or suitable water cooling.

Critical for selection: the unit cannot cool below 65 °C on its own and depends on the process heat loss.

Source
“These devices do not have an active refrigeration system.”
Selection

For air-cooled versions the application heat loss must be between 0.1 kW and 2.8 kW depending on the unit, and ambient temperature must be well below 60 °C to reach 65 °C.

Determines whether an air-cooled variant is viable for the application's heat load.

Source
“The heat loss must be between 0.1kW and 2.8kW, depending on the unit.”
Safety

Leakage current exceeds 3.5 mA, which requires appropriate protective earthing and RCD selection.

Affects electrical protection design and compliance with local installation rules.

Source
“Attention: leakage current > 3,5mA”
Cumplimiento

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and combinations such as -5% voltage with +2% frequency are not allowed.

Must be checked against the site power quality to avoid invalid operating conditions.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

The pump connection is M38x1.5 male and pipework must use ball socket per DIN 3863 with sleeve nut per DIN 3870.

Ensures correct mating of hoses and piping to the circulation pump.

Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Operation

The maximum permissible kinematic viscosity of the thermal fluid is 50 mm²/s, limiting fluid choice at low temperatures.

Constrains thermofluid selection and low-temperature operability of the pump.

Source
“max. permissible kin. viscosity 50 mm²/s”
Installation

The unit has IP20 protection and a maximum ambient temperature of 40 °C, so it must be installed in a dry, protected location.

Restricts installation environment and enclosure requirements.

Source
“Degree of Protection IP20”
Maintenance

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only suitable cooling water may be used.

Guides cooling water quality and treatment to avoid corrosion and fouling.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Performance

Cooling power is 60 kW at 300 °C and 200 °C but drops to 30 kW at 100 °C, which matters for sizing at lower set points.

Cooling capacity is temperature-dependent and must be checked at the actual operating point.

Source
“at 100°C 30 kW”