Introduces the Unistat tango w unit and its main features
Specp. 2
Lists technical data according to DIN 12876 standard
p. 2table, 18 rows
p. 1table, 28 rows
Technical data according to DIN 12876
Parameter
Value
Consumption at water 15°C, flow 0°C
42 l/h
Consumption at water 15°C, flow -20°C
24 l/h
Consumption at water 25°C, flow 0°C
108 l/h
Verbrauch b. Wasser 25°C, Vorlauf -20°C
120 l/h
min. cooling water differential pressure
0,5 bar
max. cooling water pressure
6 bar
min. filling capacity
1,5 l
Filling capacity expansion tank
2,8 l
Overall dimensions WxDxH **
426x327x631 mm
Net weight
55 kg
Power supply requirement
220-240V 1~/2~ 50/60Hz
max. current
15,5 A
Fuse
16 A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S583228
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876
Characteristic
Value
Operating temperature range
-45...250 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
3 kW
Cooling power with Thermooil at 250°C
0,7 kW
Cooling power with Thermooil at 200°C
0,7 kW
Cooling power with Thermooil at 100°C
0,7 kW
Cooling power with Ethanol at 0°C
0,7 kW
Cooling power with Ethanol at -20°C
0,4 kW
Cooling power with Ethanol at -40°C
0,06 kW
Refrigeration machine
water-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)
R-1270 (A3, H220)
Global Warming Potential (GWP)
0
Refrigerant quantity
0,09 kg
CO2 equivalent
0,0 t
Circulation pump: max. delivery
55 l/min
Circulation pump: max. delivery pressure
0,9 bar
Pump connection
M24x1,5 male
max. permissible kin. viscosity
50 mm²/s
Cooling water connection
G1/2 male
Order-No.: 1000.0055.01; Unistat tango w
Engineering Facts
13 facts · every claim linked to a page
Installation
The unit requires cooling water intake between 5°C and 20°C, so the site water supply must be verified before installation.
Cooling water outside this range will reduce cooling power or cause condensation, affecting selection and installation.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation
If cooling water temperature can fall below the dew point, the cooling water pipes inside the unit must be insulated.
Prevents condensation and corrosion/electrical risk during installation and operation.
Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance
Cooling power ratings are valid only at 20°C room temperature, 15°C cooling water inlet and 0.5 bar differential pressure.
Engineers must not assume rated cooling power at other site conditions; selection must account for actual water and ambient temperatures.
Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 0,5 bar differential pressure between cooling water inlet and -outlet.”
Installation
The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Incompatible water chemistry can corrode or damage the cooling circuit.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”
Cumplimiento
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.
Site power quality must be checked; some voltage/frequency combinations are explicitly prohibited.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
Single/two-phase devices are delivered with a power cable and country-specific plug, which must be specified at ordering.
Affects ordering and electrical installation planning.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Installation
Three-phase devices below 63 A are supplied with cable but without plug, so a suitable plug must be provided on site.
Electrical installation must include a compatible plug and connection.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Pump connections use bore shape Y (60°) per DIN 3863 and require ball socket per DIN 3863 and sleeve nut per DIN 3870.
Ensures correct mating of hoses and fittings during installation.
Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Operation
For external open baths the expansion tank is blocked off, making the thermostat atmospherically sealed and allowing it to be located below or above the application level.
Important for installation layout and operation with open baths.
Source
“For external open baths the expansion tank will be blocked off. This means that the thermostat is atmospherically sealed and can be located below or above the level of the application.”
Safety
The refrigerant is R-1270 (A3, H220) with GWP 0 and a charge of 0.09 kg, which is flammable and requires appropriate safety measures.
A3 refrigerant classification affects safety assessment and handling during installation and maintenance.
Source
“Refrigerant (ASHRAE, GHS) R-1270 (A3, H220)”
Selection
The unit has a maximum permissible kinematic viscosity of 50 mm²/s for the heat transfer fluid.
Heat transfer fluid selection must respect this limit to avoid pump damage and poor performance.
Source
“max. permissible kin. viscosity 50 mm²/s”
Operation
The pump is variable speed with soft start and integrated pressure control, with an optional external pressure sensor.
Affects hydraulic setup and control strategy for external circuits.
Source
“Powerful variable speed pump (soft start) with integrated pressure control with optional external pressure sensor.”
Maintenance
A float switch in the sight glass is available as an optional accessory for extended security.
Useful for level monitoring and maintenance safety in critical applications.
Source
“float switch in sight glass for extended security”