Induztr
Resumen técnico Visor de documentos Modelos relacionados

MO302

Technical Reference Summary

Covered Models & Document Scope

Folleto de datos técnicos para un sistema de espectrómetro de masas MALDI controlado por el software LAUNCHPAD™; abarca manipulación de muestras, visualización de muestras, fuente de ionización, láser, analizador, detector, sistema de control/datos y software. No se imprimen designaciones de modelo, tamaños ni códigos de pedido.

Technical Data

Parameter Value Condición
Mass range 1 Da to 500 kDa p.2 ↗ linear mode
Mass accuracy <30 ppm p.2 ↗ linear mode, with internal calibration
Sensitivity 250 fmol p.2 ↗ linear mode, bovine serum albumin
XY stage step 10 µm p.1 ↗ microtiter plate footprint MALDI target
XY stage repeatability 10 µm p.1 ↗ microtiter plate footprint MALDI target
Target thickness 2 mm p.1 ↗ plain, 96 and 384 sample targets
Maximum target thickness 10 mm p.1 ↗ with optional adaptor
Turbomolecular pump pumping speed 250 l/s p.1 ↗ nominal, sample introduction, rotary backing
Turbomolecular pump pumping speed 250 l/s p.1 ↗ nominal, analyzer, two pumps with rotary backing
Camera magnification 25 x p.1 ↗ monochrome CCD camera
Ion extraction energy +25 kV/-20 kV kV p.1 ↗ linear mode, variable, software controlled
Ion extraction energy +20 kV/-20 kV kV p.1 ↗ reflectron mode, variable, software controlled
Laser wavelength 337 nm p.1 ↗ nitrogen laser, fixed focus
Laser pulse width 3 ns p.1 ↗
Laser energy 100 µJ p.1 ↗ nominal, per laser shot
Maximum laser pulse rate 50 Hz p.1 ↗ 50 laser shots per second
Linear flight tube drift length 1.2 m p.1 ↗ linear mode
Reflectron effective drift length 2.0 m p.1 ↗ reflectron mode
Transient recorder sampling rate 2 GHz p.1 ↗ 8 bit, 16 bit accumulator
Second transient recorder sampling rate 125 MHz p.1 ↗ neutral detection, 8 bit, 64 kB RAM
RAM 16 GB p.1 ↗ control PC
HDD storage 256 GB p.1 ↗ control PC
SSD storage 1 TB p.1 ↗ control PC
Monitor size 24 in p.1 ↗ control PC
Operating system Microsoft® Windows® 10 64-bit p.1 ↗ control PC and LAUNCHPAD™

Operating Limits

  • Maximum laser pulse rate: 50 Hz p.1 ↗
  • Maximum target thickness with optional adaptor: 10 mm p.1 ↗
  • Mass range upper limit: 500 kDa p.2 ↗