Induztr
Résumé technique Visionneuse de documents Modèles associés

Agilent 240/280 Series Atomic Absorption Spectrometers User Guide

Sommaire

  1. Warning pp. 3-4
    Safety warnings, hazard symbols, and flashback response procedures
  2. Procedure pp. 5-6
    Connecting and configuring the Duo system
  3. Warning pp. 7-25
    Electrical, thermal, gas cylinder, hose, and flame safety guidance
  4. Overview pp. 26-27
    Graphite furnace and Zeeman operation safety notes
  5. Overview pp. 28-32
    Front and rear views of flame and graphite furnace instruments
  6. Config. pp. 33-37
    Electrical, gas supply, and PC requirements for installation
  7. Procedure pp. 38-45
    IEEE address, autosampler, USB-RS232, software startup, Tube-CAM driver
  8. Overview pp. 46-52
    SpectrAA software windows, worksheets, reports, and help
  9. Procedure pp. 53-56
    Starting the system and initial operation steps
  10. Procedure pp. 57-62
    Daily checks, surface cleaning, filters, and flashback prevention
  11. p. 3 table, 25 rows
  12. p. 4 table, 28 rows
  13. p. 5 table, 28 rows
  14. p. 6 table, 21 rows
  15. p. 7 table, 13 rows
  16. p. 30 table, 11 rows
  17. p. 33 table, 18 rows
  18. p. 36 table, 5 rows
  19. p. 37 table, 17 rows
  20. p. 56 table, 3 rows

目录

РазделСтраница
Agilent 240/280 系列原子吸收(包括塞曼)3
目录1
安全规范和危险提示7
一般信息7
安捷伦原子吸收文档8
惯例8
其他信息8
验证安全状态8
电气危险9
热量、蒸气和烟雾10
压缩气瓶的储存和处理11
气体软管和接头12
紫外线辐射12
警告符号13
彩色标记14
火焰操作15
易燃溶剂15
用于火焰操作的压缩气体和气瓶16
乙炔17
一氧化二氮18
燃烧头18
雾化器19
液封盒20
高温危险21
高氯酸21

目录

РазделСтраница
Agilent 240/280 系列原子吸收(包括塞曼)
回火22
发生回火时的应对措施25
回火后恢复运行25
石墨炉和塞曼操作26
气体26
高温危险26
蒸气和烟雾26
紫外线辐射26
磁场(仅限塞曼)26
2
简介27
仪器概述28
火焰原子吸收仪器前视图28
火焰原子吸收仪器后视图29
石墨炉(塞曼)仪器前视图30
石墨炉(塞曼)仪器后视图31
石墨炉 (D2) 仪器前视图(安装 GTA 120 附件的 240FS)32
石墨炉 (D2) 仪器后视图(安装 GTA 120 附件的 240FS)33
安装要求33
电气连接34
气体供应36
3
安装37
PC 要求37
安装流程概述38
GPIB 通信38
安装 USB-GPIB 转换器38

目录

РазделСтраница
Agilent 240/280 系列原子吸收(包括塞曼)5
安捷伦原子吸收光谱仪 Duo 系统的连接与配置39
Duo 运行设置和 IEEE 电缆的连接39
为安捷伦原子吸收 Duo 操作配置 IEEE 地址40
配置 SPS 4 自动进样器41
确定 SPS 4 COM 端口42
配置软件以使用手动进样42
USB 转 RS-232 接口43
启动 SpectrAA 软件44
SpectrAA 发行说明和软件状态公告44
安装 Tube-CAM 石墨炉摄像头选件的驱动程序45
打印机设置45
故障排除45
服务安装模块45
4
软件概述47
启动 SpectrAA 软件47
SpectrAA 窗口47
索引页面47
工作表窗口48
报告窗口49
管理窗口50
使用 SpectrAA 软件50
菜单栏50
工具栏50
对话框50
帮助51
搜索帮助51

Строка «4» напечатана без номера страницы; в таблице она оставлена как отдельная строка с пустой второй ячейкой.

目录

РазделСтраница
Agilent 240/280 系列原子吸收(包括塞曼)5
入门指南53
仪器设置53
连接53
电源54
移动仪器55
启动系统55
分析前检查清单55
维护与故障排除57
计划58
每日 — 运行前58
每日 — 运行后58
每周58
每年59
清洁59
仪器表面59
窗片60
过滤器60
气瓶60
防止回火61
备件61

安全规范和危险提示

РазделСтраница
一般信息7
安捷伦原子吸收文档8
验证安全状态8
电气危险9
热量、蒸气和烟雾10
压缩气瓶的储存和处理11
气体软管和接头12
紫外线辐射12
警告符号13
彩色标记14
火焰操作15
回火22
石墨炉和塞曼操作26

Agilent 240/280 系列原子吸收(包括塞曼)石墨炉(塞曼)仪器前视图

编号部件名称
1GTA 120 石墨管原子化器
2塞曼工作台
3PSD 120 样品盘
4PSD 120 可编程进样器
5清洗溶液
6GTA 120 仪器状态指示灯和电源开关
7塞曼原子吸收仪器状态指示灯和电源开关
8PSD 120 废液口
9PSD 120 清洗管
10PSD 120 进样针
11灯室(此处显示了八个灯,也可提供四灯室)

有关这些附件的更多信息,请参阅 GTA 120 和 PSD 120 随附的用户指南。

其中:

编号说明
1附件串行端口,如 GTA 观测/排烟附件
2USB 摄像头电缆端口
3原子吸收仪器 IEEE 488 电缆端口
4原子吸收仪器电源线
5原子吸收仪器保险丝和电源设置外壳
6UltrAA 灯选件端口
7乙炔 (C2H2) 气体管线
8空气管线
9一氧化二氮 (N2O) 气体管线
10SIPS 泵选件端口
11空气吹扫口
12GTA 120 电源线
13GTA 120 保险丝
14进水口
15出水口
16替换气进气口,通常为氮气或空气
17正常进气口,通常为氩气
18GTA 120 IEEE 488 电缆端口

气体供应

乙炔 (C2H2)空气一氧化二氮 (N2O)空气吹扫
仪器级,纯度 > 99.5%必须洁净干燥(如有需要,可使用空气过滤器)仪器级,纯度 > 99.5%
允许范围65–100 kPa (9.5–14.5psi)245–455 kPa (35–65 psi)245–455 kPa (35–65 psi)245–455 kPa (35–65 psi)
推荐75 kPa (11 psi)350 kPa (50 psi)350 kPa (50 psi)
典型流速0–10 (L/min)13.5–20 (L/min)11–16 (L/min)10 (L/min)
连接长 1.8 m(6 英尺),内径 6.3 mm(1/4 英寸)的红色软管,带有 9/16 英寸-18UNF L/H 螺纹接头#长 1.8 m(6 英尺),内径 6.3 mm(1/4 英寸)的黑色软管,带有 9/16 英寸-18UNF 接头#长 1.8 m(6 英尺),内径 6.3 mm(1/4 英寸)的蓝色软管,带有 3/4 英寸-16UNF 接头#可用于 6.3 mm(1/4 英寸)软管的倒钩

# 提供适配器;其他气体连接:样品室:用于燃烧头的推接式空气/N2O 接头;用于燃烧头的推接式 C2H2 接头

Содержание

РазделСтраница
安装
Agilent 240/280 系列原子吸收(包括塞曼)37
3
安装
PC 要求37
安装流程概述38
GPIB 通信38
安捷伦原子吸收光谱仪 Duo 系统的连接与配置39
配置 SPS 4 自动进样器41
配置软件以使用手动进样42
USB 转 RS-232 接口43
启动 SpectrAA 软件44
SpectrAA 发行说明和软件状态公告44
安装 Tube-CAM 石墨炉摄像头选件的驱动程序45
打印机设置45
故障排除45
服务安装模块45

将气体的输送压力设置为:

气体推荐 kPa推荐 psi允许范围 kPa允许范围 psi
乙炔751165–1009.5–14.5
空气35050245–45535–65
一氧化二氮35050245–45535–65

Engineering Facts

10 facts · every claim linked to a page
Safety

The instrument must only be used as specified by the manufacturer, otherwise its built-in protection may be compromised.

Defines the boundary of intended use that must be respected during selection and operation.

Source
“如果未按制造商规定的方式使用设备,则设备提供的保护可能会受到影响。”
Safety

Cylinders must be stored upright, secured, ventilated, and kept below 52 °C to avoid pressure-relief venting.

Sets lab layout and ventilation requirements for gas cylinder storage.

Source
“保持气瓶冷却。此要求适用于所有压缩气瓶。气瓶具有泄压装置,若温度超过 52 °C (125 °F),将释放气瓶内的气体”
Safety

Never use oxygen or oxygen-enriched air as oxidant and never use any gas other than acetylene as fuel.

Constrains the gas supply selection for flame operation.

Source
“切勿使用氧气或富氧空气作为助燃气,否则会引起爆炸。”
Operation

Acetylene must be supplied below 105 kPa, with the instrument designed for 65–100 kPa fuel supply pressure.

Defines the regulator setting and supply pressure needed for safe acetylene operation.

Source
“请在低于 105 kPa (15 psig) 的压力下使用乙炔。如果高于该压力,乙炔会自发爆炸。安捷伦原子吸收仪器设计在 65–100 kPa (9.5–14.5 psig) 的燃气供应压力下运行。”
Maintenance

Burner head slit widths must not exceed 0.46 mm for N2O/acetylene and 0.54 mm for air/acetylene, otherwise the head must be replaced.

Gives the wear limit for deciding when a burner head must be replaced.

Source
“Mark 7 0.46 mm(0.0181 英寸),用于 N2O/乙炔燃烧头”
Safety

A burner head that has experienced a flashback must be treated as damaged and never reused.

Defines mandatory replacement after a flashback event.

Source
“凡是经历过回火的燃烧头均应被视为已损坏和损毁。”
Installation

Only the same type and rating of fuse may be used to maintain fire protection.

Specifies the correct replacement fuse ratings for each model.

Source
“为了能持续防止火灾危险,仅可更换相同类型和额定值的保险丝。”
Installation

SpectrAA software only recognizes COM port numbers between 1 and 4 for USB-to-RS-232 adapters.

Constrains COM port assignment when configuring serial accessories.

Source
“SpectrAA 软件只识别 1 到 4 之间的值。”
Operation

Before analysis, verify acetylene cylinder pressure is above 700 kPa so acetone is not drawn into the instrument.

Sets the minimum cylinder pressure check before each analysis run.

Source
“检查乙炔气瓶压力是否高于 700 kPa(约 100 psi),确保丙酮不会被吸入仪器”
Safety

The instrument weighs over 50 kg and must be lifted by two or more people.

Affects handling and installation planning for the instrument.

Source
“仪器重量超过 50 kg (110 lb)。请勿尝试单独抬起仪器。抬动或搬运仪器时,务必由两个或更多人共同完成。”