Agilent 4150 TapeStation System Manual

Sommaire

  1. Overview pp. 17-24
    System description, ScreenTape architecture, and product overview
  2. Spec pp. 33-36
    Bench space and environmental requirements
  3. Procedure pp. 37-48
    Unpacking, working area preparation, and initial setup
  4. Overview pp. 49-60
    Controller and analysis software, settings, and security module
  5. Procedure pp. 61-96
    Sample preparation protocols, pipetting, and centrifugation recommendations
  6. Spec pp. 97-110
    RNA assays, peak integration, quantitative range, and RINe
  7. Procedure pp. 111-124
    Tests, verification procedures, and reporting
  8. Procedure pp. 127-172
    Preventive maintenance, electrode cartridge exchange, cleaning, and cover assembly
  9. Warning pp. 173-186
    Troubleshooting software messages and corrective actions
  10. Procedure pp. 187-216
    Sizing issues, DIN values, peak annotation, and degraded samples
  11. Procedure pp. 217-228
    Windows Device Manager and universal corrective actions
  12. pp. 229-231
    Additional notes and reference information
  13. p. 30 table, 14 rows
  14. p. 120 table, 2 rows
  15. p. 204 table, 4 rows
  16. p. 63 table, 8 rows
  17. p. 113 table, 7 rows
  18. p. 124 table, 2 rows
  19. p. 188 table, 5 rows
  20. p. 189 table, 7 rows
  21. p. 191 table, 2 rows
  22. p. 193 table, 4 rows

Table 2 Physical Specifications

TypeSpecification
Weight17.8 kg (39.2 lbs)
Dimensions (w×d×h)254 x 510 x 417 mm (9.6 x 20.1 x 16.4 inches)
Line voltage100 – 240 VAC ± 10 %
Line frequency50 or 60 Hz ± 5 %
Power consumption50 VA
InterfaceUSB (Instrument: USB-B; PC: USB-A)
Ambient operating temperature10 – 40 °C (50 – 104 °F)
Operating humidity15 – 80 %, non-condensing
Storage temperature-40 to 70 °C (-40 to 158 °F)
Operating altitudeUp to 3000 m (9842 ft)
Safety standardsIEC, EN, CSA, UL, Overvoltage category II, Pollution degree 2 For indoor use only.
ISM ClassificationISM Group 1 Class A According to CISPR 11
Sound pressure<70 dB (A) According to ISO 7779:1988/EN 27779/1991
Assay specific temperature range12 – 37 °C (54 – 99 °F) for D1000 ScreenTape 17 – 37 °C (63 – 99 °F) for High Sensitivity D1000 ScreenTape 15 – 30 °C (59 – 86 °F) for Genomic DNA ScreenTape 14 – 30 °C (57 – 86 °F) for RNA and High Sensitivity RNA ScreenTape 15 – 30 °C (59 – 86 °F) for D5000 ScreenTape 15 – 30 °C (59 – 86 °F) for High Sensitivity D5000 ScreenTape 16 – 30 °C (61 – 86 °F) for Cell-free DNA ScreenTape

This is an ISM Class A product intended for use in industrial environment. In a domestic environment, this product may cause radio interference, in which case the user may be required to take adequate measures. The assay specific temperature range is measured at the ScreenTape nest. Choose the location of the instrument and the ambient conditions to fit to the working range of the assay else the start of runs will be disallowed.

Table 6 Computer Environment

System Test SuiteTroubleshooting Test Suite
Computer SpecificationsChecks the computer specifications and configuration.YesYes
Software EnvironmentChecks the compatibility of TapeStation Controller Software with the computer.YesYes

Table 18 RNA - Possible causes of incorrect quantification in order of probability.

Root causeExplanationSolution
Insufficient MixingBuffer and sample was not sufficiently vortexed. The dye could not incorporate completely into the sample.See “Mixing Recommendations” on page 92 and the Agilent Information Center for more information.
Sample concentration outside recommended range for applicationLinear concentration estimation cannot be achieved for samples exceeding the limits in the specification table.Either dilute or concentrate the sample until it is within the recommended range for the application as stated in the respective assay Quick Guide; then prepare a new run. Use the High Sensitivity assay if applicable (“Quantitative range” on page 105, Table 3 on page 63).
Incorrect pipetting technique or pipette calibrationIncorrect pipetting technique or uncalibrated pipettes can lead to poor pipetting performance which in turn alter sample/buffer ratio in an unpredictable manner.See “Good Pipetting Practice” on page 83 and follow guidelines of the respective pipette vendor.
Sample not collected at bottom of vialInsufficient volume pipetted by the robotics and applied to the lane may affect the migration behavior and alter the band shape.Make sure to collect all liquid at the bottom of the vial by centrifugation. See “Centrifugation Recommendations” on page 95. Ensure that after removing strip caps and inserting the tube strips or 96-well sample plate into the TapeStation instrument liquid is still located at the bottom of the well, and is not spilled.

Table 3 ScreenTape Assay portfolio

ScreenTape AssayQuick Guide Number
Agilent D1000 ScreenTape AssayG2991-90031
Agilent High Sensitivity D1000 ScreenTape AssayG2991-90131
Agilent RNA ScreenTape AssayG2991-90021
Agilent High Sensitivity RNA ScreenTape AssayG2991-90121
Agilent Genomic DNA ScreenTape AssayG2991-90041
Agilent D5000 ScreenTape AssayG2991-90051
Agilent High Sensitivity D5000 ScreenTape AssayG2991-90151
Agilent Cell-free DNA ScreenTape AssayG2991-90060

Table 5 Test suite groups

Test Suite GroupsDescription
Computer EnvironmentEvaluation of computer specification and installation status.
SetupPrepares the instrument, requests insertion of material, checks status LED functionality. Information on current Maintenance counters, past Electrode Cartridges, past needle changes and current installed software is reported.
VisionCamera and LED illumination are checked using the TapeStation Test Tape.
ElectrophoresisThe voltage supply, current readout, and electrodes are checked using the TapeStation Test Tape.
RoboticsRobotics parts are checked by dedicated movements. Tip detection is checked using a full tip rack.
PipettingThis check procedure in the Hardware Troubleshooting visually checks pickup of liquid from tube strips.
CleanupResets the instrument to normal operational mode and requests removal of material.

Table 10 Robotics

System Test SuiteTroubleshooting Test Suite
Needle To PumpYesYes
Tape HandlingYesYes

Table 13 DNA - Possible causes of incorrect quantification results in order of probability

Root causeExplanationSolution
Insufficient MixingBuffer and sample was not sufficiently vortexed. The dye could not incorporate completely into the sample.See “Mixing Recommendations” on page 92 and the Agilent Information Center for more information.
Sample concentration outside recommended range for application.Linear concentration estimation cannot be achieved for samples exceeding the limits in the specification table.Either dilute or concentrate your sample until it is within the recommended range for application as stated in the respective assay Quick Guide, then prepare a new run. Use the High Sensitivity assay if applicable (Table 3 on page 63).
Incorrect peak integrationThe algorithm cannot establish a correct concentration estimation if marker or sample peaks are not completely encompassed as a signal from either marker or sample is left out of the calculation.Ensure that all sample peaks and markers are integrated correctly in the TapeStation Analysis software, by clicking and dragging so that the whole peak is encompassed. See the Agilent Information Center for more information.
Incorrect analysis typeThe Gel and Electropherogram view are designed to analyze discrete peaks within samples. The Region view is optimized for analyzing samples with broad peaks. Region view uses a different method to calculate concentration and sizing.Use Gel view for samples with narrow peaks and the Region view for samples with broad peaks or smears such as libraries. See “Peak Maxima Versus Average Molecular Weight Sizing and Molarity” on page 102 for more information.
Sample and the marker relevant for quantification have merged.Additional sample overlaying with the marker will alter the known concentration of the marker that is used for concentration calculation.Ensure that your sample is within the recommended sizing range for the applications, see “Sizing Range” on page 104. Contaminating beads can alter the run behavior of the sample and can introduce artifacts, see “TapeStation Analysis in the Agilent SureSelect and HaloPlex Workflow” on page 202. If using the Sure Select Protocol, ensure that all AMPure beads are removed by increasing the time that the sample is on the magnetic plate.

Table 13 DNA - Possible causes of incorrect quantification results in order of probability

Root causeExplanationSolution
Sample not collected at bottom of vialInsufficient volume pipetted by the robotics and applied to the lane may affect migration behavior and alter the band shape.Make sure to collect all liquid at the bottom of the vial by centrifugation. See “Centrifugation Recommendations” on page 95. Ensure that after removing strip caps and inserting the tube strips or 96-well sample plate into the TapeStation instrument, liquid is still located at the bottom of the well, and is not spilled.
Incorrect protocol usedDifferences exist between Standard and High Sensitivity assay protocols.Ensure that the correct sample protocol for the application is used, see “Use the Correct Protocol” on page 91.
Incorrect reagents usedReagents from different assays may differ in dye and marker concentration. Furthermore markers used are different and upper markers may not be present.Ensure that the correct reagent buffer for the application is used, see “Use the Correct Protocol” on page 91.
Sample prepared too long before analysisEvaporation or settling may alter the sample to buffer ratio.Ensure that your run is started immediately after sample preparation.
The sample/buffer mix may evaporate over time if a 96-well sample plate was used which was not sealed with the recommended seal foil.Always cover with the recommended foil seal (PN 5067-5154) and ensure tight sealing throughout the 96-well sample plate.
Incorrect pipetting technique or pipette calibration.Incorrect pipetting technique or uncalibrated pipettes can lead to poor pipetting performance which in turn alter sample/buffer ratio in an unpredictable manner.See “Good Pipetting Practice” on page 83 for more information or follow guidelines of the respective pipette vendor.
Incorrect Marker peaks picked upDepending on the sample pattern, wrong peaks can get assigned as marker if running very close to the actual marker.Use the Aligned button to detect the correct markers. Compare to aligned/unaligned data of good, for example the ladder, and questionable samples, then assign correct markers by right-clicking. See the Agilent Information Center for more information. Ensure that all sample peaks are within the recommended sizing range for the application.

Table 14 DNA - Possible causes of incorrect sizing results in order of probability

Root causeExplanationSolution
Issues with the ladder laneAny issues with the run ladder lane will be annotated within the TapeStation Analysis software sample table under Observations.Manually insert or delete affected peaks. In assays where an electronic ladder is available (D1000, High Sensitivity D1000, and Cell-free DNA Assay), insert the electronic ladder to regain sizing information, see the Agilent Information Center for more information. In assays where an electronic ladder is not available (D5000, High Sensitivity D5000, and Genomic DNA Assay), re-run affected samples and ensure that Good Measurement Practices are followed, see “Good Measurement Practices” on page 79. For Genomic DNA Ladder considerations, see “Genomic DNA Assay” on page 186.
Insufficient migration due to presence of saltSample matrices exceeding the maximum buffer strength specifications may alter currents in an unpredictable manner leading to short running lanes, see “Maximum Buffer Concentration in Sample” on page 107.Dilute your samples to ensure low levels of buffer salt. True migration profiles can be seen by unaligning your gel image by pressing the Aligned button in the TapeStation Analysis software. Please refer to the salt tolerance guidelines for your assay in the appropriate Quick Guide (Table 3 on page 63).

Table 16 Genomic DNA - Probable causes for missing or incorrect DIN results

Root causeExplanationSolution
Too few ladder peaks have been detectedIn order to calculate a DIN value a correct sizing is needed which can only be established with a correctly run and integrated ladder.Check the electropherogram trace of the ladder for incorrectly or not assigned peaks; then manually assign ladder peaks. Make use of the demo file to identify correct peaks. See the Agilent Information Center for more information. In case the biggest ladder fragment is degraded see “Genomic DNA Assay” on page 186. Assign a different lane as the ladder (if applicable).
Incorrect marker peak picked up or no marker peaks detectedSample matrices exceeding the maximum buffer strength specifications may alter currents in an unpredictable manner leading to short running lanes, see “Maximum Buffer Concentration in Sample” on page 107.Dilute your samples to ensure low levels of buffer salt. True migration profiles can be seen by unaligning your gel image by pressing the Aligned button in the TapeStation Analysis Software. Please refer to the salt tolerance guidelines for your assay in the appropriate Quick Guide (Table 3 on page 63). Rerun samples with fresh reagents.
Due to sensitivity, artifacts might be erroneously assigned as Lower Marker.Use the Aligned button in the TapeStation Analysis software to detect the correct markers. Compare to aligned/unaligned data of good and questionable samples; then assign correct markers by right-clicking. See the Agilent Information Center for more information. Ensure that all sample peaks are within the recommended sizing range for the application.
No ladder run on ScreenTape deviceFor Genomic DNA application, one ladder per ScreenTape must be run in order to ensure correct sizing and DIN calculation. The ladder should be run in the first available position on the ScreenTape device.Use the Assign button to use the correct ladder lane if a ladder has mistakenly been run in a position other than position A1 when using tube strips. Rerun samples if no ladder was used at all. In case a 96-well sample plate was used a ladder always has to be provided in position A1 from the tube strip location.

Engineering Facts

14 facts · every claim linked to a page
Selection

The instrument processes up to 16 samples per run, so throughput requirements must be matched to this entry-level capacity.

Determines whether the system fits the lab's sample throughput before purchase.

Source
“Up to 16 samples can be processed per run.”
Installation

The assay-specific temperature range is measured at the ScreenTape nest, so bench location and ambient conditions must match the assay's working range or runs will be blocked.

Site selection must account for local temperature at the nest, otherwise runs are disallowed.

Source
“The assay specific temperature range is measured at the ScreenTape nest.”
Installation

Approximately 75 mm of clear space is needed around the instrument for air circulation and electrical connections.

Bench layout must reserve clearance for cooling and cabling.

Source
“Allow approximately 75 mm (3 inches) of clear space around the instrument for air circulation and electrical connections.”
Installation

An additional ~25 cm above the listed height is required to open the lid, so low overhanging shelves must be avoided.

Prevents installing the unit under shelving that blocks lid opening.

Source
“Additional space of approximately 25 cm is required on top of the listed height in order to open the TapeStation lid.”
Installation

Two plug sockets must be available near the bench for the instrument and the laptop.

Electrical planning must provide two outlets at the bench.

Source
“Two plug sockets must be available near the allocated bench space for the instrument itself and a laptop.”
Installation

If shipment conditions are unknown or outside operating conditions, wait at least 5 h before unpacking to allow temperature and humidity equilibration.

Prevents condensation damage to electronics during commissioning.

Source
“wait for at least 5 h before unpacking to allow the instrument to adjust to the ambient temperature and humidity of the operating environment.”
Installation

Connections must be made in a specific order: power plug to socket, USB-A to laptop, then USB-B to instrument.

Correct cabling sequence avoids communication and power issues at setup.

Source
“Connect the instrument in the following order: 1. Power plug to power socket, 2. USB-A to laptop, 3. USB-B to TapeStation instrument.”
Safety

The power connector must remain easily reachable and unpluggable for emergency disconnection.

Installation must leave access behind the socket for emergency disconnect.

Source
“Make sure the power connector of the instrument can be easily reached and unplugged.”
Operation

A second ScreenTape device with sufficient lanes must be ready, with a 15-minute window once prompted.

Operators must stage a spare ScreenTape to avoid run interruption.

Source
“A time frame of 15 minutes applies once the TapeStation Controller software prompts for this.”
Operation

Loading tips that are too large cause poor quantitation accuracy, so tip size must match the pipetting step.

Tip selection directly affects measurement accuracy.

Source
“Loading tips that are too large lead to poor quantitation accuracy.”
Operation

Reagent kits are not interchangeable between standard and high sensitivity assays due to different dye and marker concentrations.

Prevents using the wrong kit with a ScreenTape type.

Source
“Reagent kits are not interchangeable between standard and high sensitivity assays due to different dye and marker concentrations.”
Maintenance

The piercing needle should be exchanged after 3840 pierces, and analysis is inhibited at 4480 pierces.

Plans consumable replacement before the instrument blocks further runs.

Source
“The piercing needle should be exchanged after 3840 pierces to ensure optimal instrument performance.”
Maintenance

An unused Needle Change Cartridge is a prerequisite for needle exchange and is single-use; an extracted needle cannot be remounted.

Spare part must be stocked before performing needle maintenance.

Source
“An unused Needle Change Cartridge (5067-5783) is prerequisite for the needle exchange and is recognized by the software.”
Operation

Hardware diagnostics can only be executed on one instrument at a time even when two are connected.

Affects scheduling of verification tests in multi-instrument setups.

Source
“Execution of hardware diagnostics is possible for one instrument at a time only.”