4200 TapeStation System Manual

Sommaire

  1. Warning pp. 10-18
    Safety warnings, EEE disposal and regulatory notices
  2. Overview pp. 19-30
    ScreenTape architecture and TapeStation product description
  3. Overview pp. 31-36
    Differences between TapeStation models and variants
  4. Procedure pp. 37-54
    Unpacking, hardware installation and setup steps
  5. Procedure pp. 55-66
    TapeStation Controller software settings, diagnostics and security
  6. Procedure pp. 67-96
    Running samples, good measurement and pipetting practices
  7. Procedure pp. 97-114
    Mixing, centrifugation and sample preparation recommendations
  8. Spec pp. 115-120
    RINe algorithm functional range and interpretation
  9. Procedure pp. 121-138
    System verification, diagnosis and troubleshooting checks
  10. Procedure pp. 139-195
    Preventive maintenance, electrode cartridge exchange and cleaning
  11. Procedure pp. 196-252
    Error scenarios, sizing issues and communication problems
  12. pp. 253-256
    Supplementary reference information and index
  13. p. 30 table, 14 rows
  14. p. 132 table, 2 rows
  15. p. 228 table, 4 rows
  16. p. 68 table, 8 rows
  17. p. 122 table, 8 rows
  18. p. 136 table, 2 rows
  19. p. 212 table, 6 rows
  20. p. 213 table, 7 rows
  21. p. 215 table, 4 rows

Table 2 Physical Specifications

TypeSpecification
Weight25.5 kg (56.3 lbs) (G2991A) 24.8 kg (54.7 lbs) (G2991B)
Dimensions (w×d×h)435 x 510 x 445 mm (17.1 x 20.0 x 17.5 inches)
Line voltage100 – 240 VAC ± 10 %
Line frequency50 or 60 Hz ± 5 %
Power consumption50 VA / 50 W
InterfaceUSB (Instrument: USB-B; PC: USB-A)
Ambient operating temperature10 – 40 °C (50 – 104 °F)
Operating humidity15 – 80 %, non-condensing
Storage temperature-40 to 70 °C (-40 to 158 °F)
Operating altitudeUp to 3000 m (9842 ft)
Safety standardsIEC, EN, CSA, UL, Overvoltage category II, Pollution degree 2 For indoor use only.
ISM ClassificationISM Group 1 Class A According to CISPR 11
Sound pressure<70 dB (A) According to ISO 7779:1988/EN 27779/1991
Assay specific temperature range12 – 37 °C (54 – 99 °F) for D1000 ScreenTape 17 – 37 °C (63 – 99 °F) for High Sensitivity D1000 ScreenTape 15 – 30 °C (59 – 86 °F) for Genomic DNA ScreenTape 14 – 30 °C (57 – 86 °F) for RNA and High Sensitivity RNA ScreenTape 15 – 30 °C (59 – 86 °F) for D5000 ScreenTape 15 – 30 °C (59 – 86 °F) for High Sensitivity D5000 ScreenTape 16 – 30 °C (61 – 86 °F) for Cell-free DNA ScreenTape

NOTE This is an ISM Group 1 Class A product intended for use in industrial environment. In a domestic environment, this product may cause radio interference, in which case the user may be required to take adequate measures. NOTE The assay specific temperature range is measured at the ScreenTape nest. Choose the location of the instrument and the ambient conditions to fit to the working range of the assay else the start of runs will be disallowed.

Table 6 Computer Environment

System Test SuiteTroubleshooting Test Suite
Computer SpecificationsChecks the computer specifications and configuration.YesYes
Software EnvironmentChecks the compatibility of TapeStation Controller Software with the computer.YesYes

Table 18 RNA - Possible causes of incorrect quantification in order of probability.

Root causeExplanationSolution
Insufficient MixingBuffer and sample was not sufficiently vortexed. The dye could not incorporate completely into the sample.See “Mixing Recommendations” on page 100 and the Agilent Information Center for more information.
Sample concentration outside recommended range for applicationLinear concentration estimation cannot be achieved for samples exceeding the limits in the specification table.Either dilute or concentrate the sample until it is within the recommended range for the application as stated in the respective assay Quick Guide; then prepare a new run. Use the High Sensitivity assay if applicable (“Quantitative range” on page 114, Table 3 on page 68).
Incorrect pipetting technique or pipette calibrationIncorrect pipetting technique or uncalibrated pipettes can lead to poor pipetting performance which in turn alter sample/buffer ratio in an unpredictable manner.See “Good Pipetting Practice” on page 91 and follow guidelines of the respective pipette vendor.
Sample not collected at bottom of vialInsufficient volume pipetted by the robotics and applied to the lane may affect the migration behavior and alter the band shape.Make sure to collect all liquid at the bottom of the vial by centrifugation. See “Centrifugation Recommendations” on page 103. Ensure that after removing strip caps and inserting the tube strips or 96-well sample plate into the TapeStation instrument liquid is still located at the bottom of the well, and is not spilled.

Table 3 ScreenTape Assay portfolio

ScreenTape AssayQuick Guide Number
Agilent D1000 ScreenTape AssayG2991-90031
Agilent High Sensitivity D1000 ScreenTape AssayG2991-90131
Agilent RNA ScreenTape AssayG2991-90021
Agilent High Sensitivity RNA ScreenTape AssayG2991-90121
Agilent Genomic DNA ScreenTape AssayG2991-90041
Agilent D5000 ScreenTape AssayG2991-90051
Agilent High Sensitivity D5000 ScreenTape AssayG2991-90151
Agilent Cell-free DNA ScreenTape AssayG2991-90060

Table 5 Test suite groups

Test suite groupsDescription
Computer EnvironmentEvaluation of computer specification and installation status.
SetupPrepares the instrument, requests insertion of material, checks status LED functionality.
Information on current Maintenance counters, past Electrode Cartridges, past needle changes and current installed software is reported.
VisionCamera and LED illumination are checked using the TapeStation Test Tape.
ElectrophoresisThe voltage supply, current readout, and electrodes are checked using the TapeStation Test Tape.
RoboticsRobotics parts are checked by dedicated movements. Tip and ScreenTape device handling are checked using a full tip rack and one regular ScreenTape device.
PipettingThis check procedure in the Hardware Troubleshooting visually checks pickup of liquid from 96-well sample plates and tube strips.
CleanupResets the instrument to normal operational mode and requests removal of material.

Table 10 Robotics

System Test SuiteTroubleshooting Test Suite
Needle To PumpYesYes
Tape HandlingYesYes

Tape Handling, Teach Pump-ML and Teach Pump-ML-Z represent tests which are only relevant for 4200 TapeStation instruments G2991A and G2991B.

Table 13 DNA - Possible causes of incorrect quantification results in order of probability

Root causeExplanationSolution
Insufficient MixingBuffer and sample was not sufficiently vortexed. The dye could not incorporate completely into the sample.See “Mixing Recommendations” on page 100 and the Agilent Information Center for more information.
Sample concentration outside recommended range for application.Linear concentration estimation cannot be achieved for samples exceeding the limits in the specification table.Either dilute or concentrate your sample until it is within the recommended range for application as stated in the respective assay Quick Guide, then prepare a new run. Use the High Sensitivity assay if applicable (Table 3 on page 68).
Incorrect peak integrationThe algorithm cannot establish a correct concentration estimation if marker or sample peaks are not completely encompassed as a signal from either marker or sample is left out of the calculation.Ensure that all sample peaks and markers are integrated correctly in the TapeStation Analysis software, by clicking and dragging so that the whole peak is encompassed. See the Agilent Information Center for more information.
Incorrect analysis typeThe Gel and Electropherogram view are designed to analyze discrete peaks within samples. The Region view is optimized for analyzing samples with broad peaks. Region view uses a different method to calculate concentration and sizing.Use Gel view for samples with narrow peaks and the Region view for samples with broad peaks or smears such as libraries. See “Peak Maxima Versus Average Molecular Weight Sizing and Molarity” on page 111 for more information.
Sample and the marker relevant for quantification have merged.Additional sample overlaying with the marker will alter the known concentration of the marker that is used for concentration calculation.Ensure that your sample is within the recommended sizing range for the applications, see “Sizing Range” on page 113.
Contaminating beads can alter the run behavior of the sample and can introduce artifacts, see “TapeStation Analysis in the Agilent SureSelect and HaloPlex Workflow” on page 226.If using the Sure Select Protocol, ensure that all AMPure beads are removed by increasing the time that the sample is on the magnetic plate.

Table 13 DNA - Possible causes of incorrect quantification results in order of probability

Root causeExplanationSolution
Sample not collected at bottom of vialInsufficient volume pipetted by the robotics and applied to the lane may affect migration behavior and alter the band shape.Make sure to collect all liquid at the bottom of the vial by centrifugation. See “Centrifugation Recommendations” on page 103. Ensure that after removing strip caps and inserting the tube strips or 96-well sample plate into the TapeStation instrument, liquid is still located at the bottom of the well, and is not spilled.
Incorrect protocol usedDifferences exist between Standard and High Sensitivity assay protocols.Ensure that the correct sample protocol for the application is used, see “Use the Correct Protocol” on page 99.
Incorrect reagents usedReagents from different assays may differ in dye and marker concentration. Furthermore markers used are different and upper markers may not be present.Ensure that the correct reagent buffer for the application is used, see “Use the Correct Protocol” on page 99.
Sample prepared too long before analysisEvaporation or settling may alter the sample to buffer ratio.Ensure that your run is started immediately after sample preparation.
The sample/buffer mix may evaporate over time if a 96-well sample plate was used which was not sealed with the recommended seal foil.Always cover with the recommended foil seal (PN 5067-5154) and ensure tight sealing throughout the 96-well sample plate.
Incorrect pipetting technique or pipette calibration.Incorrect pipetting technique or uncalibrated pipettes can lead to poor pipetting performance which in turn alter sample/buffer ratio in an unpredictable manner.See “Good Pipetting Practice” on page 91 for more information or follow guidelines of the respective pipette vendor.
Incorrect Marker peaks picked upDepending on the sample pattern, wrong peaks can get assigned as marker if running very close to the actual marker.Use the Aligned button to detect the correct markers. Compare to aligned/unaligned data of good, for example the ladder, and questionable samples, then assign correct markers by right-clicking. See the Agilent Information Center for more information. Ensure that all sample peaks are within the recommended sizing range for the application.

Table 14 DNA - Possible causes of incorrect sizing results in order of probability

Root causeExplanationSolution
Incorrect Marker peaks picked upDepending on the sample that is used, wrong peaks can get assigned as marker if the sample size is very close to the actual marker.Use the Aligned button in the TapeStation Analysis software to detect the correct markers. Compare to aligned/unaligned data of good and questionable samples; then assign correct markers by right-clicking. See the Agilent Information Center for more information. Ensure that all sample peaks are within the recommended sizing range for the application.
Incorrect analysis typeThe Gel and Electropherogram view are designed to analyze discrete peaks within samples. The Region view is optimized for analyzing samples with broad peaks. Region view uses a different method to calculate concentration and sizing.Use Gel view for samples with narrow peaks and the Region view for samples with broad peaks or smears such as libraries. See “Peak Maxima Versus Average Molecular Weight Sizing and Molarity” on page 111 for more information.
Issues with the ladder laneAny issues with the run ladder lane will be annotated within the TapeStation Analysis software sample table under Observations.Manually insert or delete affected peaks. In assays where an electronic ladder is available (D1000, High Sensitivity D1000, and Cell-free DNA Assay), insert the electronic ladder to regain sizing information, see the Agilent Information Center for more information. In assays where an electronic ladder is not available (D5000, High Sensitivity D5000, and Genomic DNA Assay), re-run affected samples and ensure that Good Measurement Practices are followed, see “Good Measurement Practices” on page 87. For Genomic DNA Ladder considerations, see “Genomic DNA Assay” on page 210.
Insufficient migration due to presence of saltSample matrices exceeding the maximum buffer strength specifications may alter currents in an unpredictable manner leading to short running lanes, see “Maximum Buffer Concentration in Sample” on page 116.Dilute your samples to ensure low levels of buffer salt. True migration profiles can be seen by unaligning your gel image by pressing the Aligned button in the TapeStation Analysis software. Please refer to the salt tolerance guidelines for your assay in the appropriate Quick Guide (Table 3 on page 68).

Engineering Facts

13 facts · every claim linked to a page
Selection

The G2991A model requires a trained service engineer for Electrode Cartridge exchange, while G2991B and G2992A allow user exchange, which affects maintenance planning and model choice.

Determines whether on-site user maintenance is possible or a service contract is needed.

Source
“Only G2991B and G2992A posses an Electrode Cartridge with lever mounting suitable for the exchange by users during maintenance. For model G2991A the Electrode Cartridge can be exchanged only by a trained service engineer during a scheduled preventive maintenance service.”
Installation

The instrument requires 75 mm clear space around it and 250 mm additional height above the listed dimensions to open the lid, so bench layout must account for this.

Prevents installation in spaces that block airflow, connections, or lid operation.

Source
“Allow approximately 75 mm (3 inches) of clear space around the instrument for air circulation and electrical connections.”
Installation

Two plug sockets must be available near the bench for the instrument and a laptop, which must be planned in the site requirements.

Ensures adequate power outlets are available before installation.

Source
“Two plug sockets must be available near the allocated bench space for the instrument itself and a laptop.”
Installation

If shipment conditions are unknown or outside operating conditions, wait at least 5 hours before unpacking to allow temperature and humidity equilibration and avoid condensation damage.

Prevents condensation-induced electronics damage during commissioning.

Source
“If actual shipment conditions are unknown or were outside operating conditions, wait for at least 5 h before unpacking to allow the instrument to adjust to the ambient temperature and humidity of the operating environment.”
Installation

The instrument is heavy and requires a second person for lifting during unpacking to prevent injury.

Ensures safe handling and prevents back strain or dropping during installation.

Source
“Because the instrument is heavy, lifting it will require the assistance of a second person to prevent injury.”
Operation

The ladder must always be loaded in location A1 on the tube strip holder for analysis.

Incorrect ladder placement will cause run failure or invalid results.

Source
“Always load the ladder in location A1 on the tube strip holder.”
Operation

Running samples from the 96-well plate and tube strips in one analysis is not supported; all samples must be combined on one format.

Prevents run setup errors that would require re-running samples.

Source
“Running samples from the 96-well sample plate and from sample tube strips in one analysis is not supported. Combine all samples you want to run on one instrument either to the 96-well sample plate or to the tube strips.”
Performance

Insufficient mixing degrades results; the recommended protocol is vortex mixing at 2000 rpm for 1 minute followed by brief centrifugation.

Ensures reproducible electropherogram and gel results by following validated mixing conditions.

Source
“The green trace shows the recommended protocol of vortex mixing using the IKA vortexer and adaptor at 2000 rpm for 1 minute followed by brief centrifugation.”
Performance

Electropherogram view reports the highest point of a peak as size, while region view reports the center of the region's mass, so sizing interpretation depends on the selected view.

Prevents misinterpretation of sizing and molarity results depending on analysis mode.

Source
“Electropherogram view is designed for use with discrete peaks, and the default size reported is that of the highest point of the peak.”
Maintenance

After an Electrode Cartridge change, a System Validation test is recommended for good laboratory practice even though it is not mandatory to continue working.

Guides maintenance workflow to confirm instrument performance after part replacement.

Source
“This additional system validation is not mandatory to continue working with the instrument but is recommended for good laboratory practice reasons.”
Safety

The instrument is not completely disconnected from mains when the power switch is off, so the power cable must be unplugged for full isolation.

Ensures safe servicing by fully removing power before maintenance.

Source
“The apparatus is not completely disconnected from the mains supply when the power switch is in the Off position”
Safety

Bearers of pacemakers or implanted defibrillators must stay at least 55 mm from the magnets, and magnets must be kept 25 mm from sensitive devices.

Prevents serious injury to personnel with implanted medical devices during installation and operation.

Source
“Bearers of heart pacemakers or implanted defibrillators must stay off at least 55 mm from the magnets.”
Conformité

The product is ISM Group 1 Class A and may cause radio interference in domestic environments, requiring adequate measures by the user.

Informs site selection and EMC compliance planning for the laboratory environment.

Source
“This is an ISM Group 1 Class A product intended for use in industrial environment. In a domestic environment, this product may cause radio interference, in which case the user may be required to take adequate measures.”