Selection
The kit only supports DNA sizing from 100 bp to 6,000 bp, so samples outside this range cannot be sized with this assay.
Determines whether the kit is suitable for the expected fragment size range before purchase.
Source
“DNA sizing range 100 bp – 6,000 bp”
Performance
Separation resolution is 8 bp at 500 bp with the 33-55 array, which sets the minimum detectable size difference near 500 bp.
Needed to judge if closely sized fragments can be resolved for the application.
Source
“8 bp @ 500 bp (33-55 Array)”
Performance
Input DNA must be between 0.005 ng/μL and 2 ng/μL, adjustable by dilution, so samples must be quantified and diluted before loading.
Prevents overloading or underloading the capillary array and ensures valid sizing.
Source
“0.005 ng/μL – 2 ng/μL input DNA (adjustable by dilution of sample)”
Installation
The gel/dye mixture volume scales with sample count (e.g., 25 mL gel with 2.5 µL dye for 48 samples), so reagent volumes must be planned per run.
Allows correct preparation of fresh gel/dye mixture for the planned number of samples.
Source
“2.5 µL 25 mL”
Maintenance
The 1x 930 dsDNA Inlet Buffer must be replaced daily, which imposes a daily maintenance step on the instrument.
Affects daily operating schedule and buffer consumption planning.
Source
“Replace 1x 930 dsDNA Inlet Buffer daily.”
Maintenance
Capillary Storage Solution must be replaced every 2-4 weeks, setting a recurring maintenance interval.
Defines the long-term capillary storage maintenance cadence.
Source
“Replace Capillary Storage Solution every 2-4 weeks.”
Operation
The MDK DNA Ladder is ready to use without dilution and must be placed at 24 µL into well D12 or H12.
Ensures the ladder is loaded correctly for sizing calibration.
Source
“Place 24 µL of MDK DNA Ladder (“ready to use”; no dilution) into well D12 or H12.”
Operation
The correct software method to select is CRP-910-(33 or 55) – CRISPR Discovery, matching the array type.
Using the wrong method would produce invalid or failed separations.
Source
“select the CRP-910-(33 or 55) – CRISPR Discovery method from the dropdown menu.”
Safety
Reagent handling carries health risks, so safety goggles, gloves, and protective clothing must be worn.
Mandatory PPE requirement for anyone handling kit reagents.
Source
“Follow the appropriate safety procedures such as wearing goggles, safety gloves and protective clothing”
Conformité
The kit is for research use only and must not be used in diagnostic procedures.
Restricts the intended use and prevents clinical/diagnostic application.
Source
“For Research Use Only. Not for use in diagnostic procedures.”