Agilent 5975 Series Mass Selective Detector (MSD) Operating Manual
Sommaire
Overviewpp. 9-28
Overview of 5975 Series MSD, CI hardware, safety warnings, hydrogen safety, symbols, recycling
Procedurepp. 29-48
Preparing, installing, and conditioning capillary columns in inlets and GC/MSD interface
Procedurepp. 49-96
Operating MSD in electron impact mode, pumping down, venting, temperature settings, tuning
Procedurepp. 97-126
Chemical ionization operation, reagent gas setup, switching sources, pressure readings
Procedurepp. 127-144
Maintenance tools, spare parts, electrostatic discharge, vacuum system, CI maintenance
Overviewpp. 145-160
Theoretical background on chemical ionization, positive and negative CI, ion-molecule reactions
p. 30table, 5 rows
p. 14table, 8 rows
p. 10table, 6 rows
p. 55table, 7 rows
p. 56table, 2 rows
p. 112table, 6 rows
p. 125table, 7 rows
p. 57table, 7 rows
p. 72table, 8 rows
p. 128table, 12 rows
Table 5 Gas flows
Feature
G3170A
G3175A
G3171A
G3176A
G3172A
G3174A
High vacuum pump
Diffusion
Standard turbo
Performance turbo
Performance turbo
Optimal gas flow, mL/min*
1
1
1 to 2
1 to 2
Maximum recommended gas flow, mL/min
1.5
2
4
4
Maximum gas flow, mL/min†
2
2.4
6.5
6.5
Maximum column id
0.25 mm (30 m)
0.32 mm (30 m)
0.53 mm (30 m)
0.53 mm (30 m)
* Total gas flow into the MSD = column flow + reagent gas flow (if applicable). † Expect degradation of spectral performance and sensitivity.
Table 3 5975 series MSD models and features
Model
G3170A
G3175A
G3171A
G3176A
G3172A
G3174A
Feature
High vacuum pump
Diffusion
Standard turbo
Performance turbo
Performance turbo
Optimal He column flow mL/min
1
1
1 to 2
1 to 2
Maximum recommended gas flow mL/min*
1.5
2.0
4.0
4
Maximum gas flow, mL/min†
2
2.4
6.5
6.5
Max column id
0.25 mm (30 m)
0.32 mm (30 m)
0.53 mm (30 m)
0.53 mm (30 m)
CI capability
No
No
No
Yes
DIP‡ capability (3rd party)
Yes
Yes
Yes
Yes
* Total gas flow into the MSD: column flow plus reagent gas flow (if applicable). † Expect degradation of spectral performance and sensitivity. ‡ Direct insertion probe.
Table 1 Available high vacuum pumps
Model name
Product number
Description
Ionization modes
5975C TAD VL MSD
G3170A
Diffusion Pump MSD
Electron impact (EI)
5975C TAD inert MSD
G3171A
Standard Turbo MSD
Electron impact (EI)
5975C TAD inert XL MSD
G3172A
Performance Turbo MSD
Electron impact (EI)
5975C TAD inert XL MSD
G3174A
CI High Mass Performance Turbo Pump
Electron impact (EI) Negative chemical ionization (NCI) Positive chemical ionization (PCI)
7820 MSD VL
G3175A
Diffusion Pump MSD
Electron impact (EI)
7820 MSD
G3176A
Standard Turbo MSD
Electron impact (EI)
Table 6 ChemStation menu
Action
Description
Run Method
Displays the current method name and starts an analysis.
Run Sequence
Displays the current sequence and starts a sequence.
Run Current Tune
Displays the current tune file and starts an autotune (EI mode only; CI tune must be started from the GC/MSD ChemStation).
# of Messages
Displays the number of messages and the text of the most recent message. Use the arrow keys to scroll through previous messages (up to 20).
Release ChemStation
Disassociates the GC/MSD ChemStation from the MSD.
Connection Status
Displays the LAN connection status for the MSD. Remote = connected to GC/MSD ChemStation online session Local = not connected to GC/MSD ChemStation online session
Name of Instrument
Displays the name of the instrument if connected to GC/MSD ChemStation online session. The name of the instrument is the name assigned to the MSD by the GC/MSD ChemStation Configuration dialogue.
Table 7 Maintenance menu
Action
Description
Prepare to vent
Reminds you to shut down the GC then prepares the instrument for venting when [Yes/Select] is pressed.
* There are no PFDTD ions formed with any reagent gas but methane. Tune with methane and use the same parameters for the other gas. † There are no negative reagent gas ions formed. To pretune in negative mode, use background ions: 17 (OH-), 35 (Cl-), and 235 (ReO3-). These ions can not be used for reagent gas flow adjustment. Set flow to 40% for NCI and adjust as necessary to get acceptable results for your application.
Displays the GC/MSD ChemStation operating system version.
Front panel
Displays the version of the LCP.
Log amplifier
Displays version information.
Sideboard
Displays the sideboard type.
Mainboard
Displays the mainboard type.
Serial number
Is assigned to the MSD by GC/MSD ChemStation Configuration dialogue.
Table 13 Micro-Ion Vacuum Gauge Reading
Column flow rate, mL/min
Gauge reading, Torr Performance turbo pump
Gauge reading, Torr Standard turbo pump
Gauge reading, Torr Diffusion pump
Foreline reading, Torr Diffusion pump
0.5
3.18E–06
1.3E–05
2.18E–05
34.7
0.7
4.42E–06
1.83E–05
2.59E–05
39.4
1
6.26E–06
2.61E–05
3.66E–05
52.86
1.2
7.33E–06
3.11E–05
4.46E–05
60.866
2
1.24E–05
5.25E–05
7.33E–05
91.784
3
1.86E–05
8.01E–05
1.13E–04
125.76
4
2.48E–05
6
3.75E–05
Table 21 Maintenance schedule
Task
Every week
Every 6 months
Every year
As needed
Tune the MSD
X
Check the foreline pump oil level
X
Check the calibration vial(s)
X
Replace the foreline pump oil*
X
Replace the diffusion pump fluid
X
Check the dry foreline pump
X
Clean the ion source
X
Check the carrier gas trap(s) on the GC and MSD
X
Replace the worn out parts
X
Lubricate sideplate or vent valve O-rings†
X
Replace CI Reagent gas supply
X
Replace GC gas supplies
X
* Every 3 months for CI MSDs using ammonia reagent gas. † Vacuum seals other than the side plate O-ring and vent valve O-ring do not need to be lubricated. Lubricating other seals can interfere with their correct function.
Engineering Facts
14 facts · every claim linked to a page
Selection
The MSD model determines the maximum column id and gas flow, so column selection must match the pump configuration.
Choosing a column with too large an id for the installed pump can degrade performance and exceed flow limits.
Source
“Max column id 0.25 mm (30 m) 0.32 mm (30 m) 0.53 mm (30 m) 0.53 mm (30 m)”
Selection
CI capability is only available on the G3174A model, so CI requirements dictate the model choice.
If CI operation is required, only the G3174A supports it; other models cannot be upgraded by column choice alone.
Source
“CI capability No No No Yes”
Installation
The column installation procedure is specific to the 5975 Series and using another instrument's procedure may damage the column or MSD.
Following the wrong procedure can cause poor sensitivity or physical damage during installation.
Source
“The column installation procedures for the 5975 Series MSDs is different from that for previous MSDs. Using the procedure from another instrument may not work and may damage the column or the MSD.”
Installation
Carrier gas purity must be at least 99.9995% to minimize column bleed and contamination.
Insufficient gas purity increases column bleed and ion source contamination, affecting sensitivity.
Source
“Always use carrier gas that is at least 99.9995% pure.”
Installation
New ferrules may loosen after heating and cooling cycles, so tightness must be checked after two or three cycles.
Loose ferrules cause leaks that degrade vacuum and performance after installation.
Source
“Because of thermal expansion, new ferrules may loosen after heating and cooling a few times. Check for tightness after two or three heating cycles.”
Safety
With hydrogen carrier gas, carrier flow must not start until the column is installed and the MSD is pumped down.
Starting hydrogen flow before pumpdown can accumulate explosive gas in the MSD.
Source
“If you are using hydrogen as a carrier gas, do not start carrier gas flow until the column is installed in the MSD and the MSD has been pumped down.”
Operation
The quadrupole must not exceed 200 °C and the source must not exceed 350 °C.
Exceeding these limits can damage the analyzer components during temperature setup.
Source
“Do not exceed 200 °C for the quadrupole or 350 °C for the source.”
Operation
Analyzer heater zones interact, so setpoints far from adjacent zones may prevent accurate temperature control.
Setting widely different zone temperatures can cause control instability and affect analysis reproducibility.
Source
“The GC/MSD interface, ion source, and quadrupole heated zones interact. The analyzer heaters may not be able to accurately control temperatures if the setpoint for one zone is much different from that of an adjacent zone.”
Operation
After switching from EI to CI or venting, the MSD must bake out for at least 2 hours before tuning.
Skipping bakeout leaves water and contaminants that degrade CI tuning and sensitivity.
Source
“After the system has been switched from EI to CI mode, or vented for any other reason, the MSD must be baked out for at least 2 hours before tuning.”
Operation
A visible peak at m/z 32 during methane pretune indicates an air leak that must be repaired before proceeding.
Operating CI with an air leak rapidly contaminates the ion source.
Source
“Make sure there is no visible peak at m/z 32. A peak there indicates an air leak. If such a peak is present, find and repair the leak before proceeding.”
Maintenance
Column bleed deposits stationary phase in the ion source, decreasing sensitivity and requiring source cleaning.
Understanding bleed helps plan maintenance intervals and column conditioning.
Source
“Column bleed deposits traces of the stationary phase in the MSD ion source. This decreases MSD sensitivity and makes cleaning the ion source necessary.”
Safety
The MSD cannot detect leaks in inlet or detector gas streams, so column fittings must always be connected or capped.
Uncapped fittings allow hydrogen or air to enter, creating explosion or contamination hazards.
Source
“The MSD cannot detect leaks in inlet and/or detector gas streams. For this reason, it is vital that column fittings should always be either connected to a column or have a cap or plug installed.”
Safety
Combustible materials under or around the foreline pump are a fire hazard and the oil pan must be kept clean.
Proper housekeeping around the pump prevents fire during operation.
Source
“Combustible materials (or flammable/non-flammable wicking material) placed under, over, or around the foreline (roughing) pump constitutes a fire hazard. Keep the pan clean, but do not leave absorbent material such as paper towels in it.”
Operation
If the foreline pump gurgling noise does not stop within a minute during pumpdown, there is a large air leak.
Persistent gurgling signals a leak that must be fixed before reliable operation.
Source
“The foreline pump will make a gurgling noise. This noise should stop within a minute. If the noise continues, there is a large air leak in your system, probably at the side plate seal, the interface column nut, or the vent valve.”