Agilent 7100 Capillary Electrophoresis System User Manual

Sommaire

  1. Overview pp. 13-18
    System overview, learning products, and documentation scope
  2. Spec pp. 19-66
    Power cords, bench space, dimensions, weight, and performance specifications in multiple languages
  3. Procedure pp. 67-90
    Initial setup, initialization mode selection, and manual configuration
  4. Procedure pp. 91-108
    Graphical user interface and direct control of instrument components
  5. Procedure pp. 109-156
    CE operation, high voltage system, preconditioning, detector parameters, and fraction collection
  6. Procedure pp. 157-168
    Operating the instrument in high pressure CE+p mode
  7. Procedure pp. 169-180
    CEC operation, preconditioning, and control via CE diagram
  8. Procedure pp. 181-210
    Diagnostics, firmware, diode array detector, and fuse replacement
  9. Procedure pp. 211-248
    Maintenance overview, EMF, pre-puncher cleaning, lamp replacement, and replenishment system care
  10. Ordering pp. 249-272
    Overview of maintenance parts and ordering information
  11. Warning pp. 273-282
    Safety symbols, WEEE directive, and legal notices
  12. pp. 283-286
    Alphabetical index and document part number
  13. p. 23 table, 3 rows
  14. p. 24 table, 4 rows
  15. p. 87 table, 6 rows
  16. p. 22 table, 9 rows
  17. p. 83 table, 5 rows
  18. p. 21 table, 12 rows
  19. p. 78 table, 2 rows
  20. p. 82 table, 3 rows
  21. p. 79 table, 4 rows
  22. p. 70 table, 27 rows

Table 3 Performance specifications for the Agilent 7100 Capillary Electrophoresis System

TypeSpecificationComments
Raw data channelsDetector signals and spectrum, voltage, current, leak current, power, cassette temperature, pressure, lamp voltage and detector temperature
Electrophoresis power• Voltage range • Current • Power • Operation under constant voltage, current or power • Programmable polarity switchsetable 0 to ± 30 kV supply setable 0–300 µA setable 0–6 W
CE specific software• Mobility report, time corrected areas, pI calibration and bio polymer size calibration

Table 6 Performance specifications for the Agilent 7100 Capillary Electrophoresis System, Capillary Cassette

TypeSpecificationComments
Capillary cassetteHigh-speed forced-air cooler with Peltier element
Temperature range10 °C below ambient, up to 60 °C (min. 10 °C cassette temperature)
Minimum total capillary length31 cm
Capillary compatibility365 µm o.d.

Table 11 Telnet commands

ValueDescription
?Displays syntax and descriptions of commands
/Displays current LAN settings
ip <x.x.x.x>Sets new IP address
sm <x.x.x.x>Sets new subnet mask
gw <x.x.x.x>Sets new default gateway
exitExits shell and saves all changes

Table 3 Performance specifications for the Agilent 7100 Capillary Electrophoresis System

TypeSpecificationComments
Safety featuresExtensive diagnostics, error detection and display, current leak detection; low current limitLiquid leak sensor; safety sensors on door and cover disabling high voltage; vial sensor;
Control and data evaluationAgilent control software (e.g. ChemStation, OpenLAB CDS ChemStation Edition, MassHunter)
CommunicationLAN, controller-area network (CAN), RS-232C, APG remote: ready, start, stop and shut-down signals. USB (future use). Analog out and in.0 to1 V
GLP featuresEarly maintenance feedback (EMF), electronic records of maintenance and errors
Diagnostic functionsRFID tag for lamp, early maintenance feedback, supported by Agilent Lab Advisor software with integrated diagnosis test suite
Pressure system• Programmable with -100 to +100 mbar on inlet • Flushing with 1 bar or with high pressure 2–12 bar • Vial pressurization with high pressure 2–12 bar on inlet and/or outlet
Injection modes• Self correcting injection system with injection from inlet or outlet • Programmable range • Pressure • Electrokineticup to 10,000 seconds –100 to +100 mbar –30 to +30 kV
Replenishment• Satellite station for refilling of inlet or outlet buffer vials with fresh buffer for automatic, continuous operation. • Selectable buffer levelling.
Vials• 100 µL (polypropylene or glas) with resealing snap caps • 1 mL (polypropylene) with resealing snap caps • 2 mL (glas) with resealing snap caps

Table 10 Link configuration switches

SW3SW4SW5Link configuration
OFFSpeed and duplex mode determined by auto-negotiation
ONOFFOFFManually set to 10 Mbps, half-duplex
ONOFFONManually set to 10 Mbps, full-duplex
ONONOFFManually set to 100 Mbps, half-duplex
ONONONManually set to 100 Mbps, full-duplex

Table 2 Physical specifications

TypeSpecificationComments
Weight35 kg (77.2 lbs)
Dimensions (width × depth × height)350 x 510 x 590 mm (13.8 x 20.1 x 23.2 inches)
Line voltage100–240 VAC ± 10%Wide-ranging capability
Line frequency50 or 60 Hz ± 5%
Power consumption350 VA / 300 W / 1024 BTU/hMaximum
Ambient operating temperature5–40 °C (41–104 °F)
Ambient non-operating temperature-40 to 70°C (-40 to 158 °F)
Humiditybelow 80% at 31 °C (87.8 °F)Non-condensing
Operating altitudeUp to 2000 m (6500 ft)
External coolingmax. 0.5 bar (7.2 psi), max. 50 °C (122 °F)Waterbath
External pressure2-12 bar (29-174 psi)Oil-free air or nitrogen
Safety standards: IEC, CSA, ULHousingInstallation category II, Pollution degree 2 All materials are recyclable For indoor use only

Table 7

Factory default settings
Initialization ('init') modeUsing default, switches 7 and 8 set to ON.
Link configurationSpeed and duplex mode determined by auto-negotiation

Table 9

Using Default parameters
IP address:192.168.254.11
Subnet mask:255.255.255.0
Default gatewayNot specified

In the "Using Default" mode, the parameters stored in the memory of the 7100 Capillary Electrophoresis instrument are not cleared automatically. If not changed by the user, they are still available when you switch back to the mode "Using Stored".

Table 8 Initialization mode switches

SW6SW7SW8Init mode
OFFOFFOFFBootp
OFFOFFONBootp & Store
OFFONOFFUsing Stored
OFFONONUsing Default

Table 6 Accessories

DescriptionPart numberQuantity
Accessory kitG7100-687051
– HPCE startup capillary kit1600-687001
– Detector interface, green, 50 µmG7100-602101
– Detector interface, red, extended 50 µmG7100-602301
– Coupler0100-16332
– O-ring 25.12-mm-id 1.78-mm-XSECT-DIA0905-11631
– O-ring 3-mm0905-1196*5
– Container1540-03071
– Fuse T6.3AH, 250 VAC2110-10182
– Fuse T6.3AL, 250 VAC2110-06232
– Fuse T3.15AL, 250 VAC2110-14174
– Air filter3150-06191
– Tubing assembly5063-65271
– Screwdriver, slot shaft8710-26071
– Screwdriver, Pozidriv shaft8710-08991
– 12 mm socket wrench8710-20761
– Lift tubing assyG1600-670031
– Vial crimp/snap polypropylene 1 mL 100/PK5182-05671
– Snap caps, polyurethane 100/PK5181-15121
– Cable, remote5061-33781
– Bottle, 500 mL, electrolyte, pressure plus9301-65312
– Bottle, 100 mL, electrolyte, pressure plus9301-65321
– CE System User’s GuideG7100-900001
– Glass filter, solvent inlet5041-21681
– Electrode alignment assyG7100-601071
– LAN cable (crossover)5023-02031
CE Installation Qualification Kit5063-65141

*Part number 0905-1196 is not orderable. Please re-order part number 5062-8544 (5/PK)

Engineering Facts

14 facts · every claim linked to a page
Installation

The instrument requires additional clearance of 2.5 cm on each side and about 8 cm at the rear for air circulation and electrical connections.

Bench space planning must account for these clearances to ensure proper cooling and cable access.

Source
“It needs an additional 2.5 cm (1.0 inch) of space on either side and approximately 8 cm (3.1 inches) in the rear for air circulation and electric connections.”
Installation

The instrument must be operated in an upright position.

Mounting or placing the instrument in any other orientation is not permitted.

Source
“The instrument must be operated in a upright position.”
Safety

The module remains partially energized when switched off as long as the power cord is plugged in.

Service or maintenance work requires unplugging the power cord to avoid shock hazard.

Source
“The module is partially energized when switched off, as long as the power cord is plugged in.”
Safety

The power plug must remain easily accessible and the cable removable at any time for emergency disconnection.

Installation layout must not obstruct the power plug for emergency disconnection.

Source
“Make sure that it is always possible to access the power plug.”
Installation

For CEC or CE+p mode an external supply of oil-free air or nitrogen with a maximum pressure of 15 bar can be used.

Sites planning CEC or CE+p operation must provide a suitable external gas supply and adapter.

Source
“an external pressure supply of oil-free air or nitrogen with a maximum pressure of 15 bar (218 psi) can be used.”
Operation

The cassette temperature setpoint must be within the specified range, with the lowest applicable temperature being 10 °C below ambient and the highest 60 °C.

Method temperature settings must respect ambient conditions and the 60 °C upper limit.

Source
“The lowest applicable temperature is 10 °C below ambient temperature. The highest temperature is 60 °C.”
Operation

Operating the cassette below the dew point can cause condensation, arcing and damage to the cassette or instrument.

Users must keep cassette temperature above dew point or reduce ambient humidity to avoid damage.

Source
“if the temperature of the cassette is below the dew point condensation might take place.”
Operation

Outlet vials must be filled with enough liquid to touch the capillaries to prevent droplets that can cause pressure and affect injection amount.

Proper outlet vial filling is required for reproducible injections and to avoid contamination.

Source
“Outlet vials (for example, waste vials) should also be filled with enough liquid to touch the capillaries.”
Operation

For replenishment, a capped vial filled with buffer or water should be placed in position 49 to prevent buffer precipitation and needle blockage.

Following this practice avoids blockage of the replenishment needle and maintains reproducibility.

Source
“it is good practice to have a capped vial filled with buffer or water in position 49”
Operation

The tray can only be moved by the Agilent ChemStation when the tray door is closed for safety reasons.

Operators must close the tray door before commanding tray movement to avoid safety issues.

Source
“For safety reasons, the tray can only be moved by the Agilent ChemStation when the tray door is closed.”
Installation

The default ramp up rate of the 7100 Capillary Electrophoresis is 5 kV/s.

Method developers should know the default voltage ramp rate when designing separation methods.

Source
“The default ramp up rate of the 7100 Capillary electrophoresis is 5 kV/s.”
Operation

When using fraction collection, all timetable entries using voltage, outlet vial or pressure settings are deleted from the timetable during the run.

Users must set voltage and ramp time in the appropriate screens because timetable entries are overridden.

Source
“When using fraction collection all timetable entries which use voltage, outlet vial or pressure settings will be deleted from the timetable.”
Installation

The 7100 CE is shipped with switches 7 and 8 set to ON, giving a fixed default IP address of 192.168.254.11.

Network configuration requires knowing the factory default IP and switch settings.

Source
“The 7100 CE is shipped with switches 7 and 8 set to ON. This means that the instrument is set to a fixed default IP: 192.168.254.11”
Installation

To perform any LAN configuration, switches SW1 and SW2 must be set to OFF.

LAN setup cannot proceed unless these switches are correctly configured.

Source
“To perform any LAN configuration, SW1 and SW2 must be set to OFF.”