ICP-OES User's Guide

Sommaire

  1. Overview p. 3
    Table of contents listing document sections
  2. Spec p. 4
    Power supply requirements and specifications
  3. Procedure p. 5
    Procedure for running samples on the instrument
  4. Warning pp. 6-13
    Safety warnings, hazards, and safe state verification
  5. p. 14
    Color coding conventions used in the document
  6. Warning pp. 15-16
    EMC declarations including South Korean Class A
  7. Overview p. 17
    Introduction to the instrument and system
  8. Spec p. 18
    Technical specifications of the instrument
  9. Spec p. 19
    Electrical power supply details
  10. Spec pp. 20-21
    Water chiller system requirements and specifications
  11. Spec p. 22
    ICP-OES instrument cooling air supply requirements
  12. Procedure pp. 23-27
    Installation procedures including USB-GPIB-HS converter and hardware
  13. Procedure pp. 28-36
    Turning on, preparing for analysis, setting conditions, running samples
  14. Procedure pp. 37-41
    Routine maintenance, induction coil maintenance, cleaning
  15. Ordering p. 42
    List of consumables and part numbers
  16. p. 20 table, 6 rows
  17. p. 7 table, 12 rows
  18. p. 27 table, 11 rows
  19. p. 37 table, 6 rows

Table 1 Gas requirements

ArgonNitrogen
Purity99.996%99.996%
Oxygen<5 ppm<5 ppm
Nitrogen (argon only)<20 ppm-
Water vapor<4 ppm<4 ppm
Permissible pressure range400–600 kPa (57 to 88 psi)
Recommended pressure550 kPa (80 psi) regulated

Safety Practices and Hazards

РазделСтраница
General7
Verifying Safe State8
Plasma8
Heat, Vapors and Fumes9
Compressed Gas Hazards9
Electrical Hazards10
Other Precautions10
Warning Symbols11
Color Coding14
CE Compliance14
Electromagnetic Compatibility15
South Korean Class A EMC declaration16

Operation

РазделСтраница
Analysis Checklist27
Turning On the Instrument and Software28
Preparing for Analysis30
Calibrating the Detector30
Aligning the Torch30
Setting Operating Conditions31
Creating/Opening a Worksheet31
Developing a Method31
Running Samples32
Printing a Report32
Turning Off the Instrument32

Table

РазделСтраница
Routine Maintenance38
Induction Coil Maintenance39
Cleaning41
Consumables42
Troubleshooting42
Spare Parts42

Engineering Facts

14 facts · every claim linked to a page
Safety

The plasma must never be operated unless the torch compartment door is closed with the locking lever fully latched and the space above the chimney is clear.

Interlock and clearance conditions are mandatory for safe plasma ignition and must be verified before operation.

Source
“The plasma must never be operated unless: the torch compartment door is closed, with the locking lever fully latched; and the space above the chimney is clear of objects.”
Safety

The torch and its surroundings remain hot for up to five minutes after plasma extinction, so cooling or heat-resistant gloves are required before work in that area.

Prevents burn injuries during maintenance or torch access immediately after plasma shutdown.

Source
“The torch and its surroundings remain hot for up to five minutes after the plasma is extinguished.”
Installation

The exhaust fan must always be switched on before igniting the plasma to extract heat, ozone, vapors and fumes.

Exhaust must be operational before plasma ignition to avoid hazardous fume exposure.

Source
“The exhaust fan must always be switched on before igniting the plasma.”
Installation

Only instrument-grade argon (99.996% purity) with specified impurity limits is acceptable for the plasma, nebulizer and optics interface purge.

Gas purity directly affects plasma stability, analytical performance and instrument safety.

Source
“Use only ‘instrument grade’ gases with your spectrometer.”
Installation

The spectrometer weighs approximately 203 kg and must be moved with a forklift or other suitable mechanical lifting device.

Site planning and rigging must account for the heavy instrument to avoid injury or damage during installation.

Source
“The spectrometer weighs approximately 203 kg (448 lb).”
Performance

For optimum analytical performance, laboratory ambient temperature should be 20–25 °C and held constant within ±2 °C throughout the working day.

Stable temperature control is required for reproducible optical emission measurements.

Source
“For optimum analytical performance, it is recommended that the ambient temperature of the laboratory be between 20 and 25 °C (68 and 77 °F).and be held constant to within ±2 °C (±3.6 °F) throughout the entire working day.”
Installation

Use of power boards or extension cables is not recommended for the instrument electrical supply.

Direct mains connection avoids voltage drop and electrical safety issues during installation.

Source
“Use of power boards or extension cables is not recommended.”
Installation

The water flow sensor stops plasma and camera Peltier operation if cooling water flow drops below 1.1 L/min.

Cooling water flow must be maintained above the trip threshold to prevent instrument shutdown and damage.

Source
“The instrument is equipped with a water flow sensor, which will stop operation of the plasma and camera Peltier assembly if the cooling water flow through the instrument drops below 1.1 L/min (0.3 gpm).”
Installation

The water cooling system must always be on before igniting the plasma.

Cooling must be established before plasma ignition to protect the torch and instrument.

Source
“Always ensure the water cooling system is on before igniting the plasma.”
Installation

A chemically inert drain vessel of at least 2 liters, not glass or narrow-necked, must be provided by the user and located under the sample compartment or on the right side of the instrument.

Proper waste drainage capacity and material are required for safe continuous operation.

Source
“A chemically inert container, not glass or of a narrow-necked style, to hold a minimum of 2 liters (4 pints) of waste must be provided by the user.”
Installation

The Agilent 700 Series ICP-OES must be installed by an Agilent-trained, Agilent-qualified or Agilent-authorized field service engineer.

Installation by authorized personnel is a prerequisite for warranty and safe operation.

Source
“The Agilent 700 Series ICP-OES must be installed by an Agilent-trained, Agilent-qualified or Agilent-authorized field service engineer.”
Operation

Both the RF supply circuit breaker and the instrument power switch must be switched on for the instrument to work.

Operators must verify both power points during startup to avoid false no-power diagnosis.

Source
“Both the RF supply circuit breaker and instrument power switch must be switched on for the instrument to work.”
Operation

The yellow Plasma emergency off button is intended only for emergencies such as a melting torch, not for routine plasma extinction.

Routine use of the emergency off button requires manual reset and is not the correct shutdown method.

Source
“The yellow Plasma emergency off button, located next to the mains power switch on the front of the instrument, is designed to be used only in an emergency—for example if the torch is melting.”
Operation

Gas supplies should be left on at all times except during long periods of non-use to maintain uninterrupted purge of the polychromator assembly.

Continuous purge minimizes moisture ingress and preserves optical performance.

Source
“It is strongly recommended that gas supplies be left on at all times except during long periods of non-use.”