Improved Determination of PCBs by EPA Method 1628

Sommaire

  1. Overview p. 2
    Background and purpose of the improved PCB determination method
  2. Procedure pp. 3-4
    Experimental results and discussion of method performance
  3. p. 5
    Summary of findings and method benefits
  4. p. 2 table, 7 rows

Table 3. Agilent 5977C GC/MSD selected ion monitoring (SIM) parameters for EPA Method 1628.

Start Time (min)Dwell Time (ms)SIM Ions (m/z)
Group 14.25320188.00, 190.00, 200.00, 202.00, 222.00, 224.00, 234.00, 236.00
Group 25.60300222.00, 224.00, 234.00, 236.00, 256.00, 258.00, 268.00, 270.00, 290.00, 291.80, 302.00, 304.00
Group 38.00300255.90, 258.00, 267.90, 270.00, 290.00, 291.80, 302.00, 304.00, 325.80, 328.00, 338.00, 340.00
Group 48.80300290.00, 291.80, 302.00, 304.00, 325.80, 328.00, 338.00, 340.00, 359.00, 362.00, 372.00, 374.00
Group 510.45300325.80, 328.00, 338.00, 340.00, 359.80, 362.00, 372.00, 374.00, 393.60, 396.00, 406.00, 408.00
Group 612.00300359.80, 362.00, 372.00, 374.00, 393.80, 396.00, 406.00, 408.00, 428.00, 429.70, 440.00, 442.00, 462.00, 463.60, 474.00, 476.00
Group 713.70400497.00, 500.00, 509.70, 512.00

Engineering Facts

13 facts · every claim linked to a page
Selection

The DB-XLB column is specified as better suited for PCB congener separation than the method-recommended column, so it should be selected for this application.

Column choice directly determines whether coeluting congeners can be resolved and whether the run can be shortened.

Source
“an Agilent J&W DB-XLB column—which is better suited for PCB congener separation”
Selection

Hydrogen carrier gas is viable here because it has a higher optimal linear velocity than helium, enabling rapid analysis.

Carrier gas selection affects achievable run time and requires confirming the instrument/source supports hydrogen.

Source
“Hydrogen, which has a higher optimal linear velocity than helium, allows for this rapid analysis.”
Selection

The HydroInert ion source is required to make hydrogen carrier gas work with the 5977C MSD.

Choosing hydrogen carrier gas dictates the ion source configuration at purchase/installation.

Source
“an 8890/5977C GC/MSD, equipped with a HydroInert ion source, was operated in selected ion monitoring (SIM) mode”
Installation

A specific low-fritted 4 mm Ultra Inert liner (p/n 5190-5112) is used in the multimode inlet.

Correct liner part number must be installed to match the validated method.

Source
“Agilent Ultra Inert, low‑fritted liner, 4 mm (p/n 5190-5112)”
Installation

The analytical column is a specific part number, 20 m × 180 µm × 0.18 µm DB-XLB.

Column dimensions and part number must match to reproduce separation and run time.

Source
“Agilent J&W DB-XLB, 20 m × 180 µm, 0.18 µm (p/n 121-1222)”
Operation

The method uses a multimode inlet in constant flow mode at 0.9 mL/min hydrogen.

Carrier flow setpoint is a core operating parameter that must be programmed correctly.

Source
“Flow 0.9 mL/min”
Operation

The oven program is a fast ramp: 50 °C hold 0.4 min, 30 °C/min to 180 °C, 10 °C/min to 300 °C with a 1.257 min hold.

The temperature program must be entered exactly to achieve the sub-20-minute separation.

Source
“50 °C (hold for 0.4 min), 30 °C/min to 180 °C, 10 °C/min to 300 °C (hold for 1.257 min)”
Operation

Acquisition is in SIM mode across 7 SIM groups with defined start times and dwell times.

SIM group timing must be configured to cover all target congeners within the run.

Source
“Number of SIM Groups 7”
Performance

The refined method achieves the required separation in just over 15 minutes, well under the 20-minute target.

Throughput expectation for the configured system; confirms the fast method actually delivers.

Source
“The run time for this method is just over 15 minutes, with PCB 209 and PCB 13C‑209 eluting at 15.02 minutes.”
Conformité

Method 1628 requires congeners 28 and 31 to have a valley below 80% of the smaller peak height, and this system meets it.

This is a pass/fail resolution criterion that must be verified during method validation.

Source
“closely eluting PCB congeners 28 and 31 have a valley between the two peaks that is less than 80% of the height of the smaller peak”
Conformité

Method 1628 requires PCB 118 to have S/N of 3:1 or greater at 10 ppb, and the system easily exceeds it.

Sensitivity compliance must be demonstrated at the 10 ppb detection limit.

Source
“PCB congener 118 have a signal-to-noise ratio of 3:1 or greater at a concentration of 10 ppb”
Performance

Calibration from 10 to 2000 ppb gave R2 greater than 0.97 for all calibrated congeners.

Confirms the linear range and quantitation quality achievable with this configuration.

Source
“R2 values were greater than 0.97 across the calibration set”
Operation

The ion source is operated at 280 °C and the quadrupole at 150 °C.

Source temperature setpoints must be programmed to match the validated method.

Source
“Ion Source Temperature 280 °C”