Centro de inclusión Avantik EM8, compuesto por dos módulos: consola de dispensación EM8H y consola criogénica EM8C. Accesorios enumerados: pinzas eléctricas calefactadas EQ6109, EQ6110, EQ6111.

Technical Data

Parameter Value Condition
Outside dimensions (W x H x D) 23.25" x 17" x 26" in p.2 ↗ EM8H dispensing console
Weight 62 lbs. p.2 ↗ EM8H dispensing console
Paraffin reservoir capacity 5 Liters p.2 ↗ EM8H dispensing console
Left & right tray capacity 1.7 liters p.2 ↗ EM8H dispensing console
Tray measurements (W x D x H) 7" x 10.5" x 2.25" in p.2 ↗ EM8H dispensing console
Paraffin reservoir temperature range 50 - 70 °C p.2 ↗ EM8H dispensing console
Left tray temperature 50 - 70 °C p.2 ↗ EM8H dispensing console
Right tray temperature 50 - 70 °C p.2 ↗ EM8H dispensing console
Working area temperature 50 - 70 °C p.2 ↗ EM8H dispensing console
Forceps receptacle temperature 50 - 70 °C p.2 ↗ EM8H dispensing console
Power supply 100-240V / 50-60Hz V / Hz p.2 ↗ EM8H dispensing console
Outside dimensions (W x H x D) 13" x 16" x 23.75" in p.2 ↗ EM8C cryo console
Cooling surface 12" x 14.75" in p.2 ↗ EM8C cryo console
Weight 55 lbs. p.2 ↗ EM8C cryo console
Cooling unit temperature 0 to -15 °C p.2 ↗ EM8C cryo console

Operating Limits

  • Paraffin reservoir temperature range: 50 - 70 °C p.2 ↗
  • Working area temperature range: 50 - 70 °C p.2 ↗
  • Cooling unit temperature range: 0 to -15 °C p.2 ↗
  • Power supply: 100-240 V / 50-60 Hz p.2 ↗