Data Sheet Model BD 23
Technical Reference Summary
Covered Models & Document Scope
Incubateur standard BINDER BD 23 (série BD, convection naturelle), volume intérieur 20 L. Deux variantes : BD023-230V (Art.-No. 9010-0187, 230 V 1~ 50/60 Hz, fiche CEE 7/7) et BD023UL-120V (Art.-No. 9010-0189, 120 V 1~ 60 Hz, fiche NEMA 5-15).
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Temperature range | +5 °C above ambient temperature to 100 °C °C p.2 ↗ | BD023-230V |
| Temperature uniformity at 37°C | 0.5 ± K p.2 ↗ | BD023-230V |
| Temperature fluctuation at 37°C | 0.2 ± K p.2 ↗ | BD023-230V |
| Heating up time to 37°C | 59 min p.2 ↗ | BD023-230V |
| Recovery time after door was opened for 30 s at 37°C | 20 min p.2 ↗ | BD023-230V |
| Rated voltage | 230 V p.2 ↗ | BD023-230V |
| Power frequency | 50/60 Hz p.2 ↗ | BD023-230V |
| Nominal power | 0,2 kW p.2 ↗ | BD023-230V |
| Unit fuse | 12,5 A p.2 ↗ | BD023-230V |
| Phase (nominal voltage) | 1~ p.2 ↗ | BD023-230V |
| Interior volume | 20 L p.2 ↗ | BD023-230V |
| Net weight of the unit (empty) | 27 kg p.2 ↗ | BD023-230V |
| Load per rack | 12 kg p.2 ↗ | BD023-230V |
| Permitted load | 25 kg p.2 ↗ | BD023-230V |
| Wall clearance back | 100 mm p.2 ↗ | BD023-230V |
| Wall clearance sidewise | 50 mm p.2 ↗ | BD023-230V |
| Width net | 435 mm p.2 ↗ | BD023-230V |
| Height net | 495 mm p.2 ↗ | BD023-230V |
| Depth net | 520 mm p.2 ↗ | BD023-230V |
| Interior width | 222 mm p.2 ↗ | BD023-230V |
| Interior height | 330 mm p.2 ↗ | BD023-230V |
| Interior depth | 277 mm p.3 ↗ | BD023-230V |
| Unit doors | 1 p.3 ↗ | BD023-230V |
| Energy consumption at 37°C | 11 Wh/h p.3 ↗ | BD023-230V |
| Number of shelves (std./max.) | 2/4 p.3 ↗ | BD023-230V |
Operating Limits
- Temperature range (min): +5 °C above ambient temperature p.2 ↗
- Temperature range (max): 100 °C p.2 ↗
- Wall clearance back: 100 mm p.2 ↗
- Wall clearance sidewise: 50 mm p.2 ↗
- Permitted load: 25 kg p.2 ↗
- Load per rack: 12 kg p.2 ↗
Type Code Decoder
| Position | Token | Meaning |
|---|---|---|
| 1-2 p.1 ↗ | BD | Product series: Standard-Incubator with natural convection |
| 3-4 p.1 ↗ | 23 | Model size (BD 23, 20 L interior volume) |
| suffix p.1 ↗ | UL | UL variant for 120 V / 60 Hz mains (NEMA 5-15 plug); absent = standard 230 V / 50/60 Hz (CEE 7/7 plug) |
| voltage field p.1 ↗ | 230V | Rated voltage 230 V, 1~, 50/60 Hz |
| voltage field p.1 ↗ | 120V | Rated voltage 120 V, 1~, 60 Hz |
Ordering Data
| Ordering Code | Part No. | Variant / Code | S'applique à | Note |
|---|---|---|---|---|
| BD023-230V p.1 ↗ | 9010-0187 | 230 V 1~ 50/60 Hz, CEE 7/7 plug, Standard | Standard model version | |
| BD023UL-120V p.1 ↗ | 9010-0189 | 120 V 1~ 60 Hz, NEMA 5-15 plug, Standard | UL model version |