Technical Reference Summary
Covered Models & Document Scope
Incubateur CO2 modèle CB-S 170 avec stérilisation à air chaud, volume intérieur 170 L. Variantes : CBS170-230V (standard, 200...230 V, fiche CEE 7/7, Art.-No. 9040-0219), CBS170-230V-L (charnière à gauche, 200...230 V, Art.-No. 9040-0220), CBS170UL-120V (standard, 100...120 V, fiche NEMA 5-20, Art.-No. 9040-0221), CBS170UL-120V-L (charnière à gauche, 100...120 V, Art.-No. 9040-0222).
Technical Data
| Parameter |
Value |
Condition |
| Interior volume |
170 L
p.3 ↗ |
all variants |
| Rated voltage |
200...230 V
p.3 ↗ |
CBS170-230V / CBS170-230V-L |
| Rated voltage |
100...120 V
p.3 ↗ |
CBS170UL-120V / CBS170UL-120V-L |
| Power frequency |
50/60 Hz
p.3 ↗ |
all variants |
| Phase (Nominal voltage) |
1~
p.3 ↗ |
all variants |
| Nominal power |
1,3 kW
p.3 ↗ |
all variants |
| Unit fuse |
10,0 A
p.3 ↗ |
CBS170-230V / CBS170-230V-L |
| Unit fuse |
16,0 A
p.3 ↗ |
CBS170UL-120V / CBS170UL-120V-L |
| Temperature range |
to 50 °C
p.3 ↗ |
all variants |
| Temperature uniformity at 37°C |
0.3 ± K
p.3 ↗ |
all variants |
| Temperature fluctuation at 37°C |
0.1 ± K
p.3 ↗ |
all variants |
| Recovery time |
6 min
p.3 ↗ |
all variants |
| Humidity range |
90 ...95 % RH
p.3 ↗ |
all variants |
| CO₂ range |
0 ...20 Vol.-% CO₂
p.3 ↗ |
all variants |
| CO₂ measuring technology |
IR
p.3 ↗ |
all variants |
| CO₂ recovery time |
5 min
p.3 ↗ |
all variants |
| Net weight of the unit (empty) |
96 kg
p.3 ↗ |
all variants |
| Load per rack |
10 kg
p.3 ↗ |
all variants |
| Permitted load |
30 kg
p.3 ↗ |
all variants |
| Wall clearance back |
100 mm
p.3 ↗ |
all variants |
| Wall clearance sidewise |
50 mm
p.3 ↗ |
all variants |
| Width net |
680 mm
p.4 ↗ |
all variants |
| Height net |
870 mm
p.4 ↗ |
all variants |
| Depth net |
715 mm
p.4 ↗ |
all variants |
| Sound-pressure level |
41 dB(A)
p.4 ↗ |
all variants |
Operating Limits
- max temperature: 50 °C
p.3 ↗
- humidity range: 90 ...95 % RH
p.3 ↗
- CO₂ range: 0 ...20 Vol.-% CO₂
p.3 ↗
- permitted load: 30 kg
p.3 ↗
- load per rack: 10 kg
p.3 ↗
- wall clearance back: 100 mm
p.3 ↗
Type Code Decoder
| Position |
Token |
Meaning |
| 1-3
p.1 ↗ |
CBS |
Series: CO2 incubator with hot air sterilization |
| 4-6
p.1 ↗ |
170 |
Interior volume size: 170 L |
| suffix
p.1 ↗ |
UL |
UL version for 100...120 V supply with NEMA 5-20 plug |
| suffix
p.1 ↗ |
-230V |
Version for 200...230 V supply with CEE 7/7 plug |
| suffix
p.1 ↗ |
-120V |
Version for 100...120 V supply with NEMA 5-20 plug |
| suffix
p.1 ↗ |
-L |
Option: with left-hand hinge |
Ordering Data
| Ordering Code |
Part No. |
Variant / Code |
S'applique à |
Note |
| CBS170-230V
p.1 ↗ |
9040-0219 |
|
200...230 V 1~ 50/60 Hz, 10,0 A, CEE 7/7 plug, standard |
Standard model |
| CBS170-230V-L
p.1 ↗ |
9040-0220 |
|
200...230 V 1~ 50/60 Hz, 10,0 A, CEE 7/7 plug, with left-hand hinge |
Left-hand hinge |
| CBS170UL-120V
p.1 ↗ |
9040-0221 |
|
100...120 V 1~ 50/60 Hz, 16,0 A, NEMA 5-20 plug, standard |
UL version |
| CBS170UL-120V-L
p.2 ↗ |
9040-0222 |
|
100...120 V 1~ 50/60 Hz, 16,0 A, NEMA 5-20 plug, with left-hand hinge |
UL version, left-hand hinge |