Data Sheet Model KB ECO 400

Technical Reference Summary

Covered Models & Document Scope

Incubateur réfrigéré BINDER KB ECO 400 (refroidissement thermoélectrique), volume intérieur 400 L. Deux variantes de tension : KBECO400-230V (Art.-No. 9020-0465, 200...230 V, fiche CEE 7/7) et KBECO400-120V (Art.-No. 9020-0466, 100...120 V, fiche NEMA 5-20), toutes deux version Standard, 1~ 50/60 Hz.

Technical Data

Parameter Value Condition
Temperature range 0...70 °C p.2 ↗ max. 28 °C under ambient temperature; KBECO400-230V / KBECO400-120V
Temperature uniformity at 37°C 0.1 ± K p.2 ↗ KBECO400-230V / KBECO400-120V
Temperature fluctuation at 37°C 0.1 ± K p.2 ↗ KBECO400-230V / KBECO400-120V
Max. heat compensation at 40°C 350 W p.2 ↗ KBECO400-230V / KBECO400-120V
Recovery time after door was opened for 30 s at 37°C 6 min p.2 ↗ KBECO400-230V / KBECO400-120V
Rated Voltage 200...230 V p.2 ↗ KBECO400-230V
Rated Voltage 100...120 V p.2 ↗ KBECO400-120V
Power frequency 50/60 Hz p.2 ↗ KBECO400-230V / KBECO400-120V
Nominal power 1,2 kW p.2 ↗ KBECO400-230V / KBECO400-120V
Unit fuse 16,0 A p.2 ↗ KBECO400-230V / KBECO400-120V
Phase (Nominal voltage) 1~ p.2 ↗ KBECO400-230V / KBECO400-120V
Interior volume 400 L p.2 ↗ KBECO400-230V / KBECO400-120V
Net weight of the unit (empty) 196 kg p.2 ↗ KBECO400-230V / KBECO400-120V
Load per rack 30 kg p.2 ↗ KBECO400-230V / KBECO400-120V
Permitted load 120 kg p.2 ↗ KBECO400-230V / KBECO400-120V
Wall clearance back 100 mm p.2 ↗ KBECO400-230V / KBECO400-120V
Wall clearance sidewise 100 mm p.2 ↗ KBECO400-230V / KBECO400-120V
Width net 925 mm p.2 ↗ KBECO400-230V / KBECO400-120V
Height net 1,946 mm p.2 ↗ KBECO400-230V / KBECO400-120V
Depth net 796 mm p.2 ↗ KBECO400-230V / KBECO400-120V
Interior width 650 mm p.2 ↗ KBECO400-230V / KBECO400-120V
Interior height 1,268 mm p.2 ↗ KBECO400-230V / KBECO400-120V
Interior depth 485 mm p.3 ↗ KBECO400-230V / KBECO400-120V
Sound-pressure level 43 dB(A) p.3 ↗ KBECO400-230V / KBECO400-120V
Energy consumption at 25°C 45 Wh/h p.3 ↗ KBECO400-230V / KBECO400-120V

Operating Limits

  • Temperature range: 0...70 °C (max. 28 °C below ambient temperature) p.2 ↗
  • Wall clearance back: 100 mm p.2 ↗
  • Wall clearance sidewise: 100 mm p.2 ↗
  • Permitted load: 120 kg p.2 ↗
  • Load per rack: 30 kg p.2 ↗

Type Code Decoder

Position Token Meaning
1-3 p.1 ↗ KB Series: cooling incubator
4-7 p.1 ↗ ECO Model version: thermoelectric cooling (ECO series)
8-10 p.1 ↗ 400 Size: interior volume 400 L
suffix p.1 ↗ 230V Mains voltage variant 200...230 V, 1~ 50/60 Hz, CEE 7/7 plug, Art.-No. 9020-0465
suffix p.1 ↗ 120V Mains voltage variant 100...120 V, 1~ 50/60 Hz, NEMA 5-20 plug, Art.-No. 9020-0466

Ordering Data

Ordering Code Part No. Variant / Code S'applique à Note
KBECO400-230V p.1 ↗ 9020-0465 200...230 V 1~ 50/60 Hz, CEE 7/7 plug, Standard Cooling incubator, 400 L
KBECO400-120V p.1 ↗ 9020-0466 100...120 V 1~ 50/60 Hz, NEMA 5-20 plug, Standard Cooling incubator, 400 L