Data Sheet Model MK COâ 1020
Technical Reference Summary
Covered Models & Document Scope
Cámara climática dinámica de CO₂ modelo MK 1020 para cambios rápidos de temperatura, volumen interior de 991 L. Dos variantes de alimentación: MK1020CO2-400V (400 V 3~/N/PE 50 Hz, n.º de art. 9020-0527) y MK1020CO2-480V (480 V 3~/N/PE 60 Hz, n.º de art. 9020-0529).
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Temperature range | -48...180 °C p.3 ↗ | MK1020CO2-400V, 50 Hz |
| Temperature uniformity dependent on set | 0.1...1.0 ± K p.3 ↗ | MK1020CO2-400V, 50 Hz |
| Temperature fluctuation dependent on set | 0.1...0.2 ± K p.3 ↗ | MK1020CO2-400V, 50 Hz |
| Cooling down time from 180°C to -40°C | 70 min p.3 ↗ | MK1020CO2-400V, 50 Hz |
| Rated voltage | 400 V p.3 ↗ | MK1020CO2-400V, 50 Hz |
| Power frequency | 50 Hz p.3 ↗ | MK1020CO2-400V |
| Nominal power | 9,6 kW p.3 ↗ | MK1020CO2-400V, 50 Hz |
| Unit fuse | 32,0 A p.3 ↗ | MK1020CO2-400V, 50 Hz |
| Phase (nominal voltage) | 3~/N/PE p.3 ↗ | MK1020CO2-400V, 50 Hz |
| Interior volume | 991 L p.3 ↗ | MK1020CO2-400V, 50 Hz |
| Net weight of the unit (empty) | 556 kg p.3 ↗ | MK1020CO2-400V, 50 Hz |
| Load per rack | 40 kg p.3 ↗ | MK1020CO2-400V, 50 Hz |
| Permitted load | 200 kg p.3 ↗ | MK1020CO2-400V, 50 Hz |
| Wall clearance back | 300 mm p.3 ↗ | MK1020CO2-400V, 50 Hz |
| Wall clearance sidewise | 200 mm p.3 ↗ | MK1020CO2-400V, 50 Hz |
| Width net | 1,580 mm p.3 ↗ | MK1020CO2-400V, 50 Hz |
| Height net | 2,005 mm p.3 ↗ | MK1020CO2-400V, 50 Hz |
| Depth net | 1,480 mm p.3 ↗ | MK1020CO2-400V, 50 Hz |
| Interior width | 1,200 mm p.4 ↗ | MK1020CO2-400V, 50 Hz |
| Interior height | 1,020 mm p.4 ↗ | MK1020CO2-400V, 50 Hz |
| Interior depth | 810 mm p.4 ↗ | MK1020CO2-400V, 50 Hz |
| Unit doors | 1 p.4 ↗ | MK1020CO2-400V, 50 Hz |
| Sound-pressure level | 62 dB(A) p.4 ↗ | MK1020CO2-400V, 50 Hz |
| Energy consumption at 25°C | 2,500 Wh/h p.4 ↗ | MK1020CO2-400V, 50 Hz |
| Number of shelves (std./max.) | 1/11 p.4 ↗ | MK1020CO2-400V, 50 Hz |
Operating Limits
- Temperature range: -48...180 °C p.3 ↗
- Permitted load: 200 kg p.3 ↗
- Load per rack: 40 kg p.3 ↗
- Wall clearance back: 300 mm p.3 ↗
- Wall clearance sidewise: 200 mm p.3 ↗
Type Code Decoder
| Position | Token | Meaning |
|---|---|---|
| 1-2 p.1 ↗ | MK | Series: dynamic climate chamber |
| 3-6 p.1 ↗ | 1020 | Model size (interior volume class, 991 L) |
| 7-9 p.1 ↗ | CO₂ | Refrigerant: natural CO₂ (R744) |
| 10-14 p.3 ↗ | 400V | Mains variant: 400 V 3~/N/PE 50 Hz (alternative 480V = 480 V 3~/N/PE 60 Hz) |
Ordering Data
| Ordering Code | Part No. | Variant / Code | S'applique à | Note |
|---|---|---|---|---|
| MK1020CO2-400V p.1 ↗ | 9020-0527 | 400 V 3~/N/PE 50 Hz, standard version | CEE 32 / 6 H plug 5-p | |
| MK1020CO2-480V p.2 ↗ | 9020-0529 | 480 V 3~/N/PE 60 Hz, standard version | NEMA L22-20P plug |