Induztr
Résumé technique Visionneuse de documents Modèles associés

Data Sheet Model MK 115

Technical Reference Summary

Covered Models & Document Scope

Cámara climática dinámica BINDER MK 115 para cambios rápidos de temperatura, volumen interior 115 L. Dos versiones de modelo: MK115-400V (estándar, 400 V 3~/N/PE 50 Hz, Art.-No. 9020-0375) y MK115-480V-C (con convertidor de tensión y frecuencia, 480 V 3~ 60 Hz, Art.-No. 9020-0303).

Technical Data

Parameter Value Condition
Temperature range -40...180 °C p.3 ↗ MK115-400V
Temperature uniformity dependent on set 0.1...2.0 ± K p.3 ↗ MK115-400V
Temperature fluctuation dependent on set 0.1...0.5 ± K p.3 ↗ MK115-400V
Cooling down time from 180°C to -40°C 90 min p.3 ↗ MK115-400V
Rated voltage 400 V p.3 ↗ MK115-400V
Power frequency 50 Hz p.3 ↗ MK115-400V
Nominal power 3,5 kW p.3 ↗ MK115-400V
Unit fuse 16,0 A p.3 ↗ MK115-400V
Phase (nominal voltage) 3~/N/PE p.3 ↗ MK115-400V
Interior volume 115 L p.3 ↗ MK115-400V
Net weight of the unit (empty) 255 kg p.3 ↗ MK115-400V
Load per rack 30 kg p.3 ↗ MK115-400V
Permitted load 60 kg p.3 ↗ MK115-400V
Wall clearance back 300 mm p.3 ↗ MK115-400V
Wall clearance sidewise 200 mm p.3 ↗ MK115-400V
Width net 980 mm p.3 ↗ MK115-400V
Height net 1,725 mm p.3 ↗ MK115-400V
Depth net 865 mm p.3 ↗ MK115-400V
Interior width 600 mm p.4 ↗ MK115-400V
Interior height 480 mm p.4 ↗ MK115-400V
Interior depth 400 mm p.4 ↗ MK115-400V
Unit doors 1 p.4 ↗ MK115-400V
Sound-pressure level 62 dB(A) p.4 ↗ MK115-400V
Energy consumption at 20°C 650 Wh/h p.4 ↗ MK115-400V
Number of shelves (std./max.) 1/4 p.4 ↗ MK115-400V

Operating Limits

  • Temperature range: -40...180 °C p.3 ↗
  • Wall clearance back: 300 mm p.3 ↗
  • Wall clearance sidewise: 200 mm p.3 ↗
  • Permitted load: 60 kg p.3 ↗
  • Load per rack: 30 kg p.3 ↗

Type Code Decoder

Position Token Meaning
1-2 p.1 ↗ MK Series: dynamic climate chamber for rapid temperature changes
3-5 p.1 ↗ 115 Model size / interior volume 115 L
suffix p.1 ↗ -400V Standard model version, 400 V 3~/N/PE 50 Hz
suffix p.2 ↗ -480V-C Model version with voltage and frequency converter, 480 V 3~ 60 Hz

Ordering Data

Ordering Code Part No. Variant / Code S'applique à Note
MK115-400V p.1 ↗ 9020-0375 Standard, 400 V 3~/N/PE 50 Hz, 16,0 A, CEE 16/6 H plug 5-pin Standard model version
MK115-480V-C p.2 ↗ 9020-0303 with voltage and frequency converter, 480 V 3~ 60 Hz, 16,0 A, CEE 16/6 H plug 5-pin Model version with voltage and frequency converter