Induztr
Résumé technique Visionneuse de documents Modèles associés

Formulas, Properties, and Tables for Vacuum Technology

Sommaire

  1. Overview pp. 1-2
    Cover and introduction from Agilent vacuum experts
  2. p. 3
    Document contents listing
  3. Spec pp. 4-8
    Vacuum formulas, gas flow characteristics, and conversion tables
  4. p. 9
    DIN 28401 vacuum technology symbols
  5. Overview pp. 10-11
    Vacuum pump accessories and valve operation modes
  6. Overview p. 12
    Overview of vacuum gauge types
  7. pp. 13-23
    Alphabetical glossary of vacuum terms and definitions
  8. Overview p. 24
    Introductory summary of vacuum concepts
  9. p. 8 table, 22 rows
  10. p. 4 table, 14 rows
  11. p. 5 table, 6 rows
  12. p. 6 table, 4 rows
  13. p. 7 table, 5 rows
  14. p. 15 table, 15 rows
  15. p. 16 table, 16 rows

Vacuum technology standards

NumberTitle
DIN 28400Vacuum Technology; Designations and Definitions
DIN 28401Graphic Symbols in Vacuum Technology
DIN 28402Vacuum Technology: Variables, Symbols, Units - Overview
DIN 28403 ISO 1609 PNEUROP 6606Vacuum Technology; Quick Connections, Small Flange Connections
DIN 28404 ISO 1609 PNEUROP 6606Vacuum Technology: Flanges, Dimensions
DIN 28410Vacuum Technology; Mass Spectrometer Partial Pressure Gauges, Definitions
DIN 28411 ISO 3530.2Mass Spectrometer type Leak Detector Calibration
DIN 28416Calibration of Vacuum Gauges – General Method
DIN 28417Measurement of Throughput by Volumetric Method
DIN 28418 ISO/DIS 3567Vacuum Gauges – Calibration by Direct Comparison
DIN 28426, part I, II ISO 1607 / 1,2 PNEUROP 6602Positive Displacement Vacuum Pumps- Measurement of Performance Characteristics. Measurement of Ultimate Pressure
DIN 28427 ISO 1608 / 1,2 PNEUROP 5607Vapor Vacuum Pumps - Measurement of Performance Characteristics. Measurement of Critical Backing Pressure
DIN 28428 PNEUROP 5608Vacuum Technology; Acceptance Specifications for Turbo Molecular Pumps
DIN 28429 PNEUROP 5615Vacuum Technology; Acceptance Specifications for Getter Pumps
DIN 28430 PNEUROP 6601Measurement of Performance of Ejector Vacuum Pumps and Ejector Compressors
ISO 1314Pressure; Basic Definitions, Units
ISO 3529 I,II,IIIVacuum Technology Vocabulary
ISO/DIS 3556 / 1Sputter Ion Pumps - Measurement of Performance Characteristics.
ISO/DIS 3568Ionization Vacuum Gauges – Calibration by direct comparison
ISO/DIS 3570 / 1Vacuum Gauges – Standard Methods for Calibration
ISO/DIS 3669Bakeable Flange Dimensions
PN5ASR CC/5Vacuum Pumps, Acceptance Specifications Refrigerator Cooled Cryopumps

ISO - International Standardization Organization – Switzerland; DIN - Deutsches Institut fuer Normung - Germany; PNEUROP – European Committee of Manufacturers of Compressors, Vacuum Technology, Pneumatic Tools, Air Treatment Equipment and Condensate Treatment Equipment - England

Physical properties of gases

GasChemical formulaMolecular weight
HydrogenH22.016
HeliumHe4.002
Deuterium2H4.028
MethaneCH416.04
AmmoniaNH317.03
Water (vapour)H2O18.02
NeonNe20.18
NitrogenN228.01
OxygenO231.99
ArgonAr39.94
Carbon dioxideCO244.01
KryptonKr83.80
XenonXe131.30
MercuryHg200.59

Some molecular relationships (at 273 K)

PressureMolecular DensityMolecular CollisionMean Free PathMonolayer Formation
Torrmolec./cm3molec./cm2 • seccmtime (sec)
7603.25 • 10193.78 • 10235.1 • 10-62.2 • 10-9
10-33.25 • 10133.78 • 10175.12.2 • 10-3
10-63.25 • 10103.78 • 101451002.2
10-93.25 • 1073.78 • 10115.1 • 1062200
10-123.25 • 1043.78 • 1085.1 • 1092.2 • 106

Pumping speed - Conversion table

PressureL/sL/minm3/hCubic Feet per Minute
1 liter per second1603.62.19
1 liter per minute0.0166610.060.0353
1 cubic meter per hour0.28716.6710.589
1 cubic foot per minute0.47228.321.701

Unit of throughput - flow - leak rate - conversion table

Flow or leak rateSTD (cc/s)atm (cc/s)mbar (L/s)molecules/s (at 0 °C)Torr (L/s)Pa (m3/s)sccmL/sL/minm3/hCubic Feet per Minute
1 STD cc/sec – 1 atm cc/sec – 1 mbar l/sec12.687 • 10190.760.160
1 molecule/s3.72 • 10-2012.86 • 10-203.72 • 10-212.23 • 10-18
1 Torr l/sec1.33.493 • 101910.1380
1 Pa m3/sec102.687 • 10207.51600
1 sccm0.0164.299 • 10170.01250.0161

Значения приведены дословно, как напечатано; пустые ячейки оставлены пустыми.

Термины и определения

ТерминОпределение
CrossoverThe pressure at which a vacuum chamber is changed from being pumped by a roughing pump to being pumped by a high vacuum pump.
Critical forepressureSee maximum tolerable foreline pressure.
CryocondensationThe pumping of gases that are condensed at cold temperatures. For example, water vapor on a liquid nitrogen trap at -196 °C undergoes cryocondensation.
ConvectionThe transfer of heat from one place to another by the circulation of currents of heated gas or other fluid.
CryosorptionThe pumping of gases that are not readily condensed (or pumped) at cold temperatures, by the process of sticking onto a cold surface.
CryotrapA device usually placed before the inlet of a high vacuum pump to “trap” or freeze out gases such as pump oil vapor and water vapor. Cryotraps commonly use liquid nitrogen as the coolant. Also called cold trap or liquid nitrogen trap.
DegassingThe removal of gas from a material, usually by application of heat under high vacuum. (See bakeout.)
DesorptionSee outgassing.
Diffusion1. The flow of one substance through another by random molecular motion. 2. The process by which molecules intermingle as a result of their thermal motion.
Diffusion pumpA vapor pump having boiler pressures of a few Torr and capable of pumping gas continuously at intake pressures. These do not exceed about 2 mTorr, and discharge pressures (forepressures) do not exceed about 500 mTorr. The term diffusion should be applied only to pumps in which the pumping action of each vapor jet occurs as follows: The gas molecules diffuse through the low-density scattered vapor into the denser, forward-moving core of the freely expanding vapor jet. Most of the gas molecules are then driven at an acute angle toward the wall and on into the fore vacuum.
FeedthroughA device used to allow some sort of utility service to go from the outside world to the inside of a vacuum system, while maintaining the integrity of the vacuum. An example is an electrical feedthrough.
EvaporationThe process that happens when a liquid or solid becomes a gas.
Dynamic sealA seal that moves. (See static seal.)
ElectronA negatively charged particle. (See atom.)
ConductionThe transfer of energy (heat, light, etc.) by direct contact. In the case of gaseous conduction, the transfer of energy by molecules directly contacting surfaces and other molecules.

Термины и определения

ТерминОпределение
Gas densityThe number of molecules per unit of volume.
Gay-lussac’s lawThe gas law that states that if the temperature of a volume of gas at 0 °C is changed by 1 °C, the volume will change (plus or minus, as appropriate) by 1/273 of its original value.
High vacuumPressure which ranges from about 10-4 Torr (0.0001 Torr) to approximately 10-8 Torr (0.00000001 Torr).
Gas loadSee pressure, gauge.
Helium mass spectrometer leak detector (hmsld)See mass spectrometer leak detector.
General gas lawThe gas law that covers pressure, volume, and temperature in one single equation, or P1V1T2 = P2V2T1.
High vacuum pumpA vacuum pump which will function in the high vacuum range. Common examples are the diffusion pump and the mechanical cryopump.
GetteringA method of pumping gases through chemical reaction of a material with gas molecules. The material usually used is an active element such as titanium. (see chemisorption.)
Foreline valveA vacuum valve placed in the foreline to permit isolation of the pump from the forepump.
High vacuum valveA large diameter valve usually placed between the vacuum chamber and the vacuum pumps. It is used to isolate the vacuum chamber from the pumps when it is necessary to work on something in the chamber. Also called hi-vac valve, gate valve, or trap valve.
ForepumpThe pump which is used to exhaust another pump, which is incapable of discharging gases at atmospheric pressure. Also called the backing pump.
FractionationA process that helps to purify the condensed fluid in a diffusion pump. This process removes contaminants produced by decomposition of pump fluid.
ImplosionIn vacuum work, the inward collapse of the walls of a vacuum system, caused by external pressure.
GasA state of matter where the individual particles are free to move in any direction and tend to expand uniformly to the confines of a container.
Gas ballastA method used with any oil-sealed rotary pump which allows a quantity of air to be admitted during the compression cycle to prevent condensation of water vapor. The amount of air admitted is regulated by the gas ballast valve. The use of the gas ballast raises the ultimate pressure of the pump.
ForelineThe section of a pump through which the gases leave. The exhaust line of a pump.

Engineering Facts

12 facts · every claim linked to a page
Operation

Viscous flow occurs when the product of pressure and diameter exceeds 0.7 mbar·cm, so piping and chamber sizing must keep p·D in the correct flow regime.

Determines whether conductance formulas for viscous or molecular flow apply when sizing forelines and roughing lines.

Source
“p • D > 0.7 (mbar cm);”
Operation

Molecular flow prevails when p·D is below 1.3·10⁻² mbar·cm, meaning high-vacuum conductance is pressure-independent and must be calculated with molecular-flow formulas.

Selecting high-vacuum plumbing and pump speed requires knowing the flow regime at the operating pressure.

Source
“p • D < 1.3 • 10-2 (mbar cm);”
Operation

The viscous-flow pump-down equation is valid only from start to about 1 mbar; below that, outgassing becomes significant and the calculation is no longer accurate.

Sets the limit for using rough pump-down time estimates and signals when outgassing must be considered.

Source
“This equation is accurate from start to approximately 1 mbar.”
Performance

Ultimate pressure is set by gas load divided by pumping speed plus the pump's compression-limited outlet contribution for each gas species.

Explains why base pressure depends on both pump speed and compression ratio, affecting pump selection for required vacuum.

Source
“The ultimate pressure of the vacuum system is”
Maintenance

Stainless steel baked at 150 °C for 24 h has an outgassing rate of 4×10⁻¹² mbar·L/s·cm², orders of magnitude lower than unbaked stainless.

Justifies bakeout of stainless chambers when low ultimate pressure is required.

Source
“Stainless – 24 h baked at 150 °C”
Maintenance

Viton A dry outgassing falls from 2×10⁻⁶ to 1×10⁻⁹ mbar·L/s·cm² between 1 h and 100 h, so elastomer seals dominate early pump-down.

Informs seal material selection and expected pump-down time when elastomer seals are used.

Source
“Viton A – Dry”
Operation

Using gas ballast admits air during compression to prevent water vapor condensation, but it raises the pump's ultimate pressure.

Affects roughing pump operation when pumping condensable vapors and the achievable base pressure.

Source
“The use of the gas ballast raises the ultimate pressure of the pump.”
Selection

A diffusion pump's intake pressure should not exceed about 2 mTorr and forepressure about 500 mTorr, defining its operating envelope.

Sets the crossover and backing pressure limits when selecting and operating a diffusion pump.

Source
“These do not exceed about 2 mTorr, and discharge”
Safety

In vacuum work, implosion is the inward collapse of vacuum system walls caused by external pressure, a hazard for large chambers and glass bell jars.

Requires pressure-rated chamber design and guarding against implosion during installation and operation.

Source
“In vacuum work, the inward collapse of the walls of a vacuum system, caused by external pressure.”
Operation

A virtual leak is caused by outgassing or trapped gas inside the system, not by a real hole, and can be distinguished using a rate-of-rise test.

Prevents misdiagnosis of outgassing as a real leak and guides proper leak-detection procedure.

Source
“A virtual leak, which is caused by outgassing of some volatile material inside a”
Operation

The rate-of-rise test isolates the system from the pump and observes pressure rise to determine whether leakage is real or virtual.

Provides a field method to distinguish real leaks from outgassing before committing to repair.

Source
“This method also”
Conformité

DIN 28411 / ISO 3530.2 governs calibration of mass-spectrometer-type leak detectors, which matters for traceable leak-rate measurements.

Identifies the standard to follow when calibrating leak detectors for compliant leak testing.

Source
“Mass Spectrometer type Leak Detector Calibration”