Table of contents listing document sections and page numbers
Warningpp. 5-10
General safety information, intended use, precautions and hazards
Specpp. 11-16
Electrical, computer, gas and environmental specifications
Procedurepp. 17-18
Installation instructions for the multicell accessories
Overviewpp. 19-20
Overview of the Cary UV-Vis flexible module
Config.pp. 21-22
System configurations and temperature probe options
Overviewpp. 23-25
Instrument components, rear panel and indicator LED
Procedurepp. 26-27
Using the software and drain outlet operation
Procedurepp. 28-30
Troubleshooting, maintenance and spill handling
Orderingpp. 31-32
Spare parts list and ordering information
p. 21table, 6 rows
p. 31table, 5 rows
p. 14table, 6 rows
p. 15table, 2 rows
p. 27table, 5 rows
Table 3 Cary 3500 UV-Vis configurations
Product number
System
Product number
Module only
G9871A
Cary 3500 Compact UV-Vis System
G9893A
Cary 3500 Compact Module
G9862AA
Cary 3500 Compact Peltier UV-Vis
G9894A
Cary Compact Peltier UV-Vis Module
G9873A
Cary 3500 Multi UV-Vis System
G9888A
Cary 3500 Multicell Module
G9866AA
Cary 3500 Multicell Peltier UV-Vis
G9889A
Cary Multicell Peltier UV-Vis Module
G9875A
Cary 3500 Flexible UV Vis System
G9886A
Cary UV-Vis Flexible Module
G9867AA
Cary 3500 Multizone Peltier UV-Vis
Table 6 Spare parts for Cary 3500 UV-Vis
Part
Part Number
Lamp module
G9864-67019
Temperature probe
G9889-60005
Magnetic stirrer bar PTFE star type
7418000400
Flow meter
K8003-60001
Purge tubing
3710043100
Table 1 Computer requirements
Component
Requirement
Processor
Intel i5, 3.0 GHz, 18MB cache
Memory
16GB RAM
Storage
512 GB Solid State Drive
Graphics
Intel® UHD Graphics 770
Communications
PCIe GbE Controller; 2nd Serial Port Adapter, 2nd LAN NIC
Operating system
Microsoft Windows 10 and 11 Professional and Enterprise 64-bit
Table 2 Purge gas purity requirements
Compressed gas
Description
Nitrogen
Water vapor <2570 ppm (dew point @ -10 °C). Acceptable sources are from bottled gas with industrial, medical or food grade nitrogen.
Air
Compressed Air: ISO 8573-1:2010 CLASS 1.3.1
Table 5 Purge gas flow rate requirements
Product number
Description
Flow rate (LPM)
G9862AA
Cary 3500 Compact Peltier UV-Vis
5
G9866AA
Cary 3500 Multicell Peltier UV-Vis
15
G9867AA
Cary 3500 Multizone Peltier UV-Vis
15
G9894A
Cary Compact Peltier UV-Vis Module
5
G9889A
Cary Multicell Peltier UV-Vis Module
15
Engineering Facts
14 facts · every claim linked to a page
Safety
The lamp module contains high-voltage components and must never be disassembled; only replace it after powering off and disconnecting the engine from mains.
Prevents electric shock during lamp replacement and defines the only user-serviceable high-voltage item.
Source
“This module contains components operating at high voltages. To avoid electric shock, NEVER disassemble the module.”
Installation
Only the sample compartment and lamp module are customer-accessible; all other screwed panels must be opened only by Agilent service engineers.
Defines service boundaries and prevents unauthorized access that could void safety compliance.
Source
“Any other panels or covers that are retained by screws on the spectrophotometer and accessories may be opened ONLY by Agilent service engineers.”
Installation
Peltier-controlled models need 20 cm clearance on both sides and about 10 cm at the rear for electrical, gas connections and airflow.
Insufficient clearance causes overheating and unreliable temperature control on Peltier systems.
Source
“Allow for 20 cm (8 inches) of space on both sides and approximately 10 cm (4 inches) in the rear for electrical, gas connections and air flow.”
Installation
Non-Peltier models require 10 cm clearance on both sides and about 10 cm at the rear for electrical connections and airflow.
Ensures adequate cooling and access for ambient (non-Peltier) configurations.
Source
“Allow for 10 cm (4 inches) of space on both sides and approximately 10 cm (4 inches) in the rear for electrical connections and air flow.”
Conformité
The system requires a 3-wire earth-grounded outlet; a separate receptacle per unit is recommended and extension cords or adapters are not recommended.
Grounding is mandatory to avoid shock hazard and to meet electrical safety requirements.
Source
“A 3-wire outlet with ground connection must be provided for the Cary system.”
Selection
Peltier modules require dry, oil-free, particle-free, non-combustible compressed gas (nitrogen or air) to prevent condensation at sub-ambient temperatures.
Gas quality directly affects condensation prevention and optical surface protection.
Source
“The supplied gas must be dry, oil and particle free and non-combustible.”
Operation
Multicell Peltier zones can be set from -5 to 110 °C and controlled by cell holder (block) or sample temperature via the probe accessory.
Defines the usable temperature range and control mode for thermal applications.
Source
“the temperature of each zone can be varied over a range of -5 to 110 °C”
Safety
The sample compartment and cuvettes can remain hot after operation; allow at least five minutes cooling before removing cuvettes.
Prevents burns during routine sample handling on heated Peltier modules.
Source
“Allow the sample compartment and cuvettes to cool for at least five minutes before attempting to remove cuvettes.”
Safety
The Flexible module contains strong magnets; keep at least 145 mm away from devices and objects that could be damaged, and follow AIMD manufacturer guidance.
Protects pacemakers, storage media and electronics from magnetic field damage.
Source
“Keep magnets at least 145 mm away from devices and objects that could be damaged by strong magnetic fields.”
Installation
The purge kit maximum pressure is 689 kPa (100 PSIG); always use an appropriate regulator and gauge to maintain correct purge pressure.
Prevents over-pressurizing the module and ensures stable purge flow.
Source
“The operating pressure of the Agilent supplied gauge is up to a maximum of 689 kPa (100 PSIG).”
Installation
Use clean flexible PVC (or equivalent) 6 mm OD/4 mm ID tubing; never use rubber tubing because talc can be blown into the optical system.
Rubber tubing contaminates optics and degrades photometric performance.
Source
“Never use rubber tubing, because it may be treated internally with talc that can be blown into the optical system.”
Installation
Never connect gas directly to the module; always use a pressure regulator meeting the manual specifications.
Direct connection risks overpressure damage to the module and seals.
Source
“Never connect the gas directly to the module. Always use a pressure regulator that meets the specifications listed in this manual.”
Maintenance
Multicell and Compact modules require a drain vessel and tubing over the 6 mm OD drain outlet; the user must supply a chemically inert container of at least 2 liters.
Prevents spilled liquids from reaching the benchtop and ensures safe waste collection.
Source
“A chemically inert container, not glass or of a narrow-necked style, to hold a minimum of 2 liters (4 pints) of waste must be provided by the user.”
Maintenance
Replace blown fuses only with the type and rating indicated on the rear of the instrument; other internal fuses are not operator-accessible.
Wrong fuse rating reduces safety protection and may cause fire.
Source
“replace the blown fuse with one of the type and rating as indicated in the on the rear of the instrument.”