Induztr
Résumé technique Visionneuse de documents Modèles associés

User's Guide

Sommaire

  1. Warning p. 2
    Legal and safety notices for the instrument
  2. pp. 3-6
    Table of contents listing guide sections
  3. Overview pp. 7-10
    Product overview and information symbols
  4. Procedure pp. 11-17
    Installation, components, power cord, environment, software, shutdown
  5. Spec p. 18
    Instrument diagnostic parameter values
  6. Procedure pp. 19-20
    Validation and spectral resolution check procedures
  7. Procedure pp. 21-23
    Sample analysis and accessory interchange procedures
  8. Procedure pp. 24-27
    Diamond ATR setup, background collection, cleaning
  9. Procedure pp. 28-31
    ZnSe ATR setup, background collection, cleaning
  10. Procedure pp. 32-34
    Germanium ATR setup, background collection, cleaning
  11. Procedure pp. 35-36
    Multireflection ZnSe ATR operation and cleaning
  12. Procedure pp. 37-39
    Transmission sampling setup and cleaning
  13. Procedure pp. 40-42
    10 and 45 degree specular reflectance operation and cleaning
  14. Procedure pp. 43-44
    TumblIR sampling accessory operation
  15. Procedure pp. 45-48
    DialPath sampling, analysis, cleaning, background collection
  16. Procedure pp. 49-52
    Diffuse reflectance setup, cleaning, background, analysis
  17. Procedure pp. 53-60
    Window and source replacement maintenance procedures
  18. Ordering pp. 61-62
    Spare parts and consumables list
  19. Spec p. 63
    Instrument technical specifications
  20. Warning pp. 64-66
    Safety information and guide conventions
  21. p. 18 table, 5 rows
  22. p. 11 table, 10 rows
  23. p. 61 table, 16 rows
  24. p. 16 table, 5 rows

Table 2 Diagnostic values

ValueOptimal (green)Marginal (yellow)Critical (red)Comment
Energy (center-burst)28,000-15,000>28,000 or <10000> 30,000 or < 3000Indicates the overall alignment and proper gain adjustment for the system.
Source2.1 A>2.4 A or <1.6 A>2.7 A or <1.4 AIndicates a problem with the infrared source control voltage, or that the source has burnt out. Both voltage and amperage are displayed; however, amperage provides a sufficient diagnostic.
Laser signal15,000-3,000>15,000 or <3,000>17,000 or < 2,000Can verify gross alignment errors even if the reflectance cap is not in place.
Detector temperature38–42<38 or >42<36 or >44Indicates a problem with the cooling circuitry or an ambient temperature over the specified range.
CPU temperature10–75>75>80Indicates the ambient temperature over the specified range.

When using the Diffuse Reflectance sampling accessory, the appropriate reflectance disk must be in place to make these measurements. The diffuse gold reference disk should be used with the Diffuse Reflectance sampling accessory. This disk is permanently mounted into Position 1 of the sampling tray.

Table

РазделСтраница
Unpacking11
Installation12
Components13
Connecting power13
Power cord selection14
Environment15
MicroLab software16
Turning On the Instrument16
Turning Off17
Performance Verification17

Оглавление раздела Getting Started; номера страниц приведены как напечатано.

Table 3 Spare parts

Part numberDescription
G8043-67483Replacement flat tip for ATR
G8043-67484Replacement pellet tip for ATR
G8043-67405Polystyrene film for ATR/Transmission sampling accessory
G8043-67406Polystyrene film for TumblIR/DialPath sampling accessory
PIKE-162-5450NIST traceable polystyrene from PIKE approximately 38 µm (for IQOQ)
925-0128NIST traceable polystyrene from ICL approximately 35 µm (for IQOQ)
G8043-67400Infrared source with purge fitting
G8043-67401Desiccant with purge fitting
G8043-67402KBr window with removal tool
G8043-67403ZnSe window with removal tool
G8043-67003Cary FTIR Swivel Diamond ATR Press
G8043-67004Cary FTIR Swivel Ge-ZnSe Press
G8043-67012Purge kit for engine and Transmission sampling accessory
G8043-67524Cary FTIR Base Plate Rubber Bumper+Screw
G8043-67407Cary 635 Power Supply
G8043-67408Cary 635 USB Cable

For service contracts, repairs and refurbishment services, contact Agilent.

Table 1 Power switch LED status

StatusLED-colorAction (% duty cycle)
System OFFNo LED illuminationN/A
System OFFRed100%
System startupRed / greenGreen 0.5 s on / red 0.5 s on
System ONGreen100%
Firmware updateRed / greenGreen blinks rapidly twice then red blinks rapidly twice

Engineering Facts

14 facts · every claim linked to a page
Installation

The instrument must be allowed to reach room temperature before unpacking to prevent condensation.

Prevents condensation damage during initial setup and ensures proper operation.

Source
“place the shipping case in a room-temperature environment and allow several hours for the contents of the container to reach the room ambient temperature.”
Installation

Shipping in non-original packaging voids the warranty.

Mandates use of original packaging for any transport to preserve warranty coverage.

Source
“Shipping the Cary 635 FTIR in packaging other than the original shipping container will void the warranty.”
Installation

The instrument must be kept away from hot surfaces and electromagnetic interference sources.

Ensures environmental conditions do not degrade performance or cause malfunction.

Source
“Your Cary 635 FTIR must be kept away from hot surfaces and sources of electromagnetic interference.”
Operation

A 5-minute warmup is required before reliable analysis.

Guarantees stable measurements and accurate results after power-on.

Source
“The spectrometer must have a warmup period of 5 minutes before reliable analysis may begin.”
Operation

Leaving the spectrometer on helps keep moisture out.

Prevents moisture ingress and extends component life when not actively measuring.

Source
“To help keep moisture out of the Cary 635 FTIR, it is recommended to leave the spectrometer on, even when not in use.”
Performance

Performance validation requires at least 120 minutes of warmup before testing.

Ensures validation tests reflect stable instrument performance and avoid false failures.

Source
“The Cary 635 FTIR system should be warmed up for at least 120 minutes before conducting any of these tests.”
Safety

Only measure samples with pH between 1 and 14 on the diamond ATR to avoid damage.

Prevents chemical damage to the diamond crystal or metal mounting plate.

Source
“Only measure samples with a pH between 1 and 14.”
Safety

The diamond ATR window is thin and can crack under extreme pressure; avoid sharp samples.

Avoids costly damage to the ATR accessory during sample pressing.

Source
“Although diamond is a very hard material, the ATR window is relatively thin (0.5 mm or less) and can crack under extreme pressure conditions.”
Safety

ZnSe ATR requires samples with pH between 5 and 9 to prevent crystal damage.

Prevents chemical attack on the ZnSe crystal and metal mounting plate.

Source
“Only measure samples with a pH between 5 and 9.”
Safety

Do not break the spectrometer seal to clean interior surfaces; doing so voids warranty.

Preserves warranty and avoids introducing contaminants into the optical path.

Source
“Do not break the spectrometer seal and attempt to clean interior surfaces. Breaking the seal will void the warranty.”
Maintenance

Replace desiccant when humidity indicator reaches 30% or turns pink.

Maintains low humidity inside the spectrometer to protect optical components.

Source
“Changing the desiccant is recommended once the humidity level has reached 30% or the color indicator is pink.”
Maintenance

After opening the main unit, allow 30 minutes to 2 hours for warmup and equilibration before use.

Ensures stable performance after maintenance that exposes internal components.

Source
“After the main unit has been exposed and opened to the environment, allow the system to sufficiently warm up and equilibrate prior to use.”
Operation

Collect a background spectrum before every sample to avoid environmental effects.

Improves measurement accuracy by accounting for current system conditions.

Source
“it is recommended that the system is configured in the software to collect a background spectrum before every sample is analyzed.”
Conformité

The multireflection ZnSe ATR complies with ASTM D-7371 for biodiesel analysis.

Indicates suitability for regulated biodiesel testing under ASTM D-7371.

Source
“the accessory complies with the ASTM D-7371 Biodiesel method, making it ideal for this application.”