Induztr
Résumé technique Visionneuse de documents Modèles associés

Agilent One-Color Microarray-Based Exon Analysis (Low Input Quick Amp WT Labeling) Protocol

Sommaire

  1. Overview pp. 1-6
    Document purpose, what's new in version 2.0, and table of contents
  2. Overview pp. 7-9
    Procedural notes and safety information before starting
  3. Spec pp. 10-15
    Required equipment, reagents, optional software, and oligo microarray details
  4. Procedure pp. 16-23
    Spike mix preparation and sample preparation procedures
  5. Procedure pp. 24-26
    Prepare labeling reaction steps and notes
  6. Procedure pp. 27-30
    Purify labeled/amplified RNA and quantify cRNA
  7. Procedure pp. 31-36
    Prepare hybridization samples, assembly, and hybridization steps
  8. Procedure pp. 37-40
    Prewarm wash buffer and wash microarray slides
  9. Procedure pp. 44-48
    Scan slides and extract data using Feature Extraction software
  10. Procedure pp. 51-68
    Additional purification, thermocycler, quality assessment, and ozone prevention procedures
  11. Procedure pp. 69-71
    Normalizing Agilent one-color microarray data
  12. Spec pp. 72-75
    Kit contents, references, and supplemental user guides
  13. Overview pp. 76-80
    Layout, orientation, scanners, and array/sample tracking
  14. Ordering pp. 81-83
    Related microarray reagents and array ordering information
  15. p. 11 table, 6 rows
  16. p. 10 table, 4 rows
  17. p. 14 table, 9 rows
  18. p. 12 table, 19 rows
  19. p. 59 table, 4 rows
  20. p. 83 table, 3 rows
  21. p. 13 table, 9 rows
  22. p. 21 table, 3 rows
  23. p. 24 table, 3 rows
  24. p. 25 table, 5 rows

Table 5 Required Equipment

DescriptionVendor and part number
Agilent Microarray ScannerAgilent p/n G4900DA, G2565CA or G2565BA
Hybridization Chamber, stainlessAgilent p/n G2534A
Hybridization gasket slides 1 microarray/slide, 5 slides/boxAgilent p/n G2534-60003
Hybridization gasket slides 2 microarrays/slide, 5 slides/boxAgilent p/n G2534-60002
Hybridization gasket slides 4 microarrays/slide, 5 slides/boxAgilent p/n G2534-60011
Hybridization gasket slides 8 microarrays/slide, 5 slides/boxAgilent p/n G2534-60014

Go to www.agilent.com/genomics to see all available kit configurations.

Table 1 Catalog SurePrint G3 Microarrays - Human

Part NumberDescription
G4832AG3 Human Exon 4×180K Microarray Kit (3 slides)
G4857A-028679G3 Human Exon 4×180K Microarray (1 slide)
G4848AG3 Human Exon 2×400K Microarray Kit (3 slides)
G4856A-028680G3 Human Exon 2×400K Microarray (1 slide)

Table 7 Optional Equipment/Reagents

DescriptionVendor and part number
2100 BioanalyzerAgilent p/n G2939AA
RNA 6000 Nano Assay Kit (RNA Series II Kit)Agilent p/n 5067-1511
Stabilization and Drying Solution*Agilent p/n 5185-5979
Ozone-Barrier Slide Cover*Agilent p/n G2505-60550
Absolutely RNA Nanoprep KitAgilent p/n 400753
slide boxCorning p/n 07201629
acetonitrileSigma p/n 271004-1L
sulfolaneSigma p/n T22209
thermal cyclerPCR 96-well plate or 0.2 mL PCR tubes

*Recommended when processing microarrays in high ozone environment.

Table 5 Required Equipment (continued)

DescriptionVendor and part number
Hybridization oven; temperature set at 65°CAgilent p/n G2545A
Hybridization oven rotator for Agilent Microarray Hybridization ChambersAgilent p/n G2530-60029
nuclease-free 1.5 mL microfuge tubeAmbion p/n 12400 or equivalent
magnetic stir bar (×2)Corning p/n 401435 or equivalent
magnetic stir plate (×2)Corning p/n 6795-410 or equivalent
circulating water baths or heat blocks set to 37°C, 40°C, 60°C, 65°C, 70°C, and 80°C,Corning p/n 6795-420 or equivalent
microcentrifugeEppendorf p/n 5417R or equivalent
sterile storage bottleNalgene 455-1000 or equivalent
spectrophotometerNanoDrop p/n ND-1000 UV-VIS or equivalent
micropipettorPipetman P-10, P-20, P-200, P-1000 or equivalent
slide-staining dish, with slide rack (×3)Thermo Shandon p/n 121 or equivalent
vacuum concentratorThermo Scientific p/n DNA 120-115 or equivalent
clean forceps
ice bucket
powder-free gloves
sterile, nuclease-free aerosol barrier pipette tips
vortex mixer
timer
nitrogen purge box for slide storage

Table 21 Analyzer and Kits

DescriptionVendor and part number
2100 BioanalyzerAgilent p/n G2938C or G2939A
RNA 6000 Nano LabChip KitAgilent p/n 5067-1511
RNA 6000 Pico LabChip KitAgilent p/n 5067-1513
SpectrophotometerNanoDrop p/n ND-1000 or equivalent

Related Microarray Reagents

DescriptionVendor and part number
Universal Human Reference RNAAgilent p/n 740000
Universal Mouse Reference RNAAgilent p/n 740100
Universal Rat Reference RNAAgilent p/n 740200

Table 6 Required Reagents

DescriptionVendor and part or catalog number
Low Input Quick Amp WT Labeling Kit, One-ColorAgilent p/n 5190-2943
RNA Spike-In Kit, One-ColorAgilent p/n 5188-5282
Gene Expression Hybridization KitAgilent p/n 5188-5242
Gene Expression Wash Buffer KitAgilent p/n 5188-5327
DNase/RNase-free distilled waterInvitrogen p/n 10977-015
RNeasy Mini KitQiagen p/n 74104
ethanol (95% to 100% molecular biology grade)Sigma-Aldrich p/n E7023-6×500ML
Milli-Q water or equivalent
isopropyl alcohol (molecular biology grade)

Table 10 Dilutions of Spike Mix for Cyanine 3-labeling

Starting Amount of RNA Total RNA (ng)Serial Dilution FirstSerial Dilution SecondSerial Dilution ThirdSerial Dilution FourthSpike Mix Volume to be used in each labeling reaction (µL)
251:201:251:201:42
501:201:251:201:22
1001:201:251:202

Use RNase-free microfuge tubes and tips. Make sure you dispense at least 2 µL with a pipette to ensure accuracy.

Table 11 WT Primer MasterMix

ComponentVolume (µL) per reactionVolume (µL) per 5 reactionVolume (µL) per 10 reactions
WT Primer1510
Diluted Spike Mix21020
Total Volume31530

Table 12 cDNA Master Mix

ComponentVolume (µL) per reactionVolume (µL) per 5 reactionVolume (µL) per 10 reactions
5× First Strand Buffer (green cap)21020
0.1 M DTT (white cap)1510
10 mM dNTP Mix (green cap)0.52.55
Affinity Script RNase Block Mix (violet cap)1.2612
Total Volume4.723.547

Engineering Facts

14 facts · every claim linked to a page
Selection

The protocol requires the C scanner because Agilent Exon Microarrays are only available on SurePrint G3 formats.

Scanner compatibility must be confirmed before selecting the microarray platform.

Source
“The C scanner is required, as Agilent Exon Microarrays are available only on the Agilent SurePrint G3 microarray formats.”
Selection

The 1-pack microarray format requires 100 ng total RNA, while 2-, 4-, and 8-pack formats require 25 ng (50 ng recommended).

Sample input amount determines which slide format can be used.

Source
“The starting input requirement for the 2-pack, 4-pack and 8-pack microarrays is 25 ng of total RNA (50 ng is recommended).”
Operation

The recommended specific activity for hybridization is 15 pmol cyanine 3 per microgram cRNA.

Labeling yield must meet this specification for successful hybridization.

Source
“The recommended Specific Activity for hybridization is increased to 15 pmol cyanine 3 per μg cRNA.”
Installation

Microarray slides must be stored at room temperature in the dark under vacuum desiccation or nitrogen purge after opening the foil pouch.

Proper storage environment prevents slide degradation before use.

Source
“After breaking foil on microarray pouch, store microarray slides at room temperature (in the dark) under a vacuum dessicator or nitrogen purge box.”
Maintenance

The hybridization oven must be calibrated regularly to ensure accuracy of collected data.

Oven calibration is a scheduled maintenance task affecting data quality.

Source
“You must calibrate the hybridization oven regularly for accuracy of the collected data.”
Operation

Hybridization sample volumes differ by slide format: 490 µL for 1-pack, 240 µL for 2-pack, 100 µL for 4-pack, and 40 µL for 8-pack.

Correct hybridization volume must be matched to the slide format to avoid leakage or insufficient coverage.

Source
“Volume to Hybridize 490 µL 240 µL 100 µL 40 µL”
Operation

Adding 0.005% Triton X-102 to wash buffers reduces the possibility of array wash artifacts.

Wash buffer additive is needed to prevent artifacts that could compromise data.

Source
“The addition of 0.005% Triton X-102 (10%) to the Gene Expression wash buffers reduces the possibility of array wash artifacts.”
Safety

The microarray wash procedure must be performed where ozone levels are 50 ppb or less.

Laboratory ozone level must be verified before installing the wash workflow.

Source
“The microarray wash procedure for the Agilent one-color platform must be done in environments where ozone levels are 50 ppb or less.”
Operation

Slides should be scanned immediately after washing to minimize environmental oxidant impact on signal intensities.

Delay between wash and scan degrades signal and should be minimized in workflow planning.

Source
“Scan slides immediately to minimize the impact of environmental oxidants on signal intensities.”
Conformité

Feature Extraction version 10.7.3 or later is required to extract Gene Expression Exon microarray data.

Software version must be verified before data extraction can proceed.

Source
“Extraction of Gene Expression Exon microarrays is supported by Feature Extraction version 10.7.3 or later.”
Operation

The thermocycler protocol can slightly lower cRNA yield compared to water baths.

Choice of incubation equipment affects yield and should be considered during setup.

Source
“The use of a thermocycler can slightly lower the yield of cRNA when compared to the use of water baths.”
Performance

Labeled cRNA should show a broad band with majority signal between 200 and 2000 nucleotides; samples with distinct bands below 200 nucleotides should not proceed.

Quality acceptance criterion for cRNA before hybridization.

Source
“The majority of signal for amplified sample should fall into the size range from 200 to 2000 nucleotides.”
Safety

Ozone levels above 50 ppb can significantly affect Cy3 signal and compromise microarray performance.

Environmental ozone control is required to protect dye signal integrity.

Source
“Beyond this level, ozone can significantly affect Cy3 signal and compromise microarray performance.”
Operation

Never allow the microarray surface to dry out during hybridization and washing steps.

Drying causes irreparable damage and must be prevented during handling.

Source
“Never allow the microarray surface to dry out during the hybridization process and washing steps.”