Selection
The bath covers a wide operating temperature range from -40 to 80 °C, suitable for both cooling and heating applications.
Determines whether the unit can meet the process temperature window without additional equipment.
Source
“Operating temperature range -40...80 °C”
Performance
Cooling capacity drops significantly at lower temperatures, so the required heat load must be checked against the capacity curve.
Ensures the selected unit can remove the process heat at the actual operating temperature.
Source
“Cooling power at 20°C 1,2 kW at 0°C 0,9 kW at -20°C 0,35 kW at -30°C 0,2 kW”
Installation
The unit requires a 230 V single-phase supply with up to 13 A current draw and a 16 A fuse.
Electrical infrastructure and protection must be sized accordingly before installation.
Source
“Power supply requirement 230V 1~ 50/60Hz max. current 13 A Fuse 16 A”
Installation
The air-cooled refrigeration machine needs sufficient ambient temperature between 5 and 40 °C and space for heat rejection.
Ambient conditions affect cooling performance and must be respected during site planning.
Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Safety
The refrigerant R-452A is classified A1 with GHS hazard H280, and the system contains 1.44 kg with a CO2 equivalent of 3.1 t.
Relevant for refrigerant handling, leak detection, and environmental reporting obligations.
Source
“Refrigerant (ASHRAE, GHS) R-452A (A1, H280) ... Refrigerant quantity 1,44 kg CO2 equivalent 3,1 t”
Safety
The unit is classified safety class III / FL according to DIN 12876 and includes adjustable overtemperature protection.
Confirms suitability for flammable liquids and the presence of critical safety cut-off.
Source
“Safety classification III / FL ... With adjustable overtemperature protection according to DIN 12876.”
Installation
The degree of protection is IP20, so the unit must be installed in a dry, protected environment.
Prevents ingress of water and dust that could damage electronics or cause hazards.
Source
“Degree of Protection IP20”
Conformité
Voltage tolerance is ±5% with simultaneous frequency tolerance of ±2%, but combined worst-case conditions are not allowed.
Ensures the power supply quality meets the unit's requirements and avoids invalid operating conditions.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2% ... -5% voltage and +2% frequency -> not allowed!”
Installation
Single-phase devices are supplied with a power cable and country-specific plug, which must be specified at ordering.
Avoids delays and ensures correct plug type for the installation site.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Conformité
The equipment is classified as Class A, Group 1 for electromagnetic disturbance according to EN55011.
Class A equipment may require additional EMC measures in residential or sensitive environments.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Operation
The controller supports external Pt100 sensor connection and digital interfaces Ethernet, USB, and RS232 for remote control and data logging.
Enables integration into process control systems and external temperature monitoring.
Source
“Sensor external connection Pt100 ... Interface digital Ethernet, USB (Host u. Device), RS232”