CC-215B Refrigerated/Heating Circulator

Sommaire

  1. Overview p. 1
    Product introduction and main features of the CC-215B circulator
  2. Spec p. 2
    Technical data according to DIN 12876 for the unit
  3. p. 1 table, 35 rows

Technical data according to DIN 12876

ParameterValue
Order-No.CC-215B
Operating temperature range25...200 °C
with water cooling20...200 °C
with refrigerator-30...200 °C
Temperature stability at 70°C0,02 K
temperature set point / display5,7" colour Touchscreen
Absolute accuracysetup for calibration
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Heating power at 208V1,6 kW
Heating power at 200V1,5 kW
Pressure pump max. delivery27 l/min
max. delivery pressure0,7 bar
Suction pumpyes
max. delivery (suction)22 l/min
max. delivery pressure (suction)0,4 bar
Pump connenction (optional)M16x1 male
Bath volume15 l
min. filling capacity11,3 l
Width bath opening WxD/ bath depth290x152/ 200 mm
Overall dimensions WxDxH **350x375x425 mm
Net weight12 kg
Power supply requirement200-240V 1~/2~ 50/60Hz
max. current10 A
min. Fuse10A
max. Fuse16A
Pressure equipment categoryN/A
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C

** Overall dimensions WxDxH

Engineering Facts

13 facts · every claim linked to a page
Selection

The operating temperature range depends on the cooling method: 25–200 °C without cooling, 20–200 °C with water cooling, and -30–200 °C with a refrigerator.

Determines whether the bath can reach sub-ambient or near-ambient setpoints and what cooling accessory must be specified.

Source
“Operating temperature range 25...200 °C with water cooling 20...200 °C with refrigerator -30...200 °C”
Installation

The unit accepts 200–240 V single- or two-phase supply at 50/60 Hz and draws up to 10 A, so branch protection must be sized between 10 A and 16 A.

Needed to size the electrical supply, fusing, and confirm the site can deliver the required current.

Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 10 A min. Fuse 10A max. Fuse 16A”
Installation

Voltage and frequency tolerances are ±10% only if they do not run in opposite directions, so a -10% voltage combined with +3% frequency is not acceptable.

Prevents commissioning on a supply whose combined voltage/frequency deviation would exceed the allowed envelope.

Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Conformité

An external branch circuit protection device according to UL 489 is required for this equipment.

Mandates a specific UL-listed breaker type in the installation, affecting panel design and compliance sign-off.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Single- and two-phase devices are delivered with a power cable and country-specific plug that must be specified at order time.

The correct plug must be ordered up front; otherwise the unit arrives without a usable mains connection.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Operation

The bath bridge has a hole for a cooling probe, allowing an immersion cooler such as TC45–TC100E to be fitted.

Enables adding external cooling for low-temperature work without changing the bath hardware.

Source
“Bath bridge with hole for cooling probe (e.g. for immersion cooler TC45-TC100E).”
Operation

When a Huber pump adapter is used, polyglycol must not be used as the heat transfer fluid.

Prevents selecting an incompatible heat transfer fluid that could damage the pump adapter or compromise safety.

Source
“When using Huber pump adapter: Polyglycol is not permissible to be used as a heat transfer fluid”
Performance

The pressure pump delivers up to 27 l/min at 0.7 bar, while the suction pump delivers up to 22 l/min at 0.4 bar.

Used to check whether the pump can supply the required flow and head for external circuits or connected equipment.

Source
“max. delivery 27 l/min max. delivery pressure 0,7 bar”
Selection

The bath holds 15 l total with a minimum filling capacity of 11.3 l, and the bath opening is 290×152 mm with 200 mm depth.

Determines whether sample vessels fit and how much heat transfer fluid must be procured and maintained.

Source
“Bath volume 15 l min. filling capacity 11,3 l”
Installation

The unit is rated IP20 and must be operated in ambient temperatures between 5 °C and 40 °C.

Sets the allowable room environment and confirms the enclosure is not protected against water ingress.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Safety

The bath includes adjustable overtemperature protection according to DIN 12876 and is classified III / FL.

Confirms a safety cut-out is present and must be configured to protect the fluid and samples.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Operation

The Pilot ONE controller offers USB, Ethernet, and RS232 interfaces, and the USB port can connect the system to a PC or notebook for remote control or data transmission with Spy software.

Supports integration into a lab network or data logging setup without additional hardware.

Source
“The USB port allows connection of the system to a PC or notebook.”
Maintenance

A drain valve with cap is available as an accessory for emptying the bath.

Simplifies fluid changes and maintenance; should be ordered with the unit if draining is required.

Source
“Drain valve with cap #6839”