CC-300BX Immersion Circulator — Technical Data Sheet

Sommaire

  1. Overview p. 1
    Product introduction and general description of the CC-300BX unit
  2. Spec p. 2
    Technical data per DIN 12876 including performance and dimensions
  3. p. 1 table, 34 rows

Technical data according to DIN 12876 — CC-300BX (Order-No.: 2046.0001.01)

CharacteristicValue
Operating temperature range28...300 °C
with water cooling20...300 °C
with refrigerator-20...300 °C
Temperature stability at 70°C0,02 K
temperature set point / display5,7" colour Touchscreen
Absolute accuracysetup for calibration
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V3,5 kW
Heating power at 230V3,5 kW
Heating power at 220V3,3 kW
Heating power at 208V3,2 kW
Heating power at 200V3 kW
Pressure pumpyes
max. delivery25 l/min
Delivery pressure (head)0,7 bar
Suction pumpyes
max. delivery (suction)18,5 l/min
Delivery suction pressure (head)0,4 bar
Pump connectionM16x1 male
max. permissible kin. viscosity50 mm²/s
Immersion depth175 mm
Overall dimensions WxDxH **345x190x392 mm
Net weight13 kg
Power supply requirement200-240V 1~/2~ 50/60Hz
max. current16 A
Fuse20 A
Pressure equipment categoryN/A
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.S436620

** Overall dimensions WxDxH

Engineering Facts

14 facts · every claim linked to a page
Selection

The operating temperature range depends on the cooling method chosen, so the required range must be matched to water cooling or a refrigerator.

Determines whether the base unit alone, water cooling, or an active refrigerator is needed for the process temperature.

Source
“Operating temperature range 28...300 °C with water cooling 20...300 °C with refrigerator -20...300 °C”
Installation

The electrical supply must be 200-240 V single or two-phase at 50/60 Hz and the branch circuit must be rated for the 16 A maximum current.

Defines the site power configuration and breaker/cabling capacity required before installation.

Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 16 A”
Safety

External branch circuit protection according to UL 489 is mandatory for this equipment.

The installer must provide a listed UL 489 protective device; omitting it is a compliance and safety defect.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Voltage and frequency tolerances are each ±10% but must not deviate in opposite directions simultaneously.

Prevents selecting a supply point where combined voltage and frequency excursions exceed the allowed envelope.

Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Installation

Single- and two-phase devices are delivered with a power cable and country-specific plug that must be specified at order time.

The plug type must be confirmed on the order, otherwise the unit may not connect to the site receptacle.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Operation

The ambient temperature at the installation site must stay between 5 °C and 40 °C.

Room conditions outside this window can affect performance and must be checked during site planning.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation

The housing is IP20, so the unit must be installed where it is protected from water and dust ingress.

Restricts placement to dry indoor locations and rules out washdown or wet environments.

Source
“Degree of Protection IP20”
Selection

The pump connection is M16x1 male, which dictates the mating hose or adapter to be ordered.

Ensures the correct external circuit connections and hoses are specified with the unit.

Source
“Pump connection M16x1 male”
Performance

The pressure pump delivers up to 25 l/min at 0.7 bar and the suction pump up to 18.5 l/min at 0.4 bar.

Used to verify that the external loop flow resistance and required flow rate are within pump capability.

Source
“max. delivery 25 l/min Delivery pressure (head) 0,7 bar Suction pump yes max. delivery (suction) 18,5 l/min Delivery suction pressure (head) 0,4 bar”
Selection

The maximum permissible kinematic viscosity of the thermofluid is 50 mm²/s.

Limits the choice of heat transfer fluid, especially at low temperatures where viscosity rises.

Source
“max. permissible kin. viscosity 50 mm²/s”
Installation

The immersion depth is 175 mm and the telescopic arm extends up to 884 mm, which sets the bath mounting geometry.

Confirms the unit fits the bath depth and that the telescopic bridge can span the bath opening.

Source
“Immersion depth 175 mm”
Conformité

The unit is classified Class A, Group 1 for electromagnetic disturbance to EN55011.

Class A equipment is intended for non-domestic environments and may require EMC measures in residential areas.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Safety

The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III/FL.

The overtemperature cut-out must be set for the application to protect the bath and medium from overheating.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Maintenance

The 4-year warranty requires registration.

Registration must be completed to preserve warranty coverage on the equipment.

Source
“4-year warranty - registration required.”