CC-505wl Immersion Circulator Datasheet

Sommaire

  1. Overview p. 1
    Product introduction and key features of the CC-505wl circulator
  2. Spec pp. 1-2
    Technical data according to DIN 12876 and performance parameters
  3. p. 2 table, 11 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
Height of bath opening565 mm
Overall dimensions WxDxH **410x480x764 mm
Net weight62 kg
Power supply requirement208V 2~ 60Hz
max. current9,5 A
min. Fuse2x10A
max. Fuse2x16A
Degree of ProtectionIP20
max. ambient temperature40 °C
min. ambient temperature5 °C
from Serial-No.:1.0/20

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

CharacteristicValue
Operating temperature range-50...200 °C
Temperature stability at -10°C0,02 K
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power1,6 kW
Cooling power at 100°C1,2 kW
Cooling power at 20°C1,2 kW
Cooling power at 0°C1 kW
Cooling power at -20°C0,6 kW
Cooling power at -40°C0,15 kW
Refrigeration machineair- and water-cooled, CFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
max. delivery25 l/min
max. delivery pressure0,7 bar
Suction pumpyes
max. delivery (suction)18,5 l/min
max. delivery pressure (suction)0,4 bar
Pump connectionM16x1 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 20°C78 l/h
Consumption at water 15°C, flow 0°C72 l/h
Consumption at water 15°C, flow -20°C54 l/h
Consumption at water 15°C, flow -40°C42 l/h
min. cooling water differential pressure3 bar
max. cooling water pressure6 bar
Bath volume5 l
min. filling capacity4 l
Width bath opening WxD / bath depth120 x 110 / 150 mm

Order-No.: CC-505wl; 2044.0004.01

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit covers a very wide temperature range from -50 to 200 °C, so it can serve both deep-freeze and high-temperature applications with one device.

Defines whether the circulator can meet the process temperature window before any other sizing step.

Source
“Operating temperature range -50...200 °C”
Performance

Cooling capacity collapses from 1.2 kW at 20 °C to only 0.15 kW at -40 °C, so low-temperature heat loads must be sized against the cold-end figure.

Prevents oversizing the process load at the cold end of the operating range.

Source
“at -40°C 0,15 kW”
Installation

The cooling water circuit requires at least 3 bar differential pressure and tolerates no more than 6 bar, which constrains the facility water supply design.

Facility water pressure must be verified before installation; too low or too high pressure will impair or damage operation.

Source
“min. cooling water differential pressure 3 bar”
Installation

Cooling water consumption is specified per operating point (78 l/h at 20 °C down to 42 l/h at -40 °C), so the facility must supply adequate flow at the coldest duty.

Needed to size the cooling water supply line and check available facility flow.

Source
“Consumption at water 15°C, flow -40°C 42 l/h”
Installation

The unit is designed for cooling water between 5 °C and 20 °C, and pipes must be insulated if the water is below dew point to prevent condensation.

Defines acceptable facility water temperature and the need for insulation to avoid condensation damage.

Source
“The temperature control unit is designed for cooling water temperatures between 5 °C and 20 °C.”
Performance

Rated output data are only valid at 20 °C room temperature, 15 °C cooling water inlet and 3 bar differential pressure, so real performance will differ outside these conditions.

Prevents expecting catalogue cooling capacity when site conditions deviate from the reference point.

Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 3 bar differential pressure between cooling water inlet and -”
Installation

The electrical supply is 208 V two-phase 60 Hz with 9.5 A max current and 2x10 A to 2x16 A fusing, which is a non-standard supply that must be planned for.

Confirms site power availability and correct fuse/breaker sizing before ordering.

Source
“Power supply requirement 208V 2~ 60Hz”
Conformité

Voltage and frequency tolerances are coupled: ±5 % voltage with ±2 % frequency, but a -5 % voltage combined with +2 % frequency is not allowed.

Governs acceptance of the site power quality and protects the compressor from damaging supply combinations.

Source
“-5% voltage and +2% frequency -> not allowed!”
Safety

The refrigerant R-452A is classified A1 but carries GHS hazard H280 and a GWP of 2141, so refrigerant handling and environmental reporting obligations apply.

Affects refrigerant safety measures, leak detection and greenhouse-gas reporting for the installation.

Source
“R-452A (A1, H280)”
Installation

The unit is IP20 and must be operated between 5 °C and 40 °C ambient, so it cannot be placed in a washdown or unconditioned hot area.

Sets the room conditions and enclosure protection required at the installation site.

Source
“max. ambient temperature 40 °C”
Installation

Overall dimensions are 410x480x764 mm at 62 kg net, and space requirements must be respected per the operating conditions document.

Needed for bench/floor space planning and access clearance around the unit.

Source
“Overall dimensions WxDxH ** 410x480x764 mm”
Operation

The pressure/suction pump delivers up to 25 l/min at 0.7 bar pressure and 18.5 l/min at 0.4 bar suction, with a maximum kinematic viscosity of 50 mm²/s.

Determines whether the external loop and heat transfer fluid viscosity are compatible with the pump.

Source
“max. permissible kin. viscosity 50 mm²/s”
Operation

The bath holds 5 l with a minimum filling of 4 l, and the bath opening is 120 x 110 mm at 150 mm depth.

Sets the fluid charge and confirms whether the immersion depth suits the application.

Source
“min. filling capacity 4 l”
Conformité

The equipment is EMC class A group 1, which is industrial-grade emission classification and may require separation from sensitive electronics.

Affects placement relative to sensitive instrumentation and EMC compliance of the installation.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”