CC-525w Circulating Bath Datasheet

Sommaire

  1. Overview p. 1
    Introduces the CC-525w circulating bath and its main features
  2. Spec pp. 1-2
    Lists operating parameters, dimensions, and performance data per DIN 12876
  3. p. 2 table, 12 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
Width bath opening WxD / bath depth270 x 150 / 200 mm
Height of bath opening903 mm
Overall dimensions WxDxH **539x629x1102 mm
Net weight163 kg
Power supply requirement (3 phase)400V 3~N 50Hz
max. current (3 Phase)13 A
Fuse (3 phase)3x16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
max. ambient temperature40 °C
min. ambient temperature5 °C
from Serial-No.:S608247

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ParameterValue
Operating temperature range-55...200 °C
Temperature stability at -10°C0,02 K
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power3 kW
Cooling power at 100°C7 kW
Cooling power at 20°C7 kW
Cooling power at 0°C5 kW
Cooling power at -20°C3 kW
Cooling power at -40°C1,5 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity1,9 kg
CO2 equivalent4,1 t
Pressure pumpyes
max. delivery25 l/min
max. delivery pressure0,7 bar
Suction pumpyes
max. delivery (suction)18,5 l/min
max. delivery pressure (suction)0,4 bar
Pump connectionM16x1 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 0°C205 l/h
min. cooling water differential pressure3 bar
max. cooling water pressure6 bar
Bath volume17 l
min. filling capacity15 l
Bath capacity with displacement rack10 l

Order-No.: CC-525w; 2023.0001.01

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a three-phase 400 V supply with a 3x16 A fuse, so the site must provide a suitable three-phase feed.

Determines electrical infrastructure and breaker sizing before installation.

Source
“Power supply requirement (3 phase) 400V 3~N 50Hz”
Installation

The cooling water circuit needs a differential pressure of at least 3 bar and must not exceed 6 bar.

Ensures the water supply can achieve the required cooling performance without overpressurizing the unit.

Source
“min. cooling water differential pressure 3 bar”
Installation

Cooling water inlet temperature must stay between 5 and 20 °C, and pipes must be insulated if below the condensation point.

Prevents condensation and ensures rated cooling power is achievable.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power drops as cooling water temperature rises, so actual capacity may be lower than the rated values.

Affects sizing of the unit for the required heat load at site conditions.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected”
Installation

The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water quality should be used.

Guides water treatment and material compatibility to avoid corrosion.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.

Defines acceptable power quality for reliable operation.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

Three-phase devices under 63 A are delivered with cable but without a plug, so a plug must be sourced separately.

Affects electrical installation planning and procurement.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety

The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III / FL.

Confirms built-in safety protection required for unattended operation.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Operation

The pump delivers up to 25 l/min at 0.7 bar pressure and 18.5 l/min suction at 0.4 bar, with M16x1 connections.

Determines whether the pump can serve the external circuit's flow and pressure needs.

Source
“max. delivery pressure 0,7 bar”
Operation

Maximum permissible kinematic viscosity of the medium is 50 mm²/s.

Limits which heat transfer fluids can be used with the pump.

Source
“max. permissible kin. viscosity 50 mm²/s”
Selection

The bath holds 17 l, with a minimum filling of 15 l and 10 l when a displacement rack is used.

Affects fluid volume, expansion volume and displacement accessory choice.

Source
“Bath capacity with displacement rack 10 l”
Installation

The unit is IP20 and must operate between 5 and 40 °C ambient, so it needs a protected indoor location.

Defines environmental limits and enclosure protection for siting.

Source
“max. ambient temperature 40 °C”
Conformité

The refrigerant R-452A has a GWP of 2141 with 1.9 kg charge, giving 4.1 t CO2 equivalent subject to F-gas rules.

Relevant for refrigerant leak monitoring and environmental compliance obligations.

Source
“CO2 equivalent 4,1 t”
Conformité

The equipment is classified Class A, Group 1 for electromagnetic disturbance to EN55011.

Class A equipment may require EMC measures in residential environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”