The cooling water circuit must be supplied at 3–6 bar differential pressure, so the building water system must be checked for adequate pressure.
Insufficient cooling water pressure will prevent the water-cooled refrigeration machine from operating correctly.
Source
“min. cooling water differential pressure 3 bar”
Installation
Cooling water intake must be between 5 °C and 20 °C, and pipes must be insulated if water can drop below the condensation point.
Defines the permissible cooling water supply conditions and required pipe insulation at the installation site.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops significantly at low temperatures, from 1.2 kW at 20 °C to only 0.14 kW at -80 °C, which must be considered for load sizing.
The available cooling capacity at the target low temperature determines whether the unit can handle the process heat load.
Source
“at -80°C 0,14 kW”
Installation
The unit is water-cooled and uses R-452A and R-23 refrigerants with a combined CO2 equivalent of about 10.1 t, which affects refrigerant reporting obligations.
Refrigerant charge and GWP determine leak-check and reporting requirements under F-gas regulations.
Source
“CO2 equivalent 1,7 t”
Operation
The pressure pump delivers up to 25 l/min at 0.7 bar and the suction pump up to 18.5 l/min, so external circuit pressure drop must be checked.
The pump head limits the permissible pressure drop in the external cooling circuit.
Source
“max. delivery pressure 0,7 bar”
Operation
The maximum permissible kinematic viscosity of the circulating fluid is 50 mm²/s, which constrains the choice of heat transfer medium at low temperatures.
Selecting a fluid with too high viscosity at operating temperature will overload the pump.
Source
“max. permissible kin. viscosity 50 mm²/s”
Installation
The bath requires a minimum filling capacity of 15 l within a 17 l bath volume, so the fluid charge must be planned accordingly.
Determines the minimum heat transfer fluid quantity needed for safe operation.
Source
“min. filling capacity 15 l”
Installation
The unit weighs 150 kg and measures 539x629x1102 mm, so floor loading and access clearance must be verified.
Needed for transport planning, doorway clearance and floor load assessment.
Source
“Overall dimensions WxDxH ** 539x629x1102 mm”
Operation
The unit is rated IP20 and must be operated between 5 °C and 40 °C ambient, so it cannot be installed in a wet or hot environment.
Defines the permissible ambient conditions and protection level at the installation location.
Source
“max. ambient temperature 40 °C”
Conformité
Voltage and frequency tolerances are ±5% and ±2% respectively, but the combination of -5% voltage with +2% frequency is not allowed.
The site power quality must be checked against these combined tolerances to avoid damage.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
Three-phase devices below 63 A are delivered with a cable but without a plug, so a suitable plug must be provided on site.
Affects the electrical installation scope and materials needed at commissioning.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety
The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III/FL, relevant for flammable or sensitive media.
Confirms the safety concept required when selecting the unit for a given application.
Source
“With adjustable overtemperature protection according to DIN 12876.”