Induztr
Résumé technique Visionneuse de documents Modèles associés

CC-E xd Immersion Circulator

Sommaire

  1. Overview p. 1
    Product introduction and model identification for CC-E xd
  2. Spec p. 2
    Technical data according to DIN 12876
  3. p. 1 table, 34 rows

Technical data according to DIN 12876

CharacteristicValue
Operating temperature range25...200 °C
with water cooling20...200 °C
with refrigerator-30...200 °C
Temperature stability at 70°C0,01 K
temperature set point / display5,7" colour Touchscreen
Absolute accuracysetup for calibration
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Heating power at 208V1,6 kW
Heating power at 200V1,5 kW
Pressure pumpyes
max. delivery22 l/min
max. delivery pressure0,4 bar
Suction pumpyes
max. delivery (suction)17 l/min
max. delivery pressure (suction)0,25 bar
Pump connenction (optional)M16x1 male
Immersion depth195 mm
Overall dimensions WxDxH **132x159x360 mm
Net weight5 kg
Power supply requirement200-240V 1~/2~ 50/60Hz
Power input2,1 kW
max. current10 A
min. Fuse10A
max. Fuse16A
Pressure equipment categoryN/A
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C

** Overall dimensions WxDxH

Engineering Facts

12 facts · every claim linked to a page
Selection

The operating temperature range depends on the cooling method chosen, so the cooling option must be specified at selection.

Determines whether the application temperature can be reached with water cooling, refrigerator, or neither.

Source
“Operating temperature range 25...200 °C with water cooling 20...200 °C with refrigerator -30...200 °C”
Installation

The unit requires a 200-240 V single or two-phase supply with 50/60 Hz and a branch fuse between 10 A and 16 A.

Electrical supply and protection must be planned before installation.

Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz”
Installation

Voltage and frequency tolerances are only acceptable if they do not run in opposite directions, so supply quality must be checked.

Prevents commissioning on a supply that would violate the allowed tolerance combination.

Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Conformité

External branch circuit protection according to UL 489 is required, which affects the electrical panel design.

Must be included in the installation to meet the stated requirement.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Single and two-phase devices are delivered with a power cable and country-specific plug, which must be specified when ordering.

Avoids delivery of the wrong plug type for the installation site.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Operation

Polyglycol must not be used as heat transfer fluid when the Huber pump adapter is used.

Prevents selection of an incompatible heat transfer fluid that could damage the pump circuit.

Source
“When using Huber pump adapter: Polyglycol is not permissible to be used as a heat transfer fluid”
Installation

The ambient temperature during operation must stay between 5 °C and 40 °C.

Defines the allowable room conditions at the installation location.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Safety

The unit has IP20 protection, so it must be installed where water ingress and dust are controlled.

Affects placement and enclosure requirements at the installation site.

Source
“Degree of Protection IP20”
Performance

The pressure pump delivers up to 22 l/min at 0.4 bar and the suction pump up to 17 l/min at 0.25 bar.

Used to verify that the external circuit flow and head requirements are met.

Source
“max. delivery 22 l/min max. delivery pressure 0,4 bar Suction pump yes max. delivery (suction) 17 l/min max. delivery pressure (suction) 0,25 bar”
Selection

Heating power depends on the supply voltage, ranging from 1.5 kW at 200 V to 2.1 kW at 240 V.

Heating capacity at the actual site voltage must be checked against the process heat demand.

Source
“Heating power at 240V 2,1 kW Heating power at 230V 2 kW Heating power at 220V 1,8 kW Heating power at 208V 1,6 kW Heating power at 200V 1,5 kW”
Installation

The immersion depth is 195 mm, which constrains how the thermostat fits into the bath or vessel.

Must match the bath depth and mounting arrangement.

Source
“Immersion depth 195 mm”
Conformité

The unit is classified as Class A, Group 1 for electromagnetic disturbance, which has implications for the installation environment.

Class A equipment may require EMC measures depending on the installation location.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”