CC-K6s Recirculating Chiller Datasheet

Sommaire

  1. Overview p. 1
    Introduces the CC-K6s chiller and its main features
  2. Spec pp. 1-2
    Performance data, dimensions, and operating parameters per DIN 12876
  3. p. 2 table, 8 rows
  4. p. 1 table, 32 rows

Technical data according to DIN 12876

ParameterValue
max. current immersion thermostat15 A
min. Fuse10A
max. Fuse15A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S449202 1.2/20

Technical data according to DIN 12876

ParameterValue
Operating temperature range-25...200 °C
Temperature stability at 70°C0,02 K
temperature set point / display5,7" colour Touchscreen
Absolute accuracysetup for calibration
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power1,5 kW
Cooling power at 20°C0,26 kW
Cooling power at 0°C0,21 kW
Cooling power at -10°C0,15 kW
Cooling power at -20°C0,05 kW
Refrigeration machineair-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-290 (A3, H220)
Global Warming Potential (GWP)0,02
Refrigerant quantity0,047 kg
CO2 equivalent0,0 t
Pressure / Suction pump max. delivery27 l/min
max. delivery pressure0,7 bar
max. delivery (suction)22 l/min
max. delivery pressure (suction)0,4 bar
Pump connectionM16x1 male
Bath volume4,5 l
Width bath opening WxD140x120 mm
Bath depth150 mm
Height of bath opening376 mm
Overall dimensions WxDxH **210x400x546 mm
Net weight25 kg
sound pressure level +/- 4 dB(A)57 dB(A)
Power supply requirement110-120V 1~ 50/60Hz
max. current refrigerated bath3,2 A

** см. сноску в оригинале (в тексте страницы пояснение отсутствует)

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit is classified as safety class III / FL per DIN 12876, which determines the required protective measures and suitability for flammable liquids.

Safety class III/FL dictates whether the bath can be used with flammable heat transfer fluids and what over-temperature protection is mandatory.

Source
“Safety classification III / FL”
Safety

The refrigerant is R-290 (propane, A3, H220), a flammable natural refrigerant, so ignition sources and ventilation must be considered during installation.

A3 flammable refrigerant affects room ventilation, ignition-source restrictions, and service procedures.

Source
“Refrigerant (ASHRAE, GHS) R-290 (A3, H220)”
Installation

The unit requires a 110-120 V single-phase 50/60 Hz supply with a maximum current of 3.2 A for the refrigerated bath.

Electrical supply must match this voltage and frequency; a different mains configuration requires a different model or transformer.

Source
“Power supply requirement 110-120V 1~ 50/60Hz”
Installation

The immersion thermostat branch can draw up to 15 A and the recommended fuse is 10 A minimum, 15 A maximum.

Circuit protection and wiring must be sized for up to 15 A on the thermostat circuit to avoid nuisance tripping or undersized cabling.

Source
“max. current immersion thermostat 15 A”
Installation

Voltage and frequency tolerances are ±5% and ±2% respectively, but the combination of -5% voltage with +2% frequency is explicitly not allowed.

Supply quality must be checked; some apparently acceptable tolerance combinations are prohibited and can damage the compressor.

Source
“-5% voltage and +2% frequency -> not allowed!”
Operation

Cooling capacity drops as ambient temperature rises above 20 °C, so performance ratings are only valid at 20 °C room temperature.

Sizing must account for actual ambient conditions; a unit adequate at 20 °C may be undersized in a warmer room.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

The permissible ambient temperature range is 5 °C to 40 °C, and the unit has IP20 protection, so it must be kept in a dry, protected location.

IP20 means no water protection; the installation site must avoid moisture and the ambient temperature must stay within 5–40 °C.

Source
“Degree of Protection IP20”
Installation

The unit is air-cooled and requires respect of space requirements around it for adequate airflow.

Insufficient clearance around an air-cooled condenser reduces cooling capacity and can cause overheating.

Source
“** Please respect space requirements. See operating conditions at www.huber-online.com”
Conformité

The equipment is EMC class A, group 1 per EN55011, so it is intended for industrial environments and may cause interference in residential areas.

Class A equipment may not be suitable for residential use and may require additional EMC measures in sensitive environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Single/two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering.

The correct plug type must be stated on the order; otherwise the unit may arrive with an incompatible mains connection.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Performance

The pressure/suction pump delivers up to 27 l/min at 0.7 bar pressure and 22 l/min suction at 0.4 bar, with M16x1 male connections.

External circuit pressure drop and hose sizing must be checked against these pump limits to avoid insufficient flow.

Source
“max. delivery 27 l/min”
Operation

The bath holds 4.5 l and the working opening is 140x120 mm with 150 mm depth, limiting the size of immersed vessels and samples.

Sample size and heat load must fit within the bath volume and opening dimensions.

Source
“Bath volume 4,5 l”
Maintenance

The unit uses only 0.047 kg of R-290 with a GWP of 0.02, giving a CO2 equivalent of 0.0 t.

The very small refrigerant charge simplifies leak reporting and F-gas obligations, but the flammable charge still requires qualified service.

Source
“Refrigerant quantity 0,047 kg”
Selection

The operating temperature range is -25 to 200 °C with 0.02 K stability at 70 °C, defining the application envelope.

The process temperature must lie within this range, and the stated stability is only valid under specified conditions.

Source
“Operating temperature range -25...200 °C”