DC32 Datasheet

Sommaire

  1. Overview p. 1
    Product introduction and key features of DC32
  2. Config. p. 1
    Available models and ordering options
  3. Spec p. 1
    Technical specifications table for DC32
  4. p. 1 table, 21 rows

Technical data according to DIN 12876

ParameterValue
Operating temperature range-30...50 °C
Safety classificationI / NFL
Cooling power at 15°C0,6 kW
Cooling power at 0°C0,47 kW
Cooling power at -20°C0,12 kW
Refrigeration machineair-cooled, natural
refrigerant
Refrigerant (ASHRAE, GHS)R-290 (A3, H220)
Global Warming Potential (GWP)0,02
Refrigerant quantity0,08 kg
fluid connectionM16x1 male
Overall dimensions WxDxH **280x340x465 mm
Net weight30 kg
Power supply requirement220-240V 1~/2~ 50/60Hz
max. current3 A
min. Fuse10A
max. Fuse16A
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.460283

** Please respect space requirements. See operating conditions at www.huber-online.com

Engineering Facts

13 facts · every claim linked to a page
Selection

The unit is rated for an operating temperature range of -30 to 50 °C, so it must be selected only for applications within these limits.

Defines the process temperature window the cooler can support.

Source
“Operating temperature range -30...50 °C”
Performance

Cooling capacity drops significantly at lower temperatures, from 0.6 kW at 15 °C to 0.12 kW at -20 °C, so sizing must use the capacity at the actual operating point.

Prevents undersizing when the application runs at low temperatures.

Source
“Cooling power at 15°C 0,6 kW at 0°C 0,47 kW at -20°C 0,12 kW”
Safety

The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, which requires appropriate safety measures and awareness during installation and service.

Flammable refrigerant classification affects site safety and service procedures.

Source
“Refrigerant (ASHRAE, GHS) R-290 (A3, H220)”
Installation

The fluid connection is M16x1 male, so mating hoses and fittings must match this thread.

Ensures correct hydraulic connection hardware is ordered.

Source
“fluid connection M16x1 male”
Installation

The power supply requirement is 220-240 V 1~/2~ 50/60 Hz with max current 3 A, so the electrical supply and protection must be sized accordingly.

Needed to plan the electrical connection and breaker rating.

Source
“Power supply requirement 220-240V 1~/2~ 50/60Hz max. current 3 A”
Installation

The minimum fuse is 10 A and maximum fuse is 16 A, so the upstream protection must fall within this range.

Avoids nuisance tripping or inadequate protection.

Source
“min. Fuse 10A max. Fuse 16A”
Installation

The degree of protection is IP20, so the unit must be installed in a dry indoor environment protected from water and dust ingress.

Restricts installation location and environmental exposure.

Source
“Degree of Protection IP20”
Operation

The ambient temperature must be between 5 °C and 40 °C, so the installation room must be kept within this range.

Ambient limits affect reliable operation and cooling performance.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance

Output data are valid for a room temperature of 20 °C, and higher ambient temperature may reduce cooling capacity, so capacity margins should account for actual room conditions.

Prevents overestimating capacity in warm installation rooms.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Conformité

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed, so supply quality must be verified.

Ensures the electrical supply stays within the allowed envelope.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Conformité

The equipment is classified as EN55011 Class A, Group 1 for disturbance, so it is intended for industrial electromagnetic environments rather than residential ones.

Affects EMC compatibility and suitability for the installation site.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Single/two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering, so the correct plug variant must be requested.

Prevents delivery with the wrong plug for the site.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Installation

Space requirements must be respected as stated in the operating conditions, so clearance around the unit must be planned before installation.

Insufficient clearance can impair air-cooled operation and access.

Source
“Please respect space requirements. See operating conditions at www.huber-online.com”