Grande Fleur

Sommaire

  1. Overview p. 1
    Product introduction and main features of Grande Fleur
  2. Spec p. 2
    Technical data according to DIN 12876
  3. p. 2 table, 16 rows
  4. p. 1 table, 30 rows

Technical data according to DIN 12876

ParameterValue
Pump connectionM24x1,5 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 15°C51 l/h
Consumption at water 15°C, flow 0°C49 l/h
Consumption at water 15°C, flow -20°C36 l/h
min. cooling water differential pressure2 bar
max. cooling water pressure6 bar
min. filling capacity1,5 l
Filling capacity expansion tank3 l
Overall dimensions WxDxH **295x530x570 mm
Power supply requirement208-240V 1~/2~ 50/60Hz
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/25

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ParameterValue
Operating temperature range-40...200 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power at 240V2 kW
Heating power at 230V1,8 kW
Heating power at 220V1,6 kW
Heating power at 208V1,5 kW
Cooling power with Thermooil at 200°C0,6 kW
Cooling power with Thermooil at 150°C0,6 kW
Cooling power with Thermooil at 100°C0,6 kW
Cooling power with Thermooil at 50°C0,6 kW
Cooling power with Thermooil at 20°C0,6 kW
Cooling power with Ethanol at 0°C0,6 kW
Cooling power with Ethanol at -20°C0,35 kW
Cooling power with Ethanol at -30°C0,2 kW
Cooling power with Ethanol at -40°C0,04 kW
Refrigeration machinewater-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-1270 (A3, H220)
Global Warming Potential (GWP)0
UN-numberUN 3358
Circulation pump: max. delivery47 l/min
Circulation pump: max. delivery pressure0,9 bar

Order-No.: 1041.0020.01; Grande Fleur w

Engineering Facts

13 facts · every claim linked to a page
Installation

The unit requires cooling water with a minimum differential pressure of 2 bar and a maximum pressure of 6 bar, so the site water supply must be sized accordingly.

To verify the available cooling water supply pressure matches the unit's hydraulic requirements before installation.

Source
“min. cooling water differential pressure 2 bar max. cooling water pressure 6 bar”
Installation

The cooling water intake temperature must be between 5°C and 20°C, and pipes must be insulated if the water temperature can fall below the dew point.

To ensure the site cooling water temperature is within the allowed range and to plan insulation to prevent condensation.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling capacity drops significantly at low temperatures with ethanol, from 0.6 kW at 0°C to only 0.04 kW at -40°C, so the process load must be matched to the actual operating temperature.

To correctly size the circulator for low-temperature duty where available cooling power is very limited.

Source
“at -40°C 0,04 kW”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

To select suitable cooling water quality and avoid corrosion or damage to wetted materials.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Conformité

Voltage and frequency tolerances are ±5% and ±2% respectively, but they must not be applied in opposite directions simultaneously.

To assess site power quality and avoid operating the unit outside its permitted supply envelope.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

The unit is IP20 and must be operated between 5°C and 40°C ambient temperature, so it needs a dry, protected indoor location.

To confirm the installation location meets environmental and ingress protection requirements.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation

The pump connection is M24x1.5 male and the cooling water connection is G1/2 male, so matching adapters and hoses must be ordered.

To prepare correct plumbing adapters and hoses before commissioning.

Source
“Pump connection M24x1,5 male”
Installation

The pump connections use bore shape Y (60°) per DIN 3863 and require ball socket fittings per DIN 3863 with sleeve nuts per DIN 3870.

To specify the correct fitting standard for leak-free pump connections.

Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Operation

The expansion tank is thermally passive, preventing HTF vapour formation and moisture absorption.

To understand why the unit maintains stable heat transfer fluid properties during operation.

Source
“No HTF vapour and no moisture absorption because the expansion tank is thermally passive.”
Selection

The unit is water-cooled with natural refrigerant R-1270 (A3, H220) having GWP 0, which affects refrigerant handling and safety classification.

To account for flammable refrigerant safety measures and environmental compliance in selection.

Source
“Refrigerant (ASHRAE, GHS) R-1270 (A3, H220) Global Warming Potential (GWP) 0”
Installation

The unit requires 208-240V single or two-phase 50/60Hz supply, and single/two-phase devices are delivered with a country-specific plug that must be specified at order.

To order the correct power cable/plug configuration and confirm site electrical supply.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Maintenance

The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s, limiting fluid selection at low temperatures.

To select a heat transfer fluid whose viscosity at the lowest operating temperature stays within the pump limit.

Source
“max. permissible kin. viscosity 50 mm²/s”
Safety

The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III/FL.

To confirm the safety concept meets the required protection class for the application.

Source
“With adjustable overtemperature protection according to DIN 12876.”