Product introduction and general description of the HB720 unit
Specp. 2
Technical data and performance parameters per DIN 12876
p. 1table, 20 rows
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
60...250 °C
Resolution of display
0,01 K
Temperature stability at 70°C
0,1 K
temperature set point / display
5,7" colour Touchscreen
Absolute accuracy
setup for calibration
Internal temperature sensor
Pt100
External sensor
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
II / FL
Heating power
72 kW
Circulation pump:
MK pump
max. delivery
200 l/min
max. delivery pressure
5,5 bar
Pump connection
M38x1,5 male
Overall dimensions WxDxH
800x1007x1607 mm
Degree of Protection
IP20
Power supply (3 Phase)
400V 3~ 50Hz
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
1.0/25
Engineering Facts
14 facts · every claim linked to a page
Selection
The unit is designed exclusively for external open operation and has no expansion tank, so the connected application must accommodate fluid volume changes.
Determines whether the unit can be used with a closed loop or requires an external open bath/reservoir.
Source
“The machine is developed exclusively for external open operation, and does not possess an expansion tank.”
Installation
Three-phase devices drawing less than 63 A are supplied with cable but without a plug, so the customer must provide and fit the plug.
Affects electrical installation planning and the need to procure a suitable plug.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Three-phase devices drawing more than 63 A are supplied without cable and without plug, requiring full customer-side power cabling.
Affects electrical installation scope and cost for high-power units.
Source
“Three-phase devices with current consumption greater than 63A --> without cable, without plug”
Conformité
Voltage and frequency tolerances of ±10% are permitted only if they do not act in opposite directions, so supply quality must be checked before installation.
Prevents installation on supplies where combined voltage and frequency deviations could exceed the allowed operating envelope.
Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Conformité
The equipment is classified as EN55011 Class A Group 1, so it is intended for industrial electromagnetic environments and may cause interference in residential areas.
Affects site selection and EMC planning, especially if installed near sensitive or residential equipment.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Operation
The minimum temperature of 60 °C depends on ambient temperature being below 60 °C and on the application heat loss, which must be between 0.1 kW and 2.8 kW depending on model.
Determines whether the unit can actually reach its stated minimum temperature in the intended application.
Source
“The heat loss must be between 0.1kW and 2.8kW, depending on the HB model, to reach the minimum temperature of 60°C.”
Operation
The unit has no active refrigeration system, so reaching 60 °C depends on ambient conditions and application heat loss rather than on active cooling.
Clarifies that the unit cannot cool below ambient and that minimum temperature performance is application-dependent.
Source
“These devices do not have an active refrigeration system.”
Installation
The unit requires a 400 V three-phase 50 Hz supply, so the site must provide a suitable three-phase power connection.
Determines the electrical supply infrastructure required at the installation site.
Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation
The pump connection is M38x1.5 male, so mating hoses or adapters must match this thread.
Ensures correct selection of connection hoses and adapters during installation.
Source
“Pump connection M38x1,5 male”
Operation
The ambient temperature must remain between 5 °C and 40 °C for correct operation.
Defines the allowable site environment and prevents operation outside the specified ambient range.
Source
“min. ambient temperature 5 °C”
Safety
The unit has adjustable overtemperature protection according to DIN 12876, which must be set appropriately for the application.
Affects safety setup and protection of the external application against overheating.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Performance
The circulation pump delivers up to 200 l/min at up to 5.5 bar, which must be sufficient for the pressure drop of the external circuit.
Used to verify that the pump can overcome the hydraulic resistance of the external application.
Source
“max. delivery 200 l/min”
Selection
The unit provides 72 kW heating power, which determines heat-up capability and must match the thermal load of the application.
Used to size the unit against the required heating rate and heat losses of the process.
Source
“Heating power 72 kW”
Installation
The degree of protection is IP20, so the unit must be installed in a dry indoor environment protected from water and dust ingress.
Restricts installation location and protects against environmental ingress.