Technical data according to DIN 12876 — KISS 104A (Order-No.: 2037.0040.98)
Characteristic
Value
Operating temperature range
25...100 °C
with water cooling
20...100 °C
with refrigerator
15...100 °C
Temperature stability at 70°C
0,05 K
temperature set point / display
digital
Absolute accuracy
setup for calibration
Internal temperature sensor
Pt100
Interface digital
USB (Device), RS232
Interface
Alarm message
Alarm message
optic, acoustic
Safety classification
III / FL
Heating power at 240V
2,1 kW
Heating power at 230V
2 kW
Heating power at 220V
1,8 kW
Heating power at 208V
1,6 kW
Heating power at 200V
1,5 kW
Pressure / Suction pump
max. delivery
14 l/min
max. delivery pressure
0,25 bar
max. delivery (suction)
10,5 l/min
max. delivery pressure (suction)
0,17 bar
Pump connection
M16x1 male
Bath volume
3 l
Filling capacity
2 l
Bath opening diameter
25 mm
Bath depth
150 mm
Height of bath opening
160 mm
Overall dimensions WxDxH **
147x234x329 mm
Net weight
5 kg
Power supply requirement
200-240V 1~/2~ 50/60Hz
max. current
10 A
min. Fuse
10A
max. Fuse
16A
Pressure equipment category
N/A
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S395787
** Overall dimensions WxDxH
Engineering Facts
14 facts · every claim linked to a page
Selection
The operating temperature range depends on the cooling method, so the choice of water cooling or refrigerator must be matched to the required process temperature.
Determines whether the unit can reach the required low temperature and which cooling option must be specified.
Source
“Operating temperature range 25...100 °C with water cooling 20...100 °C with refrigerator 15...100 °C”
Performance
Heating power depends on the supply voltage, so the available mains voltage must be checked to confirm achievable heating capacity.
Affects heat-up time and whether the unit meets the process power requirement at the site voltage.
Source
“Heating power at 240V 2,1 kW Heating power at 230V 2 kW Heating power at 220V 1,8 kW Heating power at 208V 1,6 kW Heating power at 200V 1,5 kW”
Installation
The electrical supply must be 200-240 V single or two-phase at 50/60 Hz with a branch circuit rated between 10 A and 16 A.
Needed to verify site power availability and correct fuse/breaker sizing before installation.
Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 10 A min. Fuse 10A max. Fuse 16A”
Installation
The unit has IP20 protection, so it must be installed in a dry indoor location protected from water ingress.
Prevents unsuitable installation locations that could cause electrical hazard or equipment damage.
Source
“Degree of Protection IP20”
Installation
Ambient temperature at the installation site must remain between 5 °C and 40 °C.
Ensures the unit operates within its specified environmental limits and maintains performance and safety.
Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Safety
The unit is safety class III / FL with adjustable overtemperature protection according to DIN 12876, which is relevant for flammable or sensitive bath fluids.
Confirms the safety concept and suitability for applications requiring overtemperature protection.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Operation
The autostart function restarts the unit automatically after a power failure, which may be undesirable for some processes and must be considered during commissioning.
Affects process safety and restart behavior after mains interruption.
Source
“Autostart function for power failure”
Selection
The optional Pt100 sensor connection only displays process temperature and does not control it, and it must be ordered factory-fitted.
Prevents specifying the option expecting external temperature control and ensures it is included in the factory order.
Source
“Option: Pt100 sensor connection #10688 to display (not control) e.g. of the process temperature (only available factory fitted, additional charge).”
Operation
The plastic pump is not resistant to acetone or polyglycols, so these fluids must be avoided or a more resistant plastic must be requested.
Prevents chemical attack and premature pump failure when selecting the bath fluid.
Source
“Special Case: Acetone and Polyglycol: The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.
Avoids freezing and damage when operating near or below the freezing point of water.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Conformité
An external branch circuit protection according to UL 489 is required for the installation.
Mandatory for compliant electrical installation and to avoid failing inspection.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Voltage and frequency tolerances are limited to ±10% and must not combine in opposite directions, so site power quality must be verified.
Prevents operation outside permissible supply conditions that could damage the unit or void compliance.
Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Installation
Single and two-phase devices are supplied with a power cable and country-specific plug, which must be specified at ordering.
Ensures the correct plug is supplied for the installation country and avoids delays or rework.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Installation
Space requirements around the unit must be respected according to the operating conditions.
Needed for adequate ventilation and access during installation and maintenance.
Source
“Please respect space requirements. See operating conditions at www.huber-online.com”