Sommaire

  1. Overview p. 1
    Product introduction and main features of KISS K12
  2. Spec pp. 1-2
    Technical data and performance specifications per DIN 12876
  3. p. 2 table, 2 rows
  4. p. 1 table, 39 rows

Technical data according to DIN 12876

ParameterValue
max. ambient temperature40 °C
from Serial-No.:S452707

Technical data according to DIN 12876

ParameterValue
Operating temperature range-20...200 °C
Temperature stability at 70°C0,05 K
temperature set point / displaydigital
Absolute accuracysetup for calibration
Internal temperature sensorPt100
Interface digitalUSB (Device), RS232
InterfaceAlarm message
optic, acoustic
Safety classificationIII / FL
Heating power1,5 kW
Cooling power
at 20°C0,25 kW
at 0°C0,2 kW
at -10°C0,12 kW
at -20°C0,05 kW
Refrigeration machineair-cooled, natural
refrigerant
Refrigerant (ASHRAE, GHS)R-290 (A3, H220)
Global Warming Potential (GWP)0,02
Refrigerant quantity0,041 kg
CO2 equivalent0,0 t
Pressure / Suction pump
max. delivery14 l/min
max. delivery pressure0,25 bar
max. delivery (suction)10,5 l/min
max. delivery pressure (suction)0,17 bar
Bath volume10,5 l
Width bath opening WxD290x152 mm
Bath depth150 mm
Height of bath opening265 mm
Overall dimensions WxDxH **350x560x430 mm
Net weight28 kg
Power supply requirement110-115V 1~ 50/60Hz
max. current immersion thermostat15 A
max. current refrigerated bath2,6 A
min. Fuse10A
max. Fuse16A
Degree of ProtectionIP20
min. ambient temperature5 °C

** Overall dimensions WxDxH

Engineering Facts

12 facts · every claim linked to a page
Selection

The unit is rated for an operating temperature range of -20 to 200 °C, so it must be matched to the process temperature window.

Defines whether the bath can cover the required process temperatures before selection.

Source
“Operating temperature range -20...200 °C”
Performance

Cooling capacity drops with lower temperature, from 0.25 kW at 20 °C to 0.05 kW at -20 °C, so heat load must be checked at the coldest set point.

Prevents undersizing the refrigerating unit for low-temperature duty.

Source
“at 20°C 0,25 kW at 0°C 0,2 kW at -10°C 0,12 kW at -20°C 0,05 kW”
Safety

The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, which affects room ventilation and ignition-source planning.

Flammable refrigerant classification drives installation safety and site requirements.

Source
“Refrigerant (ASHRAE, GHS) R-290 (A3, H220)”
Installation

The power supply requirement is 110-115 V single-phase 50/60 Hz with a minimum 10 A fuse, so the electrical supply must be verified on site.

Mismatch with the available mains voltage or fuse rating prevents commissioning.

Source
“Power supply requirement 110-115V 1~ 50/60Hz”
Installation

The degree of protection is IP20, so the unit must be installed in a dry, non-condensing location protected from water ingress.

IP20 limits where the bath can be placed and how it can be cleaned.

Source
“Degree of Protection IP20”
Operation

Ambient temperature must stay between 5 °C and 40 °C, and cooling capacity may fall as ambient rises.

Room conditions directly affect achievable cooling performance and must be checked at the installation site.

Source
“If the ambient temperature rises, the cooling capacity may drop.”
Operation

The plastic pump is not resistant to acetone or polyglycols, so these media must be avoided or a more resistant plastic ordered.

Prevents chemical attack and pump failure when selecting the heat transfer fluid.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than used pure.

Guides correct bath fluid preparation to avoid freezing damage.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.

Defines acceptable mains quality for reliable operation and warranty compliance.

Source
“-5% voltage and +2% frequency -> not allowed!”
Conformité

The equipment is EMC classified as Class A, Group 1 to EN55011, which has implications for use in residential environments.

Class A equipment may cause interference in domestic settings and affects site selection.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Single/two-phase devices are delivered with a country-specific plug that must be specified at ordering.

Missing plug specification at order leads to unusable power connection on delivery.

Source
“with power cable and country-specific plug (please specify when ordering)”
Selection

An optional Pt100 sensor connection can display (not control) process temperature and is only available factory fitted.

Must be decided at purchase because it cannot be retrofitted in the field.

Source
“Option: Pt100 sensor connection #10688 to display (not control) e.g. of the process temperature (only available factory fitted, additional charge).”