Minichiller 1000 OLÉ

Sommaire

  1. Overview p. 1
    Product name and cover page of the Minichiller 1000 OLÉ datasheet
  2. Spec p. 2
    Technical data according to DIN 12876 in a table
  3. p. 2 table, 4 rows

Technical data according to DIN 12876

ParameterValue
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S617619

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit's cooling capacity drops as ambient temperature rises, so size the chiller with margin if the room runs warm.

Cooling capacity is rated at 20 °C room temperature; higher ambient reduces available cooling and can undersize the loop.

Source
“If the ambient temperature rises, the cooling capacity may drop.”
Selection

The plastic circulation pump is not compatible with acetone or polyglycols, so these fluids require the optional more resistant pump.

Fluid compatibility determines whether the standard pump can be used or a special pump must be ordered.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, use water mixed with glysantine or ethylene glycol rather than pure water.

Prevents freezing and damage in the bath/loop when operating near or below 0 °C.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

Voltage and frequency tolerances must be met simultaneously; a -5% voltage with +2% frequency is not allowed.

Ensures the supply is within EN60034-1 limits and avoids invalid operating conditions.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

Single/two-phase units are supplied with a power cable and country-specific plug, which must be specified at order.

Avoids missing or wrong plug on delivery and simplifies site commissioning.

Source
“with power cable and country-specific plug (please specify when ordering)”
Installation

Three-phase units below 63 A come with cable but no plug, so the customer must provide and fit the plug.

Electrical installation planning must include a suitable plug and connection by qualified personnel.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

The unit requires 208-240 V single or two-phase supply at 50/60 Hz with max current 4 A.

Confirms the electrical supply and breaker/fuse sizing for the installation.

Source
“208-240V 1~/2~ 50/60Hz”
Installation

The chiller needs at least 2.8 l of heat transfer fluid to fill the system.

Determines the minimum fluid volume and expansion tank sizing for commissioning.

Source
“min. filling capacity 2,8 l”
Installation

The pump connection is M16x1 male, and hose connector NW12 with M16x1 sleeve nuts are included.

Specifies the plumbing interface and included fittings for connecting hoses.

Source
“hose connector NW12 #6087, sleeve nuts thread M16x1#6089”
Operation

Whisper mode reduces sound level to about 51 dB(A) but also reduces pump delivery to 14 l/min and 0.2 bar.

Noise-sensitive installations may need whisper mode, but flow and pressure must still meet process requirements.

Source
“Delivery rate: 14 l/min”
Operation

The return line temperature must not exceed 80 °C.

Exceeding this can damage the chiller or invalidate safe operation.

Source
“Permissible temperature in return line 80 °C”
Safety

The refrigerant is R-1270 (A3, H220), a flammable natural refrigerant with GWP 0.

A3 flammability affects room ventilation, ignition source control, and refrigerant handling procedures.

Source
“R-1270 (A3, H220)”
Conformité

The equipment is EMC Class A Group 1, so it may cause radio interference in residential areas.

Class A equipment is intended for industrial environments; residential use may require additional EMC measures.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Maintenance

A drain valve is available as an optional accessory for easier fluid draining.

Simplifies maintenance and fluid changes; should be ordered with the unit if draining is planned.

Source
“Drain valve #6839”