Grande Fleur Operation Manual

Sommaire

  1. Overview pp. 5-10
    Table of contents, function description, normal operation and shutdown overview
  2. Overview pp. 11-21
    Safety notes, device overview and general introductory information
  3. Procedure pp. 22-32
    Step-by-step procedures for initial setup and commissioning
  4. Overview pp. 33-44
    Detailed description of device functions and operating modes
  5. Procedure pp. 45-64
    Procedures for entering and configuring setup parameters
  6. Procedure pp. 65-66
    Procedures for everyday operation of the device
  7. Spec pp. 67-73
    Interface specifications and software update procedures
  8. Procedure pp. 74-82
    Maintenance schedules, service procedures and troubleshooting
  9. Procedure pp. 83-86
    Procedures for safe shutdown and storage
  10. pp. 87-88
    Technical data, temperature ranges and supplementary information
  11. p. 9 table, 40 rows
  12. p. 42 table, 12 rows
  13. p. 36 table, 17 rows
  14. p. 34 table, 8 rows
  15. p. 43 table, 15 rows
  16. p. 26 table, 6 rows
  17. p. 35 table, 4 rows
  18. p. 70 table, 2 rows
  19. p. 73 table, 6 rows

Содержание (оглавление) страницы 9

РазделНазваниеСтраница
4.2.5Monitoring the Pt100 temperature sensors50
4.2.6Optimum control parameters for optimum temperature control51
4.2.7Sub-category: “Select auto/expert mode”.51
4.2.8Sub-category: “Configuration auto”51
4.2.8.1Sub-category: “Find parameters”51
4.2.8.2Sub-category: “Control Dynamics”53
4.2.8.3Sub-category: “Fluid Properties”54
4.2.8.4Sub-category: “Display parameters”55
4.2.9Sub-category: “Configuration manual”56
4.2.9.1Sub-category: “Change parameters”56
4.2.9.2Sub-category: “Display parameters”57
4.2.9.3Sub-category: “Control structure”57
4.2.10Sub-category: “Reset parameters”57
4.2.11Sub-category: “Display parameters”58
4.2.12Setting the setpoint thresholds58
4.2.13Setting the setpoint59
4.3Filling, venting, degassing and draining59
4.3.1Externally closed application59
4.3.1.1Filling and venting externally closed application59
4.3.1.2Degassing externally closed applications61
4.3.1.3Draining externally closed applications63
5Normal operation65
5.1Automatic operation65
5.1.1Temperature control65
5.1.1.1Starting the temperature control process65
5.1.1.2Ending the temperature control process65
5.1.2Temperature control via a created temperature control program66
5.1.2.1Starting the temperature control program66
5.1.2.2Ending/cancelling the temperature control program66
6Interfaces and software update67
6.1Interfaces at the “Pilot ONE®” controller67
6.1.110/100 Mbps Ethernet for RJ45 network sockets67
6.1.2USB-2.0 interface67
6.1.2.1USB-2.0 interface, host68
6.1.2.2USB-2.0 interface, device68
6.2Interfaces on the temperature control unit68
6.2.1Interfaces at the back68
6.2.1.1Connector POKO (floating contact) alarm68
6.2.1.2Jack ECS (External Control Signal) standby70
6.2.1.3Service interface71

На странице напечатано оглавление (содержание), а не матрица конфигураций; транспонирование не требуется.

Restore factory settings — “X” = Value is reset, “–” = Value is not reset

Description(A)(B)(C)(D)(E)(F)
AcousticsXXX–––
Display functions Warnings; Messages; Inactive Menu Items; Resolution TemperatureXXX–––
Display modesXXX–––
AutoStart responseXXX–––
Limitations DeltaT limit; max. heating variable; max. cooling variable; max. permissible current consumption (for 230 V temperature control units)XXX–––
Com.G@te (only if Com.G@te is connected) Analog Interface (configuration input) Meaning of AIF input; response on cable break; measurement range settingXXX––X
Calibration–––––X
In case of analog error; current / voltage switching; filter constantXXX––X
Configuration Output AIF output variable; Output range settingXXX––X
Calibration–––––X
RS232/RS485 (Hardware RS, Baud rate, device address); ECS Standby; POKO AlarmXXX–––
Properties (other) Change thermal fluid (thermal fluid, liter indicator, bypass usage); cooling bath / cooling (CC-E only)XXX–––

(A) = All together; (B) = Device parameters without OT; (C) = Device Parameter; (D) Program Controller; (E) = Menu; (F) = Com.G@te

Overview of E-grade functions

FunctionE-grade BasicE-grade ExclusiveE-grade Professional
Temperature control
Controller parameterization: predefined1 / TAC2X/––/X–/X
Find parameters: Fast Identification / With Preliminary Test / Estimate Control Parameters–/–/XX/X/–X/X/–
Sensor calibration for control sensor3: x -point255
Monitoring: Low level and overtemperature4XXX
Adjustable alarm thresholdsXXX
VPC5 (Variable Pressure Control)XXX
Ventilation programXXX
Automated compressorXXX
Setpoint limitationXXX
Program controller: x programs / max x steps–/–3/1510/100
Ramp function: linear / non-linear–/–X/–X/X
Temperature control mode: Internal and process–XX
Adjustable max. heating and cooling capacity–XX
Display & operation
Temperature display: 5.7" TouchscreenXXX
Display mode: graphically / numerically large / Explore–/X/––/X/–X/X/–

X = Standard equipment, O = optional, – = not possible. 1 TAC function available as a 30 day evaluation version. 2 True Adaptive Control. 3 Internal Pt100 and External Pt100. 4 For temperature control units with integrated overtemperature protection. 5 For units with variable-speed pump or external bypass.

Designation / Specification

DesignationSpecification
Calcium carbonate per liter≤ 1.5 mmol/l; corresponds to a water hardness of: ≤ 8.4 °dH (soft)
pH valuebetween 6.0 and 8.5
Ultrapure water, distillatesAdd 0.1 g of sodium carbonate (Na2CO3) per liter
Non-approved waterDistilled, deionized, demineralized, chloric, ferruginous, ammoniacal, contaminated or untreated river water or sea water
Volume circulated (at least)3 l/min. (not valid for cooling baths)
Thermal fluid: Water without ethylene glycolUse excluded
Thermal fluid: Water-ethylene glycol mixtureUse -35 °C (with anti-free possible up to -38 °C) to +95 °C
Thermal fluid compositionThe mixture’s temperature must be 10 K below the permissible min. temperature. For the permissible temperature range, refer to the datasheet.→ From page 87, section »Annex«.

Description

Description(A)(B)(C)(D)(E)(F)
Sensor calibration––––––
Automated compressorXXX–––
Enter programX––X––
Pump settings Pump speed setpoint; Pump pressure setpoint; Control ModeXXX–––
Fluid properties Thermal fluid; liter indicatorXXX–––
Configuration manual parameters Protective function High Limit Alarm Internal; Low Limit Alarm Internal; High Limit Alarm Process; Low Limit Alarm Process; Hydrostatic Correction; Warning Time Limit (CC-E only); Min. Level (for temperature control units with analogue level sensor); Max. Level (for temperature controllers with analogue level sensor); Freeze Protection (optional)XXX–––
SetpointXXX–––
Setpoint limitation Min. Setpoint; Max. SetpointXXX–––
LanguageXXX–––
Temperature formatXXX–––
Temperature control modeXXX–––
Favourites menuX–––X–
Pilot ONE Ethernet IP address; Subnet Mask; Remote AccessXXX–––
Time formatXXX–––
2. SetpointXXX–––

Side / Distance in cm (for operation in a tub)

SideAir coolingWater cooling
[A1] TopAir outlet on top of unit: free standing–
[A2] Topcan be located under a benchcan be located under a bench
[B] Leftmin. 20min. 20
[C] Rightmin. 20min. 20
[D] Frontmin. 20min. 20
[E] Rearmin. 20min. 20

Overview of the E-grade variants

Temperature control units / E-gradeE-grade BasicE-grade ExclusiveE-grade Profes-sional
Brewing thermostats–XO
Unistat temperature control units––X
UniCAL––X
Other temperature control unitsXOO

Pin assignment

PinSignal
1.3E2
2E1

Pin assignment (front view)

Pin assignment

PinDescriptionSignal
1Current output, T external0 a)/4 - 20 mA or 0 - 10 V If using 0 - 10 V, integrate a 500 Ω resistor.
2Current output, setpoint0 a)/4 - 20 mA or 0 - 10 V
3GND for analog outputsGND
4Analog input (programmable)0 a)/4 - 20 mA or 0 - 10 V Power input: 200 Ω working resistance Voltage input: 100 kΩ input resistance
5Current output, freely programmable0 a)/4 - 20 mA or 0 - 10 V If using 0 - 10 V, integrate a 500 Ω resistor.
6GND for analog inputGND

a)Consult with our Customer Support.

Engineering Facts

11 facts · every claim linked to a page
Safety

The unit must not be installed or started in an ATEX zone without an Ex px cabinet.

Determines whether the unit can be placed in a hazardous area at all.

Source
“Do NOT install or start up the temperature control unit within an ATEX zone.”
Installation

A drip tray for condensation water/thermal fluid must be installed below the unit by the responsible body.

Affects site preparation and civil works before commissioning.

Source
“The responsible body must install a condensation water / thermal fluid drip tray below the temperature control unit.”
Installation

For natural refrigerant units, if the charge exceeds 8 g per m³ of room air, a calibrated gas warning sensor is mandatory.

Dictates room sizing and gas detection requirements for NR units.

Source
“Over 8 g refrigerant per m³ room air: A gas warning sensor must be fitted and functioning.”
Installation

Units with more than 1 kg refrigerant must not be installed below ground level.

Restricts possible installation locations for larger refrigerant charges.

Source
“Temperature control units with more than 1 kg refrigerant must not be installed below ground level (GL).”
Installation

The circulation flow must connect to the lower port of the application and the return to the higher port to vent air bubbles.

Prevents air locks and ensures proper venting of externally closed loops.

Source
“you must ensure that the >Circulation flow< [1] connection from the temperature control unit is attached to the lower connection point of the application and the >Circulation return< [2] into the temperature control unit is attached to the higher connection point of the application.”
Installation

Power cables longer than 3 m must not be used so the unit can always be disconnected from the mains.

Constrains electrical layout and disconnect accessibility at the installation site.

Source
“Do not use a power cable that is longer than 3 m to be able to disconnect the temperature control unit at any time from the mains.”
Operation

The thermal fluid viscosity must not exceed 50 mm²/s at the lowest working temperature.

Limits the choice of thermal fluid and the achievable low-temperature range.

Source
“The maximum viscosity of the thermal fluid must not exceed 50 mm²/s at the lowest working temperature!”
Operation

Different thermal fluid types must never be mixed; the circuit must be rinsed when changing fluid type.

Prevents fluid degradation and damage during fluid changes or top-ups.

Source
“Do not mix different types of thermofluid (such as mineral oil, silicone oil, synthetic oil, water, etc.) in a thermofluid circuit.”
Operation

For water cooling, the cooling water supply must be shut off even for short downtimes such as overnight.

Avoids water damage and condensation risks during idle periods.

Source
“Shut off the cooling water supply to the temperature control unit even for shorter downtimes (e.g. overnight).”
Safety

Electrical lines directly below a water-cooled unit must be protected against liquid ingress from condensation.

Affects cable routing and protection below water-cooled units.

Source
“Electrical lines directly below the temperature control unit must be protected against liquid ingress.”
Installation

The maximum torque values for pump connections must not be exceeded and a leak test must always be performed.

Prevents damage to fittings and leaks at the hydraulic connections.

Source
“The values of the maximum torque (see table) must not be exceeded.”