Engineering Facts
13 facts · every claim linked to a page
Selection
The unit covers a working temperature range of -45 to 100 °C, so it must be matched to the process temperature window before ordering.
Defines whether the cooler can reach and hold the required process setpoint.
Source
“Operating temperature range
-45...100 °C”
Performance
Cooling capacity drops sharply at low temperature (0.24 kW at 0 °C down to 0.05 kW at -40 °C), so the load must be sized against the capacity at the actual operating point.
Prevents undersizing the immersion cooler for low-temperature duty.
Source
“at -40°C
0,05 kW”
Safety
The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, which affects room ventilation and ignition-source requirements.
Flammable refrigerant classification drives installation safety and site approval.
Source
“R-290 (A3, H220)”
Installation
The unit requires a 208-240 V single- or two-phase 50/60 Hz supply with a minimum 10 A fuse, so the site power and protection must be verified.
Determines the electrical connection and upstream fuse rating needed at the installation site.
Source
“208-240V 1~/2~ 50/60Hz”
Installation
Ambient temperature must stay between 5 °C and 40 °C, so the unit cannot be placed in a cold room or hot enclosure outside this band.
Ambient limits govern where the air-cooled unit may be installed.
Source
“min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Performance
Cooling capacity is rated at 20 °C room temperature and falls if ambient rises, so capacity margins must account for the actual site ambient.
Prevents overestimating available cooling in a warm installation room.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation
Voltage and frequency tolerances are coupled: ±5 % voltage with ±2 % frequency, but opposite-sign deviations are not allowed.
Guides checking supply quality before connecting the unit.
Source
“-5% voltage and +2% frequency -> not allowed!”
Conformité
The unit is EMC class A, group 1, so it is intended for industrial environments and may cause interference in residential areas.
Affects site selection and EMC coordination with sensitive equipment.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
Single- and two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering.
Avoids a missing or wrong plug on delivery and delays at commissioning.
Source
“with power cable and country-specific plug (please specify when ordering)”
Installation
Three-phase devices below 63 A come with cable but no plug, so the customer must provide and fit the plug.
Clarifies the electrical scope of supply for three-phase variants.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Space requirements around the unit must be respected per the operating conditions, so clearance must be planned at the site.
Air-cooled units need clearance for heat rejection and service access.
Source
“Please respect space requirements. See operating conditions at www.huber-online.com”
Operation
The refrigeration unit runs continuously and temperature is controlled by a solenoid valve in the cooling circuit, which is relevant for control behavior and wear.
Explains the control principle and expected continuous compressor operation.
Source
“The refrigeration unit works continuously. The temperature control actuates a solenoid valve in the cooling circuit.”
Operation
The external Pt100 sensor connects via cable and plug, and a Pt100 sensor is included, so external temperature control is possible out of the box.
Confirms external sensor control capability and included accessory.
Source
“The temperature sensor is connected to the device with cable and plug.”