Selection
The chiller's cooling capacity drops as ambient temperature rises, so size the unit for the actual room conditions.
Cooling capacity is rated at 20°C ambient; higher ambient reduces available cooling and may require a larger model.
Source
“If the ambient temperature rises, the cooling capacity may drop.”
Installation
The unit requires a 220–240 V single-phase 50 Hz supply with a 16 A fuse and draws up to 13 A.
Electrical supply and protection must be sized accordingly to avoid nuisance tripping or undersized wiring.
Source
“Power supply requirement 220-240V 1~ 50Hz max. current 13 A Fuse 16 A”
Installation
Voltage and frequency tolerances must be met simultaneously; a -5% voltage with +2% frequency is not allowed.
Ensures the site power quality is within the allowed combined tolerance before commissioning.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
Single/two-phase devices are delivered with a country-specific plug that must be specified at ordering.
Prevents delivery of the wrong plug type and avoids rework at installation.
Source
“with power cable and country-specific plug (please specify when ordering)”
Operation
The plastic pump is not resistant to acetone or polyglycols, so these fluids must be avoided or a more resistant pump specified.
Using incompatible fluids can damage the pump and cause downtime; select alternative pump material if needed.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.
Prevents freezing and damage when operating near or below 0°C.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Safety
The refrigerant R-449A is classified A1 with GHS hazard H280, and the unit contains 0.58 kg (0.8 t CO2e).
Informs refrigerant handling, leak response, and environmental reporting requirements.
Source
“R-449A (A1, H280)”
Safety
The unit is classified IP20, so it must be installed in a dry location protected from water ingress.
IP20 offers no water protection; installation in wet or washdown areas is not permitted.
Source
“Degree of Protection IP20”
Installation
Ambient temperature must remain between 5°C and 40°C for proper operation.
Operating outside this range may affect performance, reliability, and warranty.
Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation
Space requirements around the unit must be respected per the operating conditions.
Air-cooled units need adequate clearance for heat rejection; insufficient space reduces cooling capacity.
Source
“Please respect space requirements. See operating conditions at www.huber-online.com”
Performance
The circulation pump delivers up to 29 l/min at 1 bar, with 21 l/min at 0.5 bar.
Determines whether the pump can overcome the pressure drop of the external circuit at the required flow.
Source
“max. delivery 29 l/min max. delivery pressure 1 bar Delivery at 0,5 bar 21 l/min”
Performance
Cooling capacity is 2.2 kW at 15°C, 1.6 kW at 0°C, and 1 kW at -10°C (at 20°C ambient).
Allows matching the chiller to the actual heat load at the intended operating temperature.
Source
“at 15°C 2,2 kW at 0°C 1,6 kW at -10°C 1 kW”
Conformité
The equipment is EMC Class A Group 1, which may require additional filtering in residential environments.
Class A equipment is intended for industrial use and may cause interference in residential areas.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Maintenance
The refrigerant charge is 0.58 kg with a GWP of 1397, giving 0.8 t CO2 equivalent.
Needed for leak-check intervals, refrigerant accounting, and environmental compliance.
Source
“Refrigerant quantity 0,58 kg CO2 equivalent 0,8 t”