Introduces the Unichiller 250T model and its intended use
Specp. 2
Technical data according to DIN 12876 standard
p. 1table, 35 rows
Unichiller 250T — Technical data according to DIN 12876
Characteristic
Value
Order-No.
3057.0001.01
Operating temperature range
-20...40 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
Temperature stability at -10°C
0,5 K
Safety classification
I / NFL
Cooling power at 15°C
25 kW
Cooling power at 0°C
18 kW
Cooling power at -10°C
11 kW
Cooling power at -20°C
6 kW
Refrigeration machine
air-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Circulation pump:
D3
max. delivery
210 l/min
max. delivery pressure
4,7 bar
Delivery at 1,1 bar
202 l/min
Delivery at 1,5 bar
193 l/min
Delivery at 2,0 bar
179 l/min
Delivery at 2,5 bar
159 l/min
Delivery at 3,0 bar
140 l/min
Delivery at 3,5 bar
115 l/min
Delivery at 4,0 bar
77 l/min
Delivery at 4,5 bar
31 l/min
Pump connection
G1 1/4 male
min. filling capacity
14 l
Volume of expansion
39 l
Overall dimensions WxDxH **
905x2172x1900 mm
Power supply (3 Phase)
400V 3~ 50Hz
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
** см. сноску в оригинале (Overall dimensions WxDxH **)
Engineering Facts
12 facts · every claim linked to a page
Installation
The chiller is air-cooled, so the installation site must provide adequate air flow and respect the specified space requirements.
Air-cooled units reject heat to the room; insufficient clearance or ventilation will reduce cooling capacity and can cause overheating.
Source
“** Please respect space requirements. See operating conditions at www.huber-online.com”
Installation
The unit requires a 3-phase 400 V 3~ 50 Hz supply, and three-phase devices below 63 A are delivered with cable but without a plug.
The installer must plan for a suitable 3-phase supply and provide the correct plug/connection on site.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Voltage and frequency tolerances are +/−5% and +/−2% respectively, but certain combinations such as −5% voltage with +2% frequency are not allowed.
Supply quality must be checked before commissioning; an out-of-tolerance combination can damage the compressor or pump.
Source
“-5% voltage and +2% frequency -> not allowed!”
Performance
Cooling capacity is rated at 20 °C room temperature and will drop if the ambient temperature rises.
Selection must account for the actual ambient temperature at the installation site, not just the nominal 25 kW at 15 °C.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Operation
The unit is intended for externally closed applications, so the external circuit must be closed and compatible with the stainless-steel pump and expansion tank.
Using an open external circuit can cause evaporation, contamination, or overflow of the atmospheric expansion tank.
Source
“For externally closed applications.”
Safety
The refrigerant R-449A is classified A1 with GHS hazard H280, and the unit has a GWP of 1397.
Refrigerant handling, leak detection, and disposal must follow H280 (pressurised gas) precautions, and high GWP affects environmental compliance and service planning.
Source
“R-449A (A1, H280)”
Conformité
The equipment is EMC Class A, Group 1, which is intended for industrial environments and may cause interference in residential areas.
Class A equipment may not be suitable for residential or light-industrial locations without additional filtering or separation.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
The pump connection is G1 1/4 male, and a matching hose coupling is included.
External piping and hoses must be ordered or adapted to G1 1/4 male threads; the included coupling covers only one connection.
Source
“Hose coupling for G1 1/4 male”
Operation
The minimum filling capacity is 14 l and the expansion volume is 39 l, which sets the minimum heat-transfer fluid charge and expansion allowance.
Underfilling can cause pump cavitation and poor temperature control; the expansion volume must be respected when sizing the external loop.
Source
“min. filling capacity”
Operation
The pump delivers 210 l/min maximum and 4.7 bar maximum pressure, with flow falling as pressure rises.
The external circuit pressure drop must be checked against the pump curve; high-resistance loops will reduce flow below the required rate.
Source
“max. delivery pressure”
Installation
The unit has IP20 protection and must be installed where ambient temperature is between 5 °C and 40 °C.
IP20 offers no water protection, and operation outside 5–40 °C ambient may impair cooling or damage components.
Source
“min. ambient temperature”
Selection
The temperature stability at −10 °C is 0.5 K, which defines the achievable control accuracy for sensitive processes.
Processes requiring tighter than ±0.5 K stability need a different model or additional control measures.