Product introduction and general description of the Unistat 620w unit
Specpp. 2-3
Technical data according to DIN 12876 including performance and dimensions
p. 2table, 30 rows
p. 1table, 30 rows
Technical data according to DIN 12876
Parameter
Value
Delivery at 1,0 bar
152 l/min
Delivery at 1,5 bar
123 l/min
Delivery at 2,0 bar
86 l/min
Pump connection
M38x1,5 male
max. permissible kin. viscosity
100 mm²/s
Cooling water connection
G3/4 male
Consumption at water 15°C, flow 0°C
760 l/h
min. cooling water differential pressure
1 bar
max. cooling water differential pressure
4 bar
max. cooling water pressure
6 bar
max. System pressure
6 bar
min. filling capacity
8,9 l
Filling capacity expansion tank
19,9 l
Overall dimensions Temperature Control Unit WxDxH
918x963x1771 mm
Dimensions of External Transformer WxDxH
364x740x682 mm
Net weight
616 kg
Weight External Transformer
200 kg
sound pressure level +/- 4 dB(A)
68 dB(A)
Power Supply for Temperature Control Unit
460V 3~ 60Hz
Input Voltage External Transformer
208V 3~ 60Hz
Output Voltage External Transformer
440V 3~ 60Hz
Max. current temperature control unit
25,4 A
Max. current temp. control unit with external transformer
55 A
Fuse temperature control unit
3X32 A
Fuse input voltage external transformer
3X60 A
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S607684
Delivery at 1,0 bar; Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.
Technical data according to DIN 12876 — Unistat 620w (Order-No.: 5016.0104.01)
Parameter
Value
Operating temperature range
-60...200 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Interface analog
optional
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
12 kW
Cooling power with Thermooil at 200°C
12 kW
Cooling power with Thermooil at 100°C
12 kW
Cooling power with Thermooil at 20°C
12 kW
Cooling power with Ethanol at 0°C
12 kW
Cooling power with Ethanol at -20°C
12 kW
Cooling power with Ethanol at -40°C
5,6 kW
Cooling power with Ethanol at -60°C
1,4 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-452A (A1, H280)
Global Warming Potential (GWP)
2141
Refrigerant quantity
3,6 kg
CO2 equivalent
7,7 t
UN-number
UN 2857
Circulation pump:
MK pump
max. delivery
196 l/min
max. delivery pressure
2,5 bar
Delivery at 0,5 bar
175 l/min
Engineering Facts
14 facts · every claim linked to a page
Installation
The unit requires a water-cooled cooling machine, so a cooling water supply with adequate differential pressure must be planned at the installation site.
Determines whether cooling water infrastructure is needed before selecting/installing the unit.
Source
“Refrigerated Heating Circulator with water-cooled cooling machine and electronical level indicator.”
Installation
Cooling water inlet temperature must stay between 5°C and 20°C, and pipes must be insulated if water can drop below the condensation point.
Defines cooling water supply limits and condensation/insulation requirements for installation.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops as cooling water temperature rises, so the specified 12 kW figures are only valid at the stated reference conditions.
Prevents overestimating cooling capacity when site cooling water is warmer than 15°C.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation
The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Guides water quality and material compatibility decisions for the cooling loop.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Conformité
Voltage and frequency tolerances are ±5% and ±2% respectively, but they may not be combined in opposite directions.
Defines acceptable mains quality for reliable operation and warranty compliance.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety
Leakage current exceeds 3.5 mA, which affects protective earthing and RCD selection.
Critical for electrical installation and choosing appropriate residual current protection.
Source
“Attention: leakage current > 3,5mA”
Installation
Three-phase devices below 63 A are delivered with a cable but without a plug, so a plug must be sourced separately.
Affects electrical installation planning and procurement of the mains connection.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
The temperature control unit is fed from an external transformer, so both units and their combined current must be accommodated.
Requires space and electrical planning for the transformer in addition to the main unit.
Source
“the required voltage is provided using a external transformer (supplied as standard)”
Selection
The unit is classified as pressure equipment category I, which matters for compliance documentation.
Relevant for regulatory compliance assessment during selection.
Source
“Pressure equipment category”
Operation
Ambient temperature must remain between 5°C and 40°C for correct operation.
Defines the allowable room conditions at the installation site.
Source
“max. ambient temperature”
Installation
The unit has IP20 protection, so it must be installed in a dry indoor environment protected from water ingress.
Restricts installation location and environmental protection requirements.
Source
“Degree of Protection”
Maintenance
The refrigerant R-452A has a GWP of 2141 and a CO2 equivalent of 7.7 t, which creates leak-checking and reporting obligations.
Affects maintenance obligations and environmental compliance for the refrigeration circuit.
Source
“Global Warming Potential (GWP)”
Installation
Pump connections are M38x1.5 male and cooling water connections are G3/4 male, so matching hoses and fittings must be ordered.
Needed to specify compatible hoses and adapters before installation.
Source
“Pump connection”
Selection
The unit needs a minimum filling capacity of 8.9 l and has a 19.9 l expansion tank, which sets the heat transfer fluid volume to prepare.
Determines the required heat transfer fluid quantity for commissioning.