  1. Overview p. 1
    Product introduction and general description of the Unistat 815w unit
  2. Spec pp. 1-2
    Technical data and performance specifications according to DIN 12876
  3. p. 2 table, 24 rows
  4. p. 1 table, 30 rows

Technical data according to DIN 12876

ParameterValue
max. delivery40 l/min
max. delivery pressure0,9 bar
Delivery at 0,3 bar27 l/min
Delivery at 0,5 bar20 l/min
Delivery at 0,8 bar5 l/min
Pump connectionM30x1,5 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 0°C130 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity3,2 l
Filling capacity expansion tank3,4 l
Overall dimensions WxDxH **460x604x1465 mm
Net weight222 kg
sound pressure level +/- 4 dB(A)60 dB(A)
Power supply requirement400V 3~ 50Hz
max. current13 A
Fuse3x16A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.S466955

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ParameterValue
Operating temperature range-85...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power2 kW
Cooling power with Thermooil at 250°C1,5 kW
Cooling power with Thermooil at 200°C1,5 kW
Cooling power with Thermooil at 100°C1,5 kW
Cooling power with Ethanol at 0°C1,5 kW
Cooling power with Ethanol at -20°C1,5 kW
Cooling power with Ethanol at -40°C1,4 kW
Cooling power with Ethanol at -60°C1,2 kW
Cooling power with Ethanol at -80°C0,2 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity0,95 kg
CO2 equivalent2,0 t
Refrigerant 2nd stage (ASHRAE, GHS)R-23 (A1, H280)
Global Warming Potential (GWP)14800
Refrigerant quantity 2nd stage0,39 kg
CO2 equivalent 2nd stage5,8 t
UN-numberUN 2857

Order-No.: 1053.0006.01; Unistat 815w

Engineering Facts

13 facts · every claim linked to a page
Installation

The unit requires a 400 V three-phase supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.

Determines the site power supply, cable and protection device needed before installation.

Source
“Power supply requirement 400V 3~ 50Hz”
Installation

The cooling water circuit must be supplied at 5–20 °C with at least 1 bar differential pressure and no more than 6 bar, and pipes must be insulated if water can drop below the dew point.

Defines the cooling water utility requirements and condensation risk at the installation site.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling capacity is rated at 1.5 kW down to -20 °C but falls to 0.2 kW at -80 °C, so low-temperature duty must be checked against the load.

Prevents oversizing expectations for deep-cold applications where available cooling power is small.

Source
“at -80°C 0,2 kW”
Performance

The circulation pump delivers 40 l/min maximum but only 5 l/min at 0.8 bar, so external circuit pressure drop must be checked against the pump curve.

Ensures the external loop resistance does not starve flow and degrade temperature control.

Source
“Delivery at 0,8 bar 5 l/min”
Safety

The unit is safety class III / FL with adjustable overtemperature protection per DIN 12876, which matters for flammable or sensitive fluids.

Confirms suitability and required protective settings for the intended heat transfer fluid.

Source
“Safety classification III / FL”
Safety

The refrigerant R-452A has GWP 2141 and 0.95 kg charge (2.0 t CO2e), and the second stage R-23 has GWP 14800 and 0.39 kg (5.8 t CO2e), so refrigerant regulations apply.

Affects environmental compliance, leak-check obligations and CO2-equivalent reporting.

Source
“Global Warming Potential (GWP) 2141”
Operation

The controller offers selectable internal/process control mode and a 10-program, 100-step programmer with ramp functions, enabling automated temperature profiles.

Determines whether the unit can run the required process ramps and external-sensor control without extra equipment.

Source
“selectable temperature control mode (Internal/Process), programmer with 10 programs (max. 100 steps), ramp function (linear and non-linear)”
Operation

The pump is a variable-speed soft-start unit with integrated pressure control and optional external pressure sensor, allowing adaptation to different external loops.

Supports matching pump performance to external circuit resistance and avoiding pressure shocks.

Source
“Powerful variable speed pump (soft start) with integrated pressure control with optional external pressure sensor.”
Installation

Pump connections are M30x1.5 male and cooling water connections are G1/2 male, so mating hoses and fittings must be ordered to match.

Prevents mismatch of process and cooling water plumbing during installation.

Source
“Pump connection M30x1,5 male”
Installation

The unit weighs 222 kg and measures 460x604x1465 mm, requiring adequate floor space and handling equipment.

Needed for site planning, access routes and floor loading.

Source
“Overall dimensions WxDxH ** 460x604x1465 mm”
Operation

Ambient temperature must stay between 5 °C and 40 °C and protection degree is IP20, so the location must be dry and within that range.

Out-of-range ambient or wet locations can cause malfunction or unsafe operation.

Source
“max. ambient temperature 40 °C”
Conformité

Voltage and frequency tolerances are ±5% and ±2% but must not be combined in opposite directions, so supply quality must be verified.

Avoids invalid supply combinations that could damage the compressor or void compliance.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

Three-phase devices below 63 A are delivered with cable but without a plug, so a suitable plug must be provided on site.

Affects electrical installation scope and commissioning schedule.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”