Unistat 825w Technical Data

Sommaire

  1. Overview p. 1
    Introduces the Unistat 825w unit and its main features
  2. Spec pp. 1-2
    Lists performance, temperature, and physical specifications per DIN 12876
  3. p. 2 table, 23 rows
  4. p. 1 table, 31 rows

Technical data according to DIN 12876

ParameterValue
Circulation pump: max. delivery40 l/min
max. delivery pressure0,9 bar
Delivery at 0,1 bar30 l/min
Delivery at 0,3 bar25 l/min
Delivery at 0,5 bar18 l/min
Pump connectionM30x1,5 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 0°C231 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity3 l
Filling capacity expansion tank3,6 l
Overall dimensions WxDxH **460x604x1465 mm
Net weight223 kg
Power supply requirement400V 3~ 50Hz
max. current15 A
Fuse3x16A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S466954

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Unistat 825w

ParameterValue
Operating temperature range-85...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power3 kW
Cooling power with Thermooil at 250°C2,3 kW
Cooling power with Thermooil at 200°C2,3 kW
Cooling power with Thermooil at 100°C2,3 kW
Cooling power with Ethanol at 0°C2,4 kW
Cooling power with Ethanol at -20°C2,4 kW
Cooling power with Ethanol at -40°C2,4 kW
Cooling power with Ethanol at -60°C1,5 kW
Cooling power with Ethanol at -80°C0,3 kW
Cooling power with Ethanol at -85°C0,2 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity0,95 kg
CO2 equivalent2,0 t
Refrigerant 2nd stage (ASHRAE, GHS)R-23 (A1, H280)
Global Warming Potential (GWP)14800
Refrigerant quantity 2nd stage0,39 kg
CO2 equivalent 2nd stage5,8 t
UN-numberUN 2857

Order-No.: 1079.0002.01

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a 400 V 3-phase 50 Hz supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.

Determines the site power supply, cable and breaker sizing before installation.

Source
“Power supply requirement 400V 3~ 50Hz max. current 15 A Fuse 3x16A”
Installation

The cooling water circuit needs at least 1 bar differential pressure and must not exceed 6 bar, with G1/2 male connections.

Defines the required cooling water supply pressure and connection type for proper operation.

Source
“min. cooling water differential pressure 1 bar max. cooling water pressure 6 bar”
Operation

Cooling water inlet temperature must be kept between 5 °C and 20 °C, and pipes must be insulated if it can drop below the dew point.

Operating outside this range reduces cooling power or causes condensation, affecting reliability.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power drops significantly at low temperatures, from 2.4 kW at -40 °C to only 0.2 kW at -85 °C.

Critical for selecting the unit when deep cooling loads are expected near the lower limit.

Source
“at -80°C 0,3 kW at -85°C 0,2 kW”
Safety

The refrigerant R-452A has a GWP of 2141 and 0.95 kg charge, giving 2.0 t CO2 equivalent, which matters for F-gas compliance.

Affects environmental compliance, leak monitoring obligations and refrigerant handling.

Source
“Global Warming Potential (GWP) 2141 Refrigerant quantity 0,95 kg CO2 equivalent 2,0 t”
Safety

The second-stage refrigerant R-23 has a very high GWP of 14800 with 0.39 kg charge (5.8 t CO2e).

High-GWP refrigerant drives regulatory reporting and service requirements.

Source
“Global Warming Potential (GWP) 14800 Refrigerant quantity 2nd stage 0,39 kg CO2 equivalent 2nd stage 5,8 t”
Installation

Voltage and frequency tolerances are ±5% and ±2% but must not combine in the unfavorable direction (e.g., -5% voltage with +2% frequency is not allowed).

Prevents motor and electronics damage from out-of-tolerance supply conditions.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

Three-phase devices under 63 A are delivered with cable but without a plug, so a plug must be sourced and fitted on site.

Affects installation planning and electrical connection work.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation

The pump delivers 40 l/min max and 0.9 bar max pressure, with flow falling to 18 l/min at 0.5 bar.

Needed to verify the external circuit pressure drop and achievable flow.

Source
“max. delivery 40 l/min max. delivery pressure 0,9 bar”
Operation

The pump tolerates a maximum kinematic viscosity of 50 mm²/s, limiting the choice of heat transfer fluid at low temperatures.

Constrains heat transfer fluid selection, especially for low-temperature operation.

Source
“max. permissible kin. viscosity 50 mm²/s”
Installation

The unit needs a minimum filling capacity of 3 l and has a 3.6 l expansion tank.

Determines the minimum heat transfer fluid volume and expansion allowance for the loop.

Source
“min. filling capacity 3 l Filling capacity expansion tank 3,6 l”
Installation

Ambient temperature must stay between 5 °C and 40 °C, and the unit is IP20 rated.

Defines the acceptable room conditions and protection level at the installation site.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Conformité

The equipment is EMC classified as Class A, Group 1 to EN55011, which is intended for industrial environments.

Class A equipment may cause interference in residential areas, affecting site selection.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Safety

The unit is classified as pressure equipment category I and safety class III/FL per DIN 12876.

Relevant for regulatory compliance and pressure vessel documentation.

Source
“Pressure equipment category I”