Unistat 920w Technical Data

Sommaire

  1. Overview p. 1
    Product introduction and general description of the Unistat 920w unit
  2. Spec pp. 1-2
    Technical performance data and operating parameters per DIN 12876
  3. p. 2 table, 23 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
max. delivery pressure2,5 bar
Delivery at 0,5 bar148 l/min
Delivery at 1,0 bar126 l/min
Delivery at 1,5 bar100 l/min
Delivery at 2,0 bar66 l/min
Pump connectionM38x1,5 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG3/4 male
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity12 l
Filling capacity expansion tank29 l
Overall dimensions WxDxH **954x1504x1622 mm
Net weight893 kg
sound pressure level +/- 4 dB(A)71 dB(A)
Power supply requirement400V 3~ 50Hz
max. current42 A
Fuse3x63A
Pressure equipment categoryII
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S588737

** Please respect space requirements. See operating conditions at www.huber-online.com

Unistat 920w — Technical data according to DIN 12876

ParameterValue
Order-No.1061.0022.01
Operating temperature range-90...200 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power12 kW
Cooling power with Thermooil at 200°C11 kW
Cooling power with Thermooil at 100°C11 kW
Cooling power with Thermooil at 20°C11 kW
Cooling power with Ethanol at 0°C11 kW
Cooling power with Ethanol at -20°C11 kW
Cooling power with Ethanol at -40°C10 kW
Cooling power with Ethanol at -60°C8 kW
Cooling power with Ethanol at -80°C2 kW
Cooling power with Ethanol at -90°C1 kW
Refrigeration machinewater-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-1270 (A3, H220)
Global Warming Potential (GWP)0
Refrigerant quantity1,75 kg
CO2 equivalent0,0 t
Refrigerant 2nd stage (ASHRAE, GHS)R-170 (A3, H220)
Global Warming Potential (GWP)0,437
Refrigerant quantity 2nd stage0,8 kg
CO2 equivalent 2nd stage0,0 t
UN-numberUN 3358
Circulation pump: MK pump max. delivery168 l/min

Таблица приведена в исходном виде (характеристика — значение); строки охлаждения сгруппированы по теплоносителю (Thermooil / Ethanol) и температуре.

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a 400 V three-phase supply with a 63 A fuse, so the site electrical infrastructure must be sized accordingly.

Determines the electrical supply, cable and protection device needed before installation.

Source
“Power supply requirement 400V 3~ 50Hz”
Installation

The cooling water circuit needs at least 1 bar differential pressure and must not exceed 6 bar, so the building water supply must be checked.

Insufficient or excessive cooling water pressure will prevent rated cooling performance or damage the circuit.

Source
“min. cooling water differential pressure 1 bar”
Installation

The unit is water-cooled and must be fed with cooling water between 5 °C and 20 °C, and pipes must be insulated if condensation can occur.

Cooling water temperature directly affects achievable cooling power and condensation risk.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power drops sharply at low temperatures, from 11 kW at -20 °C to only 1 kW at -90 °C, so the process load at the cold end must be small.

Prevents oversizing expectations for deep-cold process cooling duty.

Source
“at -90°C 1 kW”
Installation

The pump connections are M38x1.5 male with DIN 3863 ball-socket geometry, so matching hoses and fittings must be ordered.

Ensures compatible pipework and avoids leaks or adapters at commissioning.

Source
“Pump connection M38x1,5 male”
Operation

The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s, which limits fluid choice at low temperatures.

Wrong fluid viscosity can overload the pump and reduce circulation.

Source
“max. permissible kin. viscosity 50 mm²/s”
Safety

The refrigerant is R-1270 (A3, H220), a flammable natural refrigerant, so the installation area must be assessed for flammable refrigerant safety.

Flammable refrigerant classification affects room ventilation, ignition sources and safety compliance.

Source
“R-1270 (A3, H220)”
Safety

The unit has safety classification III / FL and adjustable overtemperature protection according to DIN 12876.

Confirms suitability for flammable fluids and the required protective measures.

Source
“Safety classification III / FL”
Installation

The unit weighs 893 kg and measures 954x1504x1622 mm, so floor loading and access route must be planned.

Affects rigging, floor strength and doorway clearance during installation.

Source
“Net weight 893 kg”
Operation

The ambient temperature must stay between 5 °C and 40 °C for correct operation.

Out-of-range ambient conditions can degrade performance and trip alarms.

Source
“min. ambient temperature 5 °C”
Conformité

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.

Site power quality must be verified to avoid invalid supply conditions and warranty issues.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

Three-phase devices below 63 A are delivered with cable but without a plug, so a suitable plug must be provided on site.

Prevents commissioning delays due to missing power connection hardware.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation

The unit has a 29 l expansion tank and a minimum filling capacity of 12 l, which sets the minimum fluid charge.

Ensures enough heat transfer fluid is available for safe circulation and expansion.

Source
“min. filling capacity 12 l”
Maintenance

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Incompatible water chemistry can cause corrosion and fouling of the cooling circuit.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”