Unistat 930w Technical Data

Sommaire

  1. Overview p. 1
    Product introduction and general description of the Unistat 930w
  2. Spec p. 2
    Technical data according to DIN 12876 standard
  3. p. 2 table, 21 rows
  4. p. 1 table, 32 rows

Technical data according to DIN 12876

ParameterValue
Delivery at 1,0 bar126 l/min
Delivery at 1,5 bar100 l/min
Delivery at 2,0 bar66 l/min
Pump connectionM38x1,5 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG3/4 male
Consumption at water 15°C, flow 0°C1300 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity12 l
Filling capacity expansion tank29 l
Overall dimensions WxDxH **950x1205x1650 mm
Net weight984 kg
Power supply requirement460V 3~ 60Hz
max. current66 A
Fuse3x80A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.1/25

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ParameterValue
Operating temperature range-90...200 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power32 kW
Cooling power with Thermooil at 200°C19 kW
Cooling power with Thermooil at 100°C19 kW
Cooling power with Thermooil at 20°C19 kW
Cooling power with Ethanol at 0°C20 kW
Cooling power with Ethanol at -20°C20 kW
Cooling power with Ethanol at -40°C20 kW
Cooling power with Ethanol at -60°C15 kW
Cooling power with Ethanol at -80°C5 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity5 kg
CO2 equivalent10,7 t
Refrigerant 2nd stage (ASHRAE, GHS)R-23 (A1, H280)
Global Warming Potential (GWP)14800
Refrigerant quantity 2nd stage2,2 kg
CO2 equivalent 2nd stage32,6 t
Circulation pump: MK pump max. delivery168 l/min
Circulation pump: MK pump max. delivery pressure2,5 bar
Circulation pump: MK pump Delivery at 0,5 bar148 l/min

Order-No.: 1082.0002.01; Unistat 930w

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a water-cooled refrigeration machine, so a cooling water supply with at least 1 bar differential pressure and no more than 6 bar must be provided at the G3/4 male connection.

Cooling water piping and pressure regulation must be designed before installation; insufficient differential pressure will prevent rated cooling.

Source
“Cooling water connection G3/4 male”
Installation

The cooling water intake temperature must be kept between 5°C and 20°C, and if it can fall below the condensation point the cooling water pipes must be insulated.

Site cooling water conditions determine whether the specified cooling power is achievable and whether pipe insulation is mandatory.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Rated output data are valid only at 20°C room temperature, 15°C cooling water inlet and 1 bar differential pressure, so higher water temperature reduces cooling power and may require higher flow.

Selection must account for actual site water temperature; otherwise the quoted 19–20 kW cooling capacity will not be met.

Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and -outlet.”
Installation

The electrical supply must be 460 V 3~ 60 Hz with a maximum current of 66 A and 3x80 A fusing, and external branch circuit protection according to UL 489 is required.

The panel and feeder must be sized for 66 A and include UL 489 branch protection; the unit is delivered without a plug for currents above 63 A.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Three-phase devices drawing more than 63 A are supplied without cable and without plug, so the customer must provide the power cable and connection.

Installation planning must include procurement and termination of the power cable, since it is not part of the delivery.

Source
“Three-phase devices with current consumption greater than 63A --> without cable, without plug”
Safety

Leakage current exceeds 3.5 mA, which affects the choice of residual current protective device and earthing arrangement.

Standard 30 mA RCDs may be unsuitable; the electrical design must account for high leakage current.

Source
“Attention: leakage current > 3,5mA”
Conformité

Voltage and frequency tolerances are ±5% and ±2% respectively, but a -5% voltage combined with +2% frequency is not allowed.

Supply quality at the site must be checked against this combined tolerance rule before commissioning.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

The unit is rated IP20 and must be operated between 5°C and 40°C ambient temperature.

The installation location must be protected from water and kept within the ambient range, otherwise protection and performance are not guaranteed.

Source
“min. ambient temperature 5 °C”
Selection

The refrigerant charges are 5 kg R-452A (GWP 2141, 10.7 t CO2e) and 2.2 kg R-23 (GWP 14800, 32.6 t CO2e), which matters for F-gas reporting and leak detection obligations.

The CO2 equivalent above 5 t triggers periodic leak checks and record-keeping under F-gas regulations.

Source
“CO2 equivalent 2nd stage 32,6 t”
Installation

The pump connection is M38x1.5 male with bore shape Y (60°) to DIN 3863, and hoses must use ball socket to DIN 3863 and sleeve nut to DIN 3870.

External piping and flexible hoses must be ordered with the correct DIN fittings to avoid leaks and mismatch.

Source
“Pump connections: Bore shape Y (60°) according to DIN 3863”
Operation

The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s, which limits fluid selection at low temperatures.

The chosen thermal fluid must stay below this viscosity over the operating range or pump delivery and cooling will suffer.

Source
“max. permissible kin. viscosity 50 mm²/s”
Maintenance

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only suitable cooling water may be used.

Water quality must be controlled to prevent corrosion and fouling of these materials in the cooling circuit.

Source
“Please use suitable cooling water.”
Selection

The unit has a minimum filling capacity of 12 l and a 29 l expansion tank, which sets the minimum fluid charge and expansion allowance.

The external loop volume and fluid expansion must be checked against these values during system design.

Source
“min. filling capacity 12 l”
Safety

The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III / FL.

The overtemperature cut-out must be set for the application and is a key safety requirement for unattended operation.

Source
“With adjustable overtemperature protection according to DIN 12876.”