Unistat GL 570w Technical Data Sheet

Sommaire

  1. Overview p. 1
    Introduces the Unistat GL 570w unit and its main features
  2. Spec pp. 1-2
    Performance, temperature range, and dimensional specifications per DIN 12876
  3. Spec p. 2
    Detailed technical parameters according to DIN 12876 standard
  4. p. 2 table, 16 rows
  5. p. 1 table, 32 rows

Technical data according to DIN 12876

ParameterValue
Pump connectionM38x1,5 male
max. permissible kin. viscosity100 mm²/s
Cooling water connectionG3/4 male
min. cooling water differential pressure1 bar
max. cooling water differential pressure4 bar
max. cooling water pressure6 bar
max. System pressure6 bar
min. filling capacity8,9 l
Filling capacity expansion tank19,9 l
Overall dimensions WxDxH **918x963x1771 mm
Power supply requirement400V 3~ 50Hz
Pressure equipment categoryII
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/24

** Please respect space requirements. See operating conditions at www.huber-online.com

Unistat GL 570w — Technical data according to DIN 12876

ParameterValue
Order-No.5041.0001.01
Operating temperature range-50...200 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Interface analogoptional
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power48 kW
Cooling power with Thermooil at 200°C57 kW
Cooling power with Thermooil at 100°C57 kW
Cooling power with Thermooil at 20°C57 kW
Cooling power with Ethanol at 0°C50 kW
Cooling power with Ethanol at -20°C35 kW
Cooling power with Ethanol at -40°C19 kW
Cooling power with Ethanol at -50°C12 kW
Refrigeration machinewater-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-744 (A1, H280)
Global Warming Potential (GWP)1
UN-numberUN 2857
Circulation pump: MK pump
max. delivery196 l/min
max. delivery pressure2,5 bar
Delivery at 0,5 bar175 l/min
Delivery at 1,0 bar152 l/min
Delivery at 1,5 bar123 l/min
Delivery at 2,0 bar86 l/min

Значения приведены дословно по тексту страницы; строки «Cooling power with Thermooil» и «Cooling power with Ethanol» объединяют соответствующие температурные подпункты.

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a 400 V three-phase supply, so the site must provide a 3~ 50 Hz feed sized for the 48 kW heater.

Determines electrical infrastructure, cable sizing and whether the facility can support the unit at all.

Source
“Power supply requirement 400V 3~ 50Hz”
Installation

Three-phase devices below 63 A are delivered with cable but without a plug, so the installer must provide a suitable plug or hard-wired connection.

Affects electrical installation scope and procurement of the mating connector.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Cooling water must be supplied at 1–4 bar differential and no more than 6 bar, with 5–20 °C intake, so the site cooling loop must be checked before commissioning.

Insufficient or excessive cooling water pressure will prevent rated cooling performance or damage the water circuit.

Source
“min. cooling water differential pressure 1 bar max. cooling water differential pressure 4 bar”
Installation

If cooling water is colder than the condensation point, the cooling water pipes inside the unit must be insulated.

Prevents condensation and corrosion/electrical risk when using cold site water.

Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance

Rated cooling capacities are valid only at 20 °C room temperature, 15 °C cooling water inlet and 1 bar differential, and will drop at higher cooling water temperatures.

Prevents overestimating available cooling power when site conditions differ from the rating basis.

Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and -”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Guides water quality and treatment to avoid corrosion or fouling of the heat exchanger.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Operation

Voltage and frequency tolerances are ±5 % and ±2 % but must not combine in the same direction, so supply quality must be verified.

Out-of-tolerance supply can trip the unit or reduce lifetime; the example clarifies the allowed combinations.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Conformité

The unit is EMC Class A Group 1, so it is intended for industrial environments and may cause interference in residential areas.

Affects installation location and EMC planning in mixed-use or sensitive environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Safety

The unit is safety class III / FL per DIN 12876 and has adjustable overtemperature protection, which must be set for the application.

Overtemperature protection setting is a key safety measure for the thermal fluid and connected load.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Selection

The pump delivers 196 l/min max at 2.5 bar, dropping to 86 l/min at 2.0 bar, so system pressure drop must be matched to the pump curve.

Ensures the external circuit can achieve the required flow at its actual pressure loss.

Source
“max. delivery 196 l/min max. delivery pressure 2,5 bar”
Installation

Pump connections are M38x1.5 male and cooling water connections are G3/4 male, so mating hoses and adapters must be ordered accordingly.

Prevents delays and leaks from mismatched connections during installation.

Source
“Pump connection M38x1,5 male”
Operation

Ambient temperature must stay between 5 °C and 40 °C, and the unit is IP20, so it must be installed in a dry, protected location.

Defines the allowable plant room conditions and protection against water/dust ingress.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Maintenance

The refrigerant is R-744 (A1, H280) with GWP 1 and UN 2857, so service and transport must follow the corresponding safety data.

Affects refrigerant handling, labeling and emergency procedures during service.

Source
“Refrigerant (ASHRAE, GHS) R-744 (A1, H280)”
Selection

Minimum filling capacity is 8.9 l and the expansion tank holds 19.9 l, so the external loop volume plus unit volume must be considered when selecting fluid quantity.

Ensures sufficient thermal fluid is ordered and the expansion volume is adequate for the application.

Source
“min. filling capacity 8,9 l Filling capacity expansion tank 19,9 l”