Unistat P405w Technical Data

Sommaire

  1. Overview p. 1
    Introduces the Unistat P405w unit and its main features
  2. Spec pp. 1-2
    Performance, temperature range, and dimensional data per DIN 12876
  3. p. 2 table, 19 rows
  4. p. 1 table, 26 rows

Technical data according to DIN 12876

ParameterValue
max. delivery pressure3 bar
Delivery at 0,5 bar58 l/min
Delivery at 1,0 bar53 l/min
Delivery at 1,5 bar46 l/min
Delivery at 2,0 bar39 l/min
Delivery at 2,5 bar29 l/min
Pump connectionM30x1,5 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG1/2 male
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity2,6 l
Filling capacity expansion tank3,7 l
Overall dimensions WxDxH **460x620x1300 mm
Power supply requirement460V 3~ 60Hz
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/24

** Please respect space requirements. See operating conditions at www.huber-online.com

Unistat P405w — Technical data according to DIN 12876

CharacteristicValue
Order-No.1002.0071.01
Operating temperature range-45...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power3 kW
Cooling power with Thermooil at 250°C1,3 kW
Cooling power with Thermooil at 200°C1,3 kW
Cooling power with Thermooil at 100°C1,3 kW
Cooling power with Thermooil at 20°C1,3 kW
Cooling power with Ethanol at 0°C1,3 kW
Cooling power with Ethanol at -20°C0,5 kW
Cooling power with Ethanol at -40°C0,1 kW
Refrigeration machinewater-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-1270 (A3, H220)
Global Warming Potential (GWP)0
UN-numberUN 3358
Circulation pumpMK pump
max. delivery63 l/min

Engineering Facts

12 facts · every claim linked to a page
Installation

The unit requires a water-cooled refrigeration machine, so cooling water supply and return must be provided at the installation site.

Determines site requirements for cooling water connections and infrastructure before purchase.

Source
“Refrigeration machine water-cooled, natural refrigerant”
Safety

The refrigerant is R-1270 (A3, H220), a flammable natural refrigerant, which affects safety classification and handling requirements.

Flammable refrigerant classification impacts installation location, ventilation, and safety compliance.

Source
“R-1270 (A3, H220)”
Installation

Cooling water differential pressure must be at least 1 bar and must not exceed 6 bar at the connection.

Ensures adequate cooling water flow and protects the unit from overpressure during installation.

Source
“min. cooling water differential pressure 1 bar”
Installation

The unit is designed for cooling water intake temperatures between 5°C and 20°C; below the condensation point the cooling water pipes must be insulated.

Defines acceptable cooling water supply conditions and required insulation to prevent condensation.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power drops significantly at low temperatures with ethanol, from 1.3 kW at 0°C to 0.1 kW at -40°C.

Critical for selecting the unit for low-temperature applications where cooling capacity is limited.

Source
“at -40°C 0,1 kW”
Installation

The electrical supply must be 460V 3~ 60Hz, and voltage/frequency tolerances of ±5% and ±2% apply simultaneously.

Ensures correct power supply design and prevents invalid combinations such as -5% voltage with +2% frequency.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Conformité

External branch circuit protection according to UL 489 is required for this equipment.

Mandatory for electrical installation compliance in UL jurisdictions.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Three-phase devices with current consumption less than 63A are supplied with cable but without plug.

Installer must provide a suitable plug or hardwired connection at the site.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation

The expansion tank is thermally passive, preventing HTF vapour and moisture absorption, and for external open baths it is blocked off so the thermostat is atmospherically sealed.

Allows flexible placement below or above the application level and avoids moisture ingress.

Source
“No HTF vapour and no moisture absorption because the expansion tank is thermally passive.”
Installation

Pump connections use M30x1.5 male threads with bore shape Y (60°) according to DIN 3863, requiring matching ball socket and sleeve nut.

Ensures correct selection of hoses and fittings for leak-free connection.

Source
“Pump connection M30x1,5 male”
Operation

The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s.

Limits choice of heat transfer fluid to ensure proper pump operation and cooling performance.

Source
“max. permissible kin. viscosity 50 mm²/s”
Installation

The unit has IP20 protection and must be operated in ambient temperatures between 5°C and 40°C.

Restricts installation environment to dry indoor locations and defines ambient temperature limits.

Source
“Degree of Protection IP20”