Unistat P425w Technical Data

Sommaire

  1. Overview p. 1
    Product introduction and general description of the Unistat P425w
  2. Spec pp. 1-2
    Technical data and performance specifications per DIN 12876
  3. p. 2 table, 21 rows
  4. p. 1 table, 30 rows

Technical data according to DIN 12876

ParameterValue
Delivery at 2,0 bar48 l/min
Delivery at 2,5 bar30 l/min
Pump connectionM30x1,5 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 0°C150 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity3,6 l
Filling capacity expansion tank3,3 l
Overall dimensions WxDxH **460x554x1453 mm
Net weight178 kg
sound pressure level +/- 4 dB(A)61 dB(A)
Power supply requirement400V 3~ 50Hz
max. current10 A
Fuse3x16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S469298

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Unistat P425w (Order-No.: 1050.0033.01)

ParameterValue
Operating temperature range-40...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power2 kW
Cooling power with Thermooil at 250°C2,8 kW
Cooling power with Thermooil at 200°C2,8 kW
Cooling power with Thermooil at 100°C2,8 kW
Cooling power with Thermooil at 20°C2,8 kW
Cooling power with Ethanol at 0°C2,5 kW
Cooling power with Ethanol at -20°C1,8 kW
Cooling power with Ethanol at -40°C0,1 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity0,97 kg
CO2 equivalent2,1 t
Circulation pump: MK pump
max. delivery97 l/min
max. delivery pressure3 bar
Delivery at 0,5 bar86 l/min
Delivery at 1,0 bar75 l/min
Delivery at 1,5 bar63 l/min

Значения приведены дословно по тексту страницы; строки «Cooling power with Thermooil» и «Cooling power with Ethanol» объединены с соответствующими температурными точками, так как в исходном тексте они напечатаны как отдельные подзаголовки.

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a 400 V three-phase supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.

Determines the site power supply and breaker/fuse rating needed before installation.

Source
“Power supply requirement 400V 3~ 50Hz”
Installation

The cooling water circuit must be supplied at 1–6 bar with 150 l/h at 15 °C, and the intake water must stay between 5 °C and 20 °C.

Cooling water pressure, flow and temperature limits dictate the facility water supply design.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

If cooling water is colder than the condensation point, the cooling water pipes inside the unit must be insulated.

Prevents condensation and corrosion/electrical risk when using cold facility water.

Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Installation

The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Guides water quality and material compatibility to avoid corrosion or fouling.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation

The pump connection is M30x1.5 male and the cooling water connection is G1/2 male, so matching hoses/fittings must be ordered.

Ensures correct plumbing adapters and hoses are available at installation.

Source
“Pump connection M30x1,5 male”
Installation

Pump connections use bore shape Y (60°) per DIN 3863 with ball socket and sleeve nut per DIN 3870, so standard DIN fittings are required.

Specifies the exact fitting standard to avoid leaks or mismatched connections.

Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Installation

The unit is IP20 and must be installed where ambient temperature stays between 5 °C and 40 °C.

Defines environmental limits and protection against dust/water at the installation site.

Source
“Degree of Protection IP20”
Installation

The unit weighs 178 kg and measures 460x554x1453 mm, so floor loading and access clearance must be planned.

Needed for rigging, floor loading and space planning during installation.

Source
“Overall dimensions WxDxH ** 460x554x1453 mm”
Operation

The operating temperature range is -40 to 250 °C with 0.01 K stability, defining the process window.

Confirms the unit can cover the required process temperatures with high stability.

Source
“Operating temperature range -40...250 °C”
Performance

Cooling power with ethanol drops to 0.1 kW at -40 °C, so low-temperature capacity is very limited.

Critical for sizing heat loads at the cold end of the range.

Source
“at -40°C 0,1 kW”
Performance

The pump delivers 97 l/min max and 3 bar max, with 86 l/min at 0.5 bar, so pressure-drop calculations are needed.

Determines whether the pump can overcome the external circuit pressure drop.

Source
“max. delivery 97 l/min”
Safety

Leakage current exceeds 3.5 mA, so protective earth and RCD selection must account for it.

Affects electrical safety design and residual current device selection.

Source
“Attention: leakage current > 3,5mA”
Conformité

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and -5% voltage with +2% frequency is not allowed.

Prevents operation outside EN60034-1 tolerances that could damage the compressor/pump.

Source
“-5% voltage and +2% frequency -> not allowed!”
Conformité

The refrigerant is R-452A with GWP 2141 and 0.97 kg charge (2.1 t CO2e), relevant for F-gas reporting.

Needed for refrigerant compliance, leak reporting and CO2-equivalent accounting.

Source
“Refrigerant (ASHRAE, GHS) R-452A (A1, H280)”