Installation
The unit requires a three-phase 400 V supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.
Determines the site power supply, cable and protection device rating before installation.
Source
“Power supply factory configured (3 Phase) 400V 3~ 50Hz”
Installation
Three-phase devices below 63 A are delivered with a cable but without a plug, so the customer must provide and fit the plug.
Affects installation planning and the bill of materials for the electrical connection.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety
Leakage current exceeds 3.5 mA, which may require a suitable residual current device or protective earthing arrangement.
Important for electrical safety design and RCD selection at the installation site.
Source
“Attention: leakage current > 3,5mA”
Installation
The pump connections use M30x1.5 male threads with Y-bore and ball-socket fittings per DIN 3863/3870, so matching hoses and fittings must be ordered.
Ensures correct mating of process pipework and flexible tempering hoses.
Source
“Pump connection M30x1,5 male”
Operation
The unit must be operated between 5 °C and 40 °C ambient, and cooling capacity may drop above 20 °C room temperature.
Affects site ventilation and expected performance in warm rooms.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Performance
Cooling capacity falls sharply at low temperature with ethanol, from 3.5 kW at 0 °C to only 0.15 kW at -40 °C.
Critical for sizing the unit for low-temperature applications near the lower limit.
Source
“at -40°C 0,15 kW”
Selection
The pump delivers 97 l/min maximum but only 30 l/min at 2.5 bar, so high-pressure-drop loops need careful pump curve checking.
Determines whether the circulation pump can supply the required flow through restrictive external circuits.
Source
“Delivery at 2,5 bar 30 l/min”
Selection
The refrigerant is R-1270 (A3, H220) with GWP 0 and 0.75 kg charge, which is flammable and requires appropriate site precautions.
Flammable refrigerant classification affects room safety, ventilation and insurance requirements.
Source
“R-1270 (A3, H220)”
Installation
The unit is IP20 and weighs 197 kg with dimensions 460x570x1629 mm, so it needs a dry indoor location and adequate floor space and handling equipment.
Needed for site layout, access routes and floor loading assessment.
Source
“Overall dimensions WxDxH ** 460x570x1629 mm”
Operation
Minimum filling capacity is 4 l and the expansion tank holds 3.7 l, so the external loop volume must be checked against these limits.
Prevents underfilling or overflow of the expansion tank in external circuits.
Source
“min. filling capacity 4 l”
Conformité
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.
Must be verified against the site supply quality to avoid voiding performance or warranty.
Source
“-5% voltage and +2% frequency -> not allowed!”
Safety
The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III/FL.
Relevant for protecting temperature-sensitive process fluids and meeting safety requirements.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Operation
The controller supports external Pt100 sensors and internal/process control modes, enabling control at the external process point.
Allows accurate control at the application rather than at the circulator outlet.
Source
“selectable temperature control mode (Internal/Process)”