Unistat P620w Technical Data

Sommaire

  1. Overview p. 1
    Product introduction and general description of the Unistat P620w unit
  2. Spec p. 2
    Technical data according to DIN 12876 standard
  3. p. 2 table, 17 rows
  4. p. 1 table, 31 rows

Technical data according to DIN 12876

ParameterValue
Delivery at 5,0 bar42 l/min
Pump connectionM38x1,5 male
max. permissible kin. viscosity100 mm²/s
Cooling water connectionG3/4 male
min. cooling water differential pressure1 bar
max. cooling water differential pressure4 bar
max. cooling water pressure6 bar
max. System pressure6 bar
min. filling capacity8,9 l
Filling capacity expansion tank19,9 l
Overall dimensions WxDxH **918x963x1771 mm
Power supply requirement400V 3~ 50Hz
Pressure equipment categoryII
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/24

** Please respect space requirements. See operating conditions at www.huber-online.com

Unistat P620w — Technical data according to DIN 12876

CharacteristicValue
Order-No.5016.0002.01
Operating temperature range-60...200 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Interface analogoptional
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power12 kW
Cooling power with Thermooil at 200°C12 kW
Cooling power with Thermooil at 100°C12 kW
Cooling power with Thermooil at 20°C12 kW
Cooling power with Ethanol at 0°C12 kW
Cooling power with Ethanol at -20°C12 kW
Cooling power with Ethanol at -40°C5,6 kW
Cooling power with Ethanol at -60°C1,4 kW
Refrigeration machinewater-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-1270 (A3, H220)
Global Warming Potential (GWP)0
Circulation pumpMK pump
max. delivery201 l/min
max. delivery pressure5,3 bar
Delivery at 1,0 bar179 l/min
Delivery at 2,0 bar154 l/min
Delivery at 3,0 bar124 l/min
Delivery at 4,0 bar87 l/min

Engineering Facts

12 facts · every claim linked to a page
Installation

The unit requires a 400 V three-phase supply, so the electrical infrastructure must be verified before installation.

Determines whether the site has the correct three-phase power available and affects cabling and breaker sizing.

Source
“Power supply requirement 400V 3~ 50Hz”
Installation

Cooling water must be supplied at 1–4 bar differential pressure and no more than 6 bar, so the facility water loop must be checked against these limits.

Incorrect cooling water pressure can reduce cooling capacity or damage the water circuit.

Source
“min. cooling water differential pressure 1 bar max. cooling water differential pressure 4 bar max. cooling water pressure 6 bar”
Operation

The unit is designed for cooling water inlet temperatures between 5 °C and 20 °C, and lower temperatures require insulating the cooling water pipes.

Operating outside this range risks condensation and reduced performance.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling capacity drops significantly at low temperatures with ethanol, from 12 kW at -20 °C to 5.6 kW at -40 °C and 1.4 kW at -60 °C.

Process design must account for reduced cooling power at the lowest operating temperatures.

Source
“at -40°C 5,6 kW at -60°C 1,4 kW”
Installation

The pump connection is M38x1.5 male, so mating hoses or adapters must match this thread.

Ensures correct connection of external circuit plumbing and avoids leaks or adapters.

Source
“Pump connection M38x1,5 male”
Safety

The refrigerant is R-1270 (A3, H220), a flammable natural refrigerant, which affects site safety classification and handling procedures.

Flammable refrigerant requires appropriate safety measures and may affect insurance or room classification.

Source
“R-1270 (A3, H220)”
Conformité

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.

Power quality must be verified to avoid voiding warranty or causing malfunction.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

Three-phase devices with current consumption less than 63 A are supplied with cable but without a plug, so a suitable plug must be provided on site.

Affects electrical installation planning and cost.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation

The maximum permissible kinematic viscosity of the heat transfer fluid is 100 mm²/s, which limits fluid selection at low temperatures.

Using a fluid with too high viscosity can overload the pump and reduce flow.

Source
“max. permissible kin. viscosity 100 mm²/s”
Installation

The unit has IP20 protection, so it must be installed in a dry indoor environment protected from water ingress.

Prevents installation in wet or dusty environments that could damage electronics.

Source
“Degree of Protection IP20”
Operation

The ambient temperature must be between 5 °C and 40 °C for proper operation.

Ensures the unit operates within its design envelope and maintains performance.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Maintenance

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so suitable cooling water quality must be used to avoid corrosion.

Poor water quality can cause corrosion and fouling, reducing cooling efficiency and requiring maintenance.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”