Product introduction and general description of Unistat P905
Specpp. 1-2
Technical data and performance specifications per DIN 12876
p. 2table, 21 rows
p. 1table, 31 rows
Technical data according to DIN 12876
Parameter
Value
max. delivery
65 l/min
max. delivery pressure
3 bar
Delivery at 0,5 bar
60 l/min
Delivery at 1,0 bar
54 l/min
Delivery at 1,5 bar
47 l/min
Delivery at 2,0 bar
39 l/min
Delivery at 2,5 bar
29 l/min
Pump connection
M30x1,5 male
max. permissible kin. viscosity
50 mm²/s
min. filling capacity
3,2 l
Filling capacity expansion tank
5,3 l
Overall dimensions WxDxH **
540x654x1500 mm
Net weight
272 kg
Power supply factory configured (3 Phase)
460V 3~ 60Hz
max. current (3 Phase)
19 A
Fuse (3 phase)
3x20A
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
35 °C
from Serial-No.:
S573024
** Please respect space requirements. See operating conditions at www.huber-online.com
Unistat P905 Technical data according to DIN 12876
Characteristic
Value
Order-No.
1054.0011.01
Operating temperature range
-90...250 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
6 kW
Cooling power with Thermooil at 250°C
3,6 kW
Cooling power with Thermooil at 200°C
3,6 kW
Cooling power with Thermooil at 100°C
3,6 kW
Cooling power with Ethanol at 0°C
3,6 kW
Cooling power with Ethanol at -20°C
3,5 kW
Cooling power with Ethanol at -40°C
3,5 kW
Cooling power with Ethanol at -60°C
2 kW
Cooling power with Ethanol at -80°C
0,4 kW
Refrigeration machine
air-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-452A (A1, H280)
Global Warming Potential (GWP)
2141
Refrigerant quantity
2,5 kg
CO2-Äquvivalent
5,4 t
Refrigerant 2nd stage (ASHRAE, GHS)
R-23 (A1, H280)
Global Warming Potential (GWP)
14800
Refrigerant quantity 2nd stage
0,75 kg
CO2-Äquvivalent 2. Stufe
11,1 t
Circulation pump:
MK pump
Engineering Facts
14 facts · every claim linked to a page
Selection
The unit is a refrigerated heating circulator with an air-cooled refrigeration machine, suitable for both externally closed and externally open applications.
Confirms the equipment type and that it can serve open external loops, which affects loop design and accessories.
Source
“Refrigerated Heating Circulator with air-cooled cooling machine and optical level indicator.”
Selection
The 'P' model pump is intended for high-pressure-drop applications such as flow-through chemistry or semiconductor industry.
Determines whether this pump variant matches your pressure-drop requirements before ordering.
Source
“Unistat „P“ Models: Circulating pumps with a high discharge pressure for applications with high pressure drops, e.g. in the Flow-Through chemistry or in the Semicon industry.”
Installation
The pump connection is M30x1.5 male, so mating hoses or fittings must match this thread.
Prevents ordering incompatible hoses or adapters during installation.
Source
“Pump connection M30x1,5 male”
Installation
Piping and flexible tempering hoses must use ball sockets per DIN 3863 and sleeve nuts per DIN 3870.
Specifies the exact fitting standard required for leak-free connections.
Source
“pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Installation
The three-phase unit is supplied with cable but without a plug, so a suitable plug must be field-installed.
Affects electrical installation planning and bill of materials.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
External branch circuit protection according to UL 489 is required.
Mandates specific upstream protection hardware in the electrical panel.
Source
“External branch circuit protection according to UL 489 required.”
Safety
Leakage current exceeds 3.5 mA, which may require additional protective measures such as a higher-rated RCD.
Impacts residual-current device selection and electrical safety compliance.
Source
“Attention: leakage current > 3,5mA”
Operation
Cooling capacity is rated at 20 °C room temperature and may drop if ambient temperature rises.
Requires derating or ventilation planning in warm rooms to meet process cooling needs.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Operation
The unit must be operated within an ambient temperature range of 5 °C to 35 °C.
Defines the allowable site conditions for reliable operation.
Source
“min. ambient temperature 5 °C”
Operation
Voltage and frequency tolerances are ±5% and ±2% respectively, but not simultaneously in opposite directions.
Guides power quality assessment at the installation site.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Conformité
The equipment is classified as Class A, Group 1 for electromagnetic disturbance per EN55011.
Class A equipment may require additional EMC measures in residential or light-industrial environments.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Maintenance
The refrigerant is R-452A with a GWP of 2141 and a charge of 2.5 kg, giving a CO2 equivalent of 5.4 t.
Relevant for refrigerant handling, leak reporting thresholds, and environmental compliance.
Source
“Refrigerant quantity 2,5 kg”
Maintenance
The second-stage refrigerant is R-23 with a GWP of 14800 and a charge of 0.75 kg, giving 11.1 t CO2 equivalent.
Affects service procedures and environmental reporting for the cascade circuit.
Source
“Refrigerant quantity 2nd stage 0,75 kg”
Selection
The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s.
Constrains fluid selection, especially at low temperatures where viscosity rises.